Heterogeneous physical phantom for I-125 dose measurements and dose-to-medium determination.
Antunes, Paula Cristina Guimarães; Siqueira, Paulo de Tarso Dalledone; Shorto, Julian Marco Barbosa; Yoriyaz, Hélio.
Afiliação
  • Antunes PCG; Instituto de Pesquisas Energéticas e Nucleares - IPEN-CNEN/SP, Sao Paulo, Brazil; Institute of Physics, University of Sao Paulo, Sao Paulo, Brazil. Electronic address: pacrisguian@gmail.com.
  • Siqueira PTD; Instituto de Pesquisas Energéticas e Nucleares - IPEN-CNEN/SP, Sao Paulo, Brazil.
  • Shorto JMB; Instituto de Pesquisas Energéticas e Nucleares - IPEN-CNEN/SP, Sao Paulo, Brazil.
  • Yoriyaz H; Instituto de Pesquisas Energéticas e Nucleares - IPEN-CNEN/SP, Sao Paulo, Brazil.
Brachytherapy ; 23(1): 73-84, 2024.
Article em En | MEDLINE | ID: mdl-38016863
PURPOSE: In this paper we present a further step in the implementation of a physical phantom designed to generate sets of "true" independent reference data as requested by TG-186, intending to address and mitigate the scarcity of experimental studies on brachytherapy (BT) validation in heterogeneous media. To achieve this, we incorporated well-known heterogeneous materials into the phantom in order to perform measurements of 125I dose distribution. The work aims to experimentally validate Monte Carlo (MC) calculations based on MBDCA and determine the conversion factors from LiF response to absorbed dose in different media, using cavity theory. METHODS AND MATERIALS: The physical phantom was adjusted to incorporate tissue equivalent materials, such as: adipose tissue, bone, breast and lung with varying thickness. MC calculations were performed using MCNP6.2 code to calculate the absorbed dose in the LiF and the dose conversion factors (DCF). RESULTS: The proposed heterogeneous phantom associated with the experimental procedure carried out in this work yielded accurate dose data that enabled the conversion of the LiF responses into absorbed dose to medium. The results showed a maximum uncertainty of 6.92 % (k = 1), which may be considered excellent for dosimetry with low-energy BT sources. CONCLUSIONS: The presented heterogeneous phantom achieves the required precision in dose evaluations due to its easy reproducibility in the experimental setup. The obtained results support the dose conversion methodology for all evaluated media. The experimental validation of the DCF in different media holds great significance for clinical procedures, as it can be applied to other tissues, including water, which remains a widely utilized reference medium in clinical practice.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Temas: Geral Base de dados: MEDLINE Assunto principal: Braquiterapia / Radioisótopos do Iodo Limite: Humans Idioma: En Revista: Brachytherapy Assunto da revista: RADIOTERAPIA Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Temas: Geral Base de dados: MEDLINE Assunto principal: Braquiterapia / Radioisótopos do Iodo Limite: Humans Idioma: En Revista: Brachytherapy Assunto da revista: RADIOTERAPIA Ano de publicação: 2024 Tipo de documento: Article