Clinical outcomes of a toric continuous range of vision presbyopia-correcting intraocular lens.
Orbegozo, Javier; Pérez, Angelica; Basterra, Iñaki; Noguera, Helena; González, Laura; Lauzirika, Gorka; Piñero, David P.
Afiliação
  • Orbegozo J; From the Centro Oftalmológico Integral Bilbao Berri SL, Miranza COI Bilbao, Bilbao, Spain (Orbegozo, Pérez, Basterra, Noguera); R&D Department of Miranza Group, Barcelona, Spain (González, Lauzirika); Instituto de Microcirugía Ocular (IMO) Barcelona, Miranza Group, Barcelona, Spain (González); Miranza Begitek, Donostia-San Sebastián, Spain (Lauzirika); Department of Optics, Pharmacology and Anatomy, University of Alicante, Alicante, Spain (Piñero).
J Cataract Refract Surg ; 50(7): 724-732, 2024 Jul 01.
Article em En | MEDLINE | ID: mdl-38530027
ABSTRACT

PURPOSE:

To evaluate the clinical outcomes of the toric version of a presbyopia-correcting intraocular lens (IOL) based on the combination of a diffractive-based extended depth-of-focus (EDOF) pattern and a diffractive multifocal platform.

SETTING:

Miranza COI Bilbao, Bilbao, Spain.

DESIGN:

Prospective case series.

METHODS:

35 patients (51 to 84 years) with corneal astigmatism ranging from 0.75 to 2.19 diopters (D) undergoing bilateral cataract surgery with implantation of the Synergy Toric II IOL were evaluated during a 3-month follow-up. Visual acuity, refraction, defocus curve, and patient-reported outcomes with the Catquest-9SF questionnaire were analyzed. A vectorial analysis was used to analyze the accuracy of astigmatic correction.

RESULTS:

Mean 3-month monocular postoperative uncorrected distance, intermediate (80 cm) and near (40 cm) visual acuities were 0.06 ± 0.11 logMAR, 0.13 ± 0.12 logMAR, and 0.13 ± 0.09 logMAR, respectively. Mean monocular distance-corrected intermediate (80 cm) and near visual acuity (40 cm) were 0.11 ± 0.12 logMAR and 0.10 ± 0.10 logMAR, respectively. Mean binocular defocus curve showed visual acuities of 0.10 logMAR or better for defocus levels from +0.50 to -2.50 D. Residual cylinder was within ±0.50 D in 97.0% of eyes. The surgically induced astigmatism prediction error ranged between -0.49 D and 0.50 D, with a mean value of 0.04 ± 0.16 D. Mean absolute IOL rotation was 3.79 ± 2.94 degrees. Significant improvements were found in all Rasch-calibrated scores obtained with Catquest-9SF ( P < .001).

CONCLUSIONS:

The implantation of the toric presbyopia-correcting IOL evaluated provides an efficacious astigmatic correction while providing a fully restoration of the visual function across different distances.
Assuntos

Texto completo: 1 Coleções: 01-internacional Temas: Geral Base de dados: MEDLINE Assunto principal: Presbiopia / Refração Ocular / Astigmatismo / Acuidade Visual / Implante de Lente Intraocular / Lentes Intraoculares Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Cataract Refract Surg / J. cataract. refract. surg / Journal of cataract & refractive surgery Assunto da revista: OFTALMOLOGIA Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Temas: Geral Base de dados: MEDLINE Assunto principal: Presbiopia / Refração Ocular / Astigmatismo / Acuidade Visual / Implante de Lente Intraocular / Lentes Intraoculares Limite: Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: J Cataract Refract Surg / J. cataract. refract. surg / Journal of cataract & refractive surgery Assunto da revista: OFTALMOLOGIA Ano de publicação: 2024 Tipo de documento: Article