1.
Circ Arrhythm Electrophysiol ; 17(7): e012854, 2024 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-38758741

RESUMO

BACKGROUND: Unlike conventional microsecond pulsed electrical fields that primarily target the cell membranes, nanosecond pulses are thought to primarily electroporate intracellular organelles. We conducted a comprehensive preclinical assessment of catheter-based endocardial nanosecond pulsed field ablation in swine. METHODS: A novel endocardial nanosecond pulsed field ablation system was evaluated in a total of 25 swine. Using either a low-dose (5-second duration) or high-dose (15-second duration) strategy, thoracic veins and discrete atrial and ventricular sites were ablated. Predetermined survival periods were <1 (n=1), ≈2 (n=7), ≈7 (n=6), 14 (n=2), or ≈28 (n=9) days, and venous isolation was assessed before euthanasia. Safety assessments included evaluation of esophageal effects, phrenic nerve function, and changes in venous caliber. All tissues were subject to careful gross pathological and histopathologic examination. RESULTS: All (100%) veins (13 low-dose, 34 high-dose) were acutely isolated, and all reassessed veins (6 low-dose, 15 high-dose) were durably isolated. All examined vein lesions (10 low-dose, 22 high-dose) were transmural. Vein diameters (n=15) were not significantly changed. Of the animals assessed for phrenic palsy (n=9), 3 (33%) demonstrated only transient palsy. There were no differences between dosing strategies. Thirteen mitral isthmus lesions were analyzed, and all 13 (100%) were transmural (depth, 6.4±0.4 mm). Ventricular lesions were 14.7±4.5 mm wide and 7.1±1.3 mm deep, with high-dose lesions deeper than low-dose (7.9±1.2 versus 6.2±0.8 mm; P=0.007). The esophagus revealed nontransmural adventitial surface lesions in 5 of 5 (100%) animals euthanized early (2 days) post-ablation. In the 10 animals euthanized later (14-28 days), all animals demonstrated significant esophageal healing-8 with complete resolution, and 2 with only trace fibrosis. CONCLUSIONS: A novel, endocardial nanosecond pulsed field ablation system provides acute and durable venous isolation and linear lesions. Transient phrenic injury and nontransmural esophageal lesions can occur with worst-case assessments suggesting limits to pulsed field ablation tissue selectivity and the need for dedicated assessments during clinical studies.


Assuntos
Estudos de Viabilidade , Nervo Frênico , Animais , Suínos , Fatores de Tempo , Miocárdio/patologia , Ablação por Cateter/métodos , Ablação por Cateter/efeitos adversos , Veias/fisiopatologia , Modelos Animais , Ventrículos do Coração/fisiopatologia , Ventrículos do Coração/patologia , Esôfago , Átrios do Coração/fisiopatologia , Átrios do Coração/patologia
2.
JAMA Cardiol ; 6(6): 697-705, 2021 06 01.
Artigo em Inglês | MEDLINE | ID: mdl-33909022

RESUMO

Importance: Early rhythm control of atrial fibrillation (AF) with either antiarrhythmic drugs (AADs) or catheter ablation has been reported to improve cardiovascular outcomes compared with usual care; however, the optimal therapeutic modality to achieve early rhythm control is unclear. Objective: To assess the safety and efficacy of AF ablation as first-line therapy when compared with AADs in patients with paroxysmal AF. Data Sources: PubMed/MEDLINE, Scopus, Google Scholar, and various major scientific conference sessions from January 1, 2000, through November 23, 2020. Study Selection: Randomized clinical trials (RCTs) published in English that had at least 12 months of follow-up and compared clinical outcomes of ablation vs AADs as first-line therapy in adults with AF. The quality of individual studies was assessed using the Cochrane risk of bias tool. Six RCTs met inclusion criteria, including 1212 patients. Data Extraction and Synthesis: Two investigators independently extracted data. Reporting was performed in compliance with the PRISMA (Preferred Reporting Items for Systematic Reviews and Meta-analysis) guidelines. Analysis was performed using a random-effects model with the Mantel-Haenszel method, and results are presented as 95% CIs. Main Outcomes and Measures: Main outcomes were safety and efficacy of AF ablation as first-line therapy when compared with AADs. Trials were evaluated as having low risk of selection and attrition biases, high risk of performance bias, and with unclear risk for detection biases due to unblinding and open-label designs. Results: A total of 6 RCTs involving 1212 patients with AF were included (609 were randomized to AF ablation and 603 to drug therapy; mean [SD] age, 56 [11] years). Compared with AADs, catheter ablation use was associated with reductions in recurrent atrial arrhythmia (32.3% vs 53%; risk ratio [RR], 0.62; 95% CI, 0.51-0.74; P < .001; I2 = 40%), with a number needed to treat with ablation to prevent 1 arrhythmia of 5. Use of ablation was also associated with reduced symptomatic atrial arrhythmia (11.8% vs 26.4%; RR, 0.44; 95% CI, 0.27-0.72; P = .001; I2 = 54%) and hospitalization (5.6% vs 18.7%; RR, 0.32; 95% CI, 0.19-0.53; P < .001) with no significant difference in serious adverse events between the groups (4.2% vs 2.8%; RR, 1.52; 95% CI, 0.81-2.85; P = .19). Conclusions and Relevance: In this meta-analysis of randomized clinical trials including first-line therapy of patients with paroxysmal AF, catheter ablation compared with antiarrhythmic drugs was associated with reductions in recurrence of atrial arrhythmias and hospitalizations, with no difference in major adverse events.


Assuntos
Antiarrítmicos/uso terapêutico , Fibrilação Atrial/terapia , Ablação por Cateter , Hospitalização , Humanos , Ensaios Clínicos Controlados Aleatórios como Assunto
3.
Pacing Clin Electrophysiol ; 44(4): 711-719, 2021 04.
Artigo em Inglês | MEDLINE | ID: mdl-33686695

RESUMO

BACKGROUND: Same-day discharge (SDD) after atrial fibrillation (AF) ablation is increasingly being considered. This study examined the barriers and financial impact associated with SDD in a contemporary cohort of patients undergoing elective AF ablation. METHODS: A single center retrospective review was conducted of the 249 first case-of-the-day outpatient AF ablations performed in 2019 to evaluate the proportion of patients that could have undergone SDD. Barriers to SDD were defined as any intervention that prevented SDD by 8 p.m. The financial impact of SDD was based on savings from avoidance of the overnight hospital stay and revenue related to management of chest pain facilitated by a vacant hospital bed. RESULTS: SDD could have occurred in 157 patients (63%) without change in management and in up to 200 patients (80%) if avoidable barriers were addressed. Barriers to SDD included non-clinical logistical issues (43%), prolonged post-procedure recovery (42%) and minor procedural complications (15%). On multivariate analysis, factors associated with barriers to SDD included increasing age (P = .01), left ventricular ejection fraction ≤ 35% (P = .04), and severely dilated left atrium (P = .04). The financial gain from SDD would have ranged from $1,110,096 (assuming discharge of 63% of eligible patients) to $1,480,128 (assuming 80% discharge) over the course of a year. CONCLUSIONS: Up to 80% of patients undergoing outpatient AF ablation were amenable to SDD if avoidable delays in care had been anticipated. Based on reduced hospital operating expenses and increased revenue from management of individuals with chest pain, this would translate to a financial savings of ∼$1.5 million.


Assuntos
Assistência Ambulatorial/economia , Fibrilação Atrial/cirurgia , Ablação por Cateter/métodos , Tempo de Internação/economia , Alta do Paciente/economia , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Seleção de Pacientes , Estudos Retrospectivos
4.
Circ Arrhythm Electrophysiol ; 13(6): e008716, 2020 06.
Artigo em Inglês | MEDLINE | ID: mdl-32370542

RESUMO

BACKGROUND: A novel ablation and mapping system can toggle between delivering biphasic pulsed field (PF) and radiofrequency energy from a 9-mm lattice-tip catheter. We assessed the preclinical feasibility and safety of (1) focal PF-based thoracic vein isolation and linear ablation, (2) combined PF and radiofrequency focal ablation, and (3) PF delivered directly atop the esophagus. METHODS: Two cohorts of 6 swine were treated with pulsed fields at low dose (PFLD) and high dose (PFHD) and followed for 4 and 2 weeks, respectively, to isolate 25 thoracic veins and create 5 right atrial (PFLD), 6 mitral (PFHD), and 6 roof lines (radiofrequency+PFHD). Baseline and follow-up voltage mapping, venous potentials, ostial diameters, and phrenic nerve viability were assessed. PFHD and radiofrequency lesions were delivered in 4 and 1 swine from the inferior vena cava onto a forcefully deviated esophagus. All tissues were submitted for histopathology. RESULTS: Hundred percent of thoracic veins (25 of 25) were successfully isolated with 12.4±3.6 applications/vein with mean PF times of <90 seconds/vein. Durable isolation improved from 61.5% PFLD to 100% with PFHD (P=0.04), and all linear lesions were successfully completed without incurring venous stenoses or phrenic injury. PFHD sections had higher transmurality rates than PFLD (98.3% versus 88.1%; P=0.03) despite greater mean thickness (2.5 versus 1.3 mm; P<0.001). PF lesions demonstrated homogenous fibrosis without epicardial fat, nerve, or vessel involvement. In comparison, radiofrequency+PFHD sections revealed similar transmurality but expectedly more necrosis, inflammation, and epicardial fat, nerve, and vessel involvement. Significant ablation-related esophageal necrosis, inflammation, and fibrosis were seen in all radiofrequency sections, as compared with no PF sections. CONCLUSIONS: The lattice-tip catheter can deliver focal PF to durably isolate veins and create linear lesions with excellent transmurality and without complications. The PF lesions did not damage the phrenic nerve, vessels, and the esophagus.


Assuntos
Ablação por Cateter/métodos , Átrios do Coração/cirurgia , Veias Pulmonares/cirurgia , Irrigação Terapêutica , Potenciais de Ação , Animais , Cateteres Cardíacos , Ablação por Cateter/efeitos adversos , Ablação por Cateter/instrumentação , Técnicas Eletrofisiológicas Cardíacas , Estudos de Viabilidade , Feminino , Átrios do Coração/patologia , Átrios do Coração/fisiopatologia , Frequência Cardíaca , Modelos Animais , Complicações Pós-Operatórias/patologia , Complicações Pós-Operatórias/fisiopatologia , Veias Pulmonares/patologia , Veias Pulmonares/fisiopatologia , Sus scrofa , Irrigação Terapêutica/efeitos adversos , Irrigação Terapêutica/instrumentação
5.
JACC Clin Electrophysiol ; 3(10): 1092-1102, 2017 10.
Artigo em Inglês | MEDLINE | ID: mdl-29759491

RESUMO

OBJECTIVES: This study sought to compare a novel lesion dimension estimation approach to actual measurements of lesion dimensions on necropsy in porcine atria and ventricles. BACKGROUND: An irrigated-tip, force-sensing radiofrequency catheter with 6 temperature (tip-tissue interface) sensors allows for assessment of lesion dimensions based on estimated tissue temperature. Lesion dimension assessment has not been attempted previously in atrial tissue. METHODS: Ablations were performed using this catheter in all chambers. Irrigated radiofrequency was delivered using 20 to 50 W for durations that ranged from 15 to 90 s with contact force ranging from 5 to 45 g to replicate a wide spectrum of clinical conditions. All swine were then sacrificed and lesions were identified and photographed. Three independent observers made offline measurements, which were then averaged to obtain lesion width and depth for comparison with estimated dimensions based on interface tissue temperature. RESULTS: In 9 swine, 54 atrial and 61 ventricular lesions were assessed. In the atria, the mean difference between the measured and estimated depth and width was 0.9 ± 0.74 mm and 1.2 ± 0.9 mm, respectively. Eighty percent of all lesions had a difference of ≤1.7 mm for depth and ≤1.74 mm for width. In the ventricle, the mean difference between the measured and estimated depth and width was 0.75 ± 0.6 mm and 1.66 ± 1.1 mm, respectively. Eighty percent of all lesions had a difference of ≤1.1 mm ventricular depth and ≤2.6 mm for width. CONCLUSIONS: Estimation of lesion dimensions can be achieved with clinically relevant accuracy using unique temperature signatures. These data have important implications for understanding the adequacy of lesion overlap and assessment of transmurality.


Assuntos
Temperatura Corporal , Ablação por Cateter/métodos , Átrios do Coração/cirurgia , Ventrículos do Coração/cirurgia , Irrigação Terapêutica , Animais , Ablação por Cateter/instrumentação , Desenho de Equipamento , Átrios do Coração/patologia , Ventrículos do Coração/patologia , Microeletrodos , Modelos Animais , Suínos
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA