Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros

Bases de dados
Ano de publicação
Tipo de documento
Intervalo de ano de publicação
1.
Phys Med ; 120: 103337, 2024 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-38552274

RESUMO

The metrics used for assessing image quality in computed tomography (CT) do not integrate the influence of temporal resolution. A shortcoming in the assessment of image quality for imaging protocols where motion blur can therefore occur. We developed a method to calculate the temporal resolution of standard CT protocols and introduced a specific spatiotemporal formulation of the non-prewhitening with eye filter (NPWE) model observer to assess the detectability of moving objects as a function of their speed. We scanned a cubic water phantom with a plexiglass cylindrical insert (120 HU) using a large panel of acquisition parameters (rotation times, pitch factors and collimation widths) on two systems (GE Revolution Apex and Siemens SOMATOM Force) to determine the in-plane task-based transfer functions (TTF) and noise power spectra (NPS). The phantom set in a uniform rectilinear motion in the transverse plane allowed the temporal modulation transfer function (MTF) calculation. The temporal MTF appropriately compared the temporal resolution of the various acquisition protocols. The longitudinal TTF was measured using a thin tungsten wire. The detectability index showed the advantage of applying high rotation speed, wide collimations and high pitch for object detection in the presence of motion. No counterpart to the increase in these three parameters was found in the in-plane and longitudinal image quality.


Assuntos
Interpretação de Imagem Radiográfica Assistida por Computador , Tomografia Computadorizada por Raios X , Doses de Radiação , Tomografia Computadorizada por Raios X/métodos , Imagens de Fantasmas , Algoritmos
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA