1.
J Electrocardiol
; 67: 73-76, 2021.
Artigo
em Inglês
| MEDLINE
| ID: mdl-34090117
RESUMO
Subcutaneous implantable cardioverter-defibrillators are a modern alternative to classic cardioverter-defibrillators. Prior to implantation of the device, qualification based on assessment of the heart's specific rhythm shape during a resting electrocardiogram examination must be performed. In the presented case, a patient with a subcutaneous implantable cardioverter-defibrillator experienced numerous discharges during exercise. An exercise test was performed, which revealed a bundle branch block that appeared during exercise. The wide QRS complex was double- or triple-counted by the device, resulting in an inaccurate heart rate estimation and multiple discharges. Optimising the sensing vectors during exercise solved this problem.