Highly sensitive determination of estrone and estradiol in human serum by liquid chromatography-electrospray ionization tandem mass spectrometry.
Steroids
; 72(11-12): 819-27, 2007 Oct.
Article
em En
| MEDLINE
| ID: mdl-17716700
A highly sensitive and specific quantification method of estrone and estradiol in human serum was described based upon the use of picolinoyl derivatization and liquid chromatography-electrospray ionization mass spectrometry (LC-ESI-MS) in a positive mode. Estrogens were treated with picolinoyl chloride hydrochloride or picolinic acid and 2-methyl-6-nitrobenzoic anhydride followed by a solid-phase extraction with ODS cartridge. Picolinoyl derivatization proceeded quantitatively even in a microscale, and the picolinoyl esters provided simple positive ESI-mass spectra showing [M+H](+) as base peaks for these estrogens. The picolinoyl derivatives of these estrogens showed 100-fold higher detection response compared to underivatized intact molecules by LC-ESI-MS (selected reaction monitoring). Using this derivatization, estrogens spiked in the charcoal treated human serum samples were analyzed with limit of quantification (LOQ), intra-day accuracy and precision of 1.0pg/ml, 96.0% and 9.9% for estrone, and 0.5pg/ml, 84.4% and 12.8% for estradiol, respectively. Estrone and estradiol added to the crude serum samples were recovered with comparable LOQ and accuracy obtained for the charcoal treated serum samples as well.
Buscar no Google
Temas:
ECOS
/
Estado_mercado_regulacao
Bases de dados:
MEDLINE
Assunto principal:
Espectrometria de Massas por Ionização por Electrospray
/
Estradiol
/
Estrona
/
Espectrometria de Massas em Tandem
Tipo de estudo:
Diagnostic_studies
Limite:
Female
/
Humans
Idioma:
En
Revista:
Steroids
Ano de publicação:
2007
Tipo de documento:
Article
País de afiliação:
Japão