ABSTRACT
A rapid antigen assay for malaria was performed on blood samples collected during a simultaneous outbreak of falciparum malaria and vivax malaria on a remote island in the Indonesian archipelago. During the outbreak, a total of 89 patients (4.3% of the population) were diagnosed with malaria within a week. Microscopic examination revealed 78 malaria slide-positive cases, of whom 49 (62.8%) were identified as P. falciparum, 7 (9.0%) as P. vivax and 22 (28.2%) as mixed P. falciparum and P. vivax infections. The rapid malaria assay showed excellent correlation with expert-confirmed routine microscopy for P. falciparum and P. vivax monoinfections and mixed infections with a parasite density >50 parasites/microl. Several slide-negative blood samples collected from febrile patients with clinical malaria tested positive in the rapid test. The estimated sensitivity calculated for the rapid test (91.0%) was slightly higher than that of microscopy (87.6%). The result indicates that rapid antigen detection for malaria could be a useful alternative to microscopy to reduce the workload during emergency outbreak situations.
Texto completo:
1
Temas:
ECOS
/
Aspectos_gerais
Bases de dados:
MEDLINE
Assunto principal:
Kit de Reagentes para Diagnóstico
/
Malária Vivax
/
Malária Falciparum
/
Sistemas Automatizados de Assistência Junto ao Leito
/
Antígenos de Protozoários
Tipo de estudo:
Diagnostic_studies
/
Prognostic_studies
Limite:
Adolescent
/
Adult
/
Animals
/
Child
/
Female
/
Humans
/
Male
/
Middle aged
Idioma:
En
Revista:
Trans R Soc Trop Med Hyg
Ano de publicação:
2008
Tipo de documento:
Article
País de afiliação:
Indonésia