Influence of secondary neutrons induced by proton radiotherapy for cancer patients with implantable cardioverter defibrillators.
Hashimoto, Takayuki; Isobe, Tomonori; Hashii, Haruko; Kumada, Hiroaki; Tada, Hiroshi; Okumura, Toshiyuki; Tsuboi, Koji; Sakae, Takeji; Aonuma, Kazutaka; Sakurai, Hideyuki.
Afiliação
  • Hashimoto T; Department of Radiation Oncology, Graduate School of Comprehensive Human Sciences, University of Tsukuba, 1-1-1 Tennodai, Tsukuba, Ibaraki 305-8575, Japan. hashimoto@pmrc.tsukuba.ac.jp
Radiat Oncol ; 7: 10, 2012 Jan 29.
Article em En | MEDLINE | ID: mdl-22284700
BACKGROUND: Although proton radiotherapy is a promising new approach for cancer patients, functional interference is a concern for patients with implantable cardioverter defibrillators (ICDs). The purpose of this study was to clarify the influence of secondary neutrons induced by proton radiotherapy on ICDs. METHODS: The experimental set-up simulated proton radiotherapy for a patient with an ICD. Four new ICDs were placed 0.3 cm laterally and 3 cm distally outside the radiation field in order to evaluate the influence of secondary neutrons. The cumulative in-field radiation dose was 107 Gy over 10 sessions of irradiation with a dose rate of 2 Gy/min and a field size of 10 × 10 cm². After each radiation fraction, interference with the ICD by the therapy was analyzed by an ICD programmer. The dose distributions of secondary neutrons were estimated by Monte-Carlo simulation. RESULTS: The frequency of the power-on reset, the most serious soft error where the programmed pacing mode changes temporarily to a safety back-up mode, was 1 per approximately 50 Gy. The total number of soft errors logged in all devices was 29, which was a rate of 1 soft error per approximately 15 Gy. No permanent device malfunctions were detected. The calculated dose of secondary neutrons per 1 Gy proton dose in the phantom was approximately 1.3-8.9 mSv/Gy. CONCLUSIONS: With the present experimental settings, the probability was approximately 1 power-on reset per 50 Gy, which was below the dose level (60-80 Gy) generally used in proton radiotherapy. Further quantitative analysis in various settings is needed to establish guidelines regarding proton radiotherapy for cancer patients with ICDs.
Assuntos

Texto completo: 1 Temas: ECOS / Financiamentos_gastos Bases de dados: MEDLINE Assunto principal: Prótons / Desfibriladores Implantáveis / Análise de Falha de Equipamento / Neoplasias / Nêutrons Tipo de estudo: Guideline / Health_economic_evaluation Limite: Humans Idioma: En Revista: Radiat Oncol Assunto da revista: NEOPLASIAS / RADIOTERAPIA Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Japão

Texto completo: 1 Temas: ECOS / Financiamentos_gastos Bases de dados: MEDLINE Assunto principal: Prótons / Desfibriladores Implantáveis / Análise de Falha de Equipamento / Neoplasias / Nêutrons Tipo de estudo: Guideline / Health_economic_evaluation Limite: Humans Idioma: En Revista: Radiat Oncol Assunto da revista: NEOPLASIAS / RADIOTERAPIA Ano de publicação: 2012 Tipo de documento: Article País de afiliação: Japão