Assessment of radiation exposure in dental cone-beam computerized tomography with the use of metal-oxide semiconductor field-effect transistor (MOSFET) dosimeters and Monte Carlo simulations.
Oral Surg Oral Med Oral Pathol Oral Radiol
; 114(3): 393-400, 2012 Sep.
Article
em En
| MEDLINE
| ID: mdl-22862982
OBJECTIVES: The aims of this study were to assess the organ and effective dose (International Commission on Radiological Protection (ICRP) 103) resulting from dental cone-beam computerized tomography (CBCT) imaging using a novel metal-oxide semiconductor field-effect transistor (MOSFET) dosimeter device, and to assess the reliability of the MOSFET measurements by comparing the results with Monte Carlo PCXMC simulations. STUDY DESIGN: Organ dose measurements were performed using 20 MOSFET dosimeters that were embedded in the 8 most radiosensitive organs in the maxillofacial and neck area. The dose-area product (DAP) values attained from CBCT scans were used for PCXMC simulations. The acquired MOSFET doses were then compared with the Monte Carlo simulations. RESULTS: The effective dose measurements using MOSFET dosimeters yielded, using 0.5-cm steps, a value of 153 µSv and the PCXMC simulations resulted in a value of 136 µSv. CONCLUSIONS: The MOSFET dosimeters placed in a head phantom gave results similar to Monte Carlo simulations. Minor vertical changes in the positioning of the phantom had a substantial affect on the overall effective dose. Therefore, the MOSFET dosimeters constitute a feasible method for dose assessment of CBCT units in the maxillofacial region.
Texto completo:
1
Temas:
ECOS
/
Aspectos_gerais
/
Financiamentos_gastos
Bases de dados:
MEDLINE
Assunto principal:
Doses de Radiação
/
Radiometria
/
Transistores Eletrônicos
/
Radiografia Dentária
/
Tomografia Computadorizada de Feixe Cônico
/
Cabeça
Tipo de estudo:
Health_economic_evaluation
Limite:
Humans
/
Male
Idioma:
En
Revista:
Oral Surg Oral Med Oral Pathol Oral Radiol
Ano de publicação:
2012
Tipo de documento:
Article
País de afiliação: