A method for optimizing EEG electrode number and configuration for signal acquisition in P300 speller systems.
Speier, William; Deshpande, Aniket; Pouratian, Nader.
Afiliação
  • Speier W; University of California, Los Angeles, Department of Bioengineering, David Geffen School of Medicine at UCLA, Los Angeles, CA, USA; University of California, Los Angeles, Medical Imaging Informatics Group, David Geffen School of Medicine at UCLA, Los Angeles, CA, USA.
  • Deshpande A; University of California, Los Angeles, Department of Bioengineering, David Geffen School of Medicine at UCLA, Los Angeles, CA, USA.
  • Pouratian N; University of California, Los Angeles, Department of Bioengineering, David Geffen School of Medicine at UCLA, Los Angeles, CA, USA; University of California, Los Angeles, Department of Neurosurgery, David Geffen School of Medicine at UCLA, Los Angeles, CA, USA; University of California, Los Angeles, Interdepartmental Program in Neuroscience, David Geffen School of Medicine at UCLA, Los Angeles, CA, USA; University of California, Los Angeles, Brain Research Institute, David Geffen School of Medic
Clin Neurophysiol ; 126(6): 1171-1177, 2015 Jun.
Article em En | MEDLINE | ID: mdl-25316166
ABSTRACT

OBJECTIVE:

The P300 speller is intended to restore communication to patients with advanced neuromuscular disorders, but clinical implementation may be hindered by several factors, including system setup, burden, and cost. Our goal was to develop a method that can overcome these barriers by optimizing EEG electrode number and placement for P300 studies within a population of subjects.

METHODS:

A Gibbs sampling method was developed to find the optimal electrode configuration given a set of P300 speller data. The method was tested on a set of data from 15 healthy subjects using an established 32-electrode pattern. Resulting electrode configurations were then validated using online prospective testing with a naïve Bayes classifier in 15 additional healthy subjects.

RESULTS:

The method yielded a set of four posterior electrodes (PO8, PO7, POZ, CPZ), which produced results that are likely sufficient to be clinically effective. In online prospective validation testing, no significant difference was found between subjects' performances using the reduced and the full electrode configurations.

CONCLUSIONS:

The proposed method can find reduced sets of electrodes within a subject population without reducing performance.

SIGNIFICANCE:

Reducing the number of channels may reduce costs, set-up time, signal bandwidth, and computation requirements for practical online P300 speller implementation.
Assuntos
Palavras-chave

Texto completo: 1 Temas: ECOS / Aspectos_gerais Bases de dados: MEDLINE Assunto principal: Potenciais Evocados P300 / Eletroencefalografia / Interfaces Cérebro-Computador Tipo de estudo: Observational_studies Limite: Adult / Female / Humans / Male Idioma: En Revista: Clin Neurophysiol Assunto da revista: NEUROLOGIA / PSICOFISIOLOGIA Ano de publicação: 2015 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Temas: ECOS / Aspectos_gerais Bases de dados: MEDLINE Assunto principal: Potenciais Evocados P300 / Eletroencefalografia / Interfaces Cérebro-Computador Tipo de estudo: Observational_studies Limite: Adult / Female / Humans / Male Idioma: En Revista: Clin Neurophysiol Assunto da revista: NEUROLOGIA / PSICOFISIOLOGIA Ano de publicação: 2015 Tipo de documento: Article País de afiliação: Estados Unidos