Multicenter study to evaluate a new enumeration method for hematopoietic stem cell collection management.
Grommé, Monique; Russcher, Henk; Braakman, Eric; Klinkspoor, J Henriëtte; Dobber, Johan A; de Greef, Inge; de Wit, Norbert C J.
Afiliação
  • Grommé M; Department of Internal Medicine, Division of Haematology, Erasmus Medical Centre, Rotterdam, the Netherlands.
  • Russcher H; Clinical Chemistry, Erasmus Medical Centre, Rotterdam, the Netherlands.
  • Braakman E; Department of Hematology, Erasmus Medical Centre, Rotterdam, the Netherlands.
  • Klinkspoor JH; Laboratory for General Clinical Chemistry, Academic Medical Center, Amsterdam, the Netherlands.
  • Dobber JA; Laboratory for Specialized Haematology, Academic Medical Center, Amsterdam, the Netherlands.
  • de Greef I; Department of Hematology, Erasmus Medical Centre, Rotterdam, the Netherlands.
  • de Wit NCJ; Central Diagnostic Laboratory, Maastricht University Medical Center, Maastricht, the Netherlands.
Transfusion ; 57(8): 1949-1955, 2017 08.
Article em En | MEDLINE | ID: mdl-28653370
ABSTRACT

BACKGROUND:

CD34 flow cytometry is the gold standard for stem cell enumeration in peripheral blood at the mobilization stage and in the final apheresis product. The new stem cell mode of the Sysmex XN Series analyzer enumerates an immature cell population in the white progenitor and pathological cell (WPC) channel, based on the cell size, internal cellular complexity, and fluorescence intensity. STUDY DESIGN AND

METHODS:

In this multicenter study we analyzed 147 peripheral blood samples, 22 samples during collection of stem cells, and 45 samples from the apheresis product of 18 healthy allogeneic donors and 84 autologous patients.

RESULTS:

In this multicenter study we demonstrate that the XN stem cell enumeration method correlates well with viable CD34+ cells determined by flow cytometry during the stem cell mobilization phase to determine apheresis start time, during apheresis for real-time monitoring and adjustment, and for quality control of the final stem cell harvest.

CONCLUSION:

Our data show that there is an improvement in the correlation of XN stem cells and CD34+ cells in the peripheral blood during stem cell mobilization as well as in stem cell harvests compared to SE or XE Series analyzers. The XN stem cell enumeration method has a number of advantages compared to CD34 flow cytometry it is fast, simple, reproducible, and less expensive. CE marking for the European market has been obtained, making the stem cell count on the XN analyzer a reportable clinical variable.
Assuntos

Texto completo: 1 Temas: ECOS / Aspectos_gerais / Financiamentos_gastos Bases de dados: MEDLINE Assunto principal: Contagem de Células Sanguíneas / Células-Tronco Hematopoéticas Tipo de estudo: Clinical_trials / Evaluation_studies / Health_economic_evaluation Limite: Humans Idioma: En Revista: Transfusion Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Holanda

Texto completo: 1 Temas: ECOS / Aspectos_gerais / Financiamentos_gastos Bases de dados: MEDLINE Assunto principal: Contagem de Células Sanguíneas / Células-Tronco Hematopoéticas Tipo de estudo: Clinical_trials / Evaluation_studies / Health_economic_evaluation Limite: Humans Idioma: En Revista: Transfusion Ano de publicação: 2017 Tipo de documento: Article País de afiliação: Holanda