Your browser doesn't support javascript.
loading
Monte Carlo simulation of secondary neutron dose for scanning proton therapy using FLUKA.
Lee, Chaeyeong; Lee, Sangmin; Lee, Seung-Jae; Song, Hankyeol; Kim, Dae-Hyun; Cho, Sungkoo; Jo, Kwanghyun; Han, Youngyih; Chung, Yong Hyun; Kim, Jin Sung.
Afiliação
  • Lee C; Department of Radiological Science, Yonsei University, Wonju, Korea.
  • Lee S; Program in Biomedical Radiation Sciences, Department of Transdisciplinary Studies, Graduate School of Convergence Science and Technology, Seoul National University, Seoul, Republic of Korea.
  • Lee SJ; Department of Radiological Science, Yonsei University, Wonju, Korea.
  • Song H; Department of Radiological Science, Yonsei University, Wonju, Korea.
  • Kim DH; Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
  • Cho S; Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
  • Jo K; Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
  • Han Y; Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
  • Chung YH; Department of Radiological Science, Yonsei University, Wonju, Korea.
  • Kim JS; Department of Radiation Oncology, Samsung Medical Center, Sungkyunkwan University School of Medicine, Seoul, Republic of Korea.
PLoS One ; 12(10): e0186544, 2017.
Article em En | MEDLINE | ID: mdl-29045491
ABSTRACT
Proton therapy is a rapidly progressing field for cancer treatment. Globally, many proton therapy facilities are being commissioned or under construction. Secondary neutrons are an important issue during the commissioning process of a proton therapy facility. The purpose of this study is to model and validate scanning nozzles of proton therapy at Samsung Medical Center (SMC) by Monte Carlo simulation for beam commissioning. After the commissioning, a secondary neutron ambient dose from proton scanning nozzle (Gantry 1) was simulated and measured. This simulation was performed to evaluate beam properties such as percent depth dose curve, Bragg peak, and distal fall-off, so that they could be verified with measured data. Using the validated beam nozzle, the secondary neutron ambient dose was simulated and then compared with the measured ambient dose from Gantry 1. We calculated secondary neutron dose at several different points. We demonstrated the validity modeling a proton scanning nozzle system to evaluate various parameters using FLUKA. The measured secondary neutron ambient dose showed a similar tendency with the simulation result. This work will increase the knowledge necessary for the development of radiation safety technology in medical particle accelerators.
Assuntos

Texto completo: 1 Temas: ECOS / Financiamentos_gastos Bases de dados: MEDLINE Assunto principal: Dosagem Radioterapêutica / Simulação por Computador / Método de Monte Carlo / Terapia com Prótons / Nêutrons Tipo de estudo: Health_economic_evaluation / Prognostic_studies Idioma: En Revista: PLoS One Assunto da revista: CIENCIA / MEDICINA Ano de publicação: 2017 Tipo de documento: Article

Texto completo: 1 Temas: ECOS / Financiamentos_gastos Bases de dados: MEDLINE Assunto principal: Dosagem Radioterapêutica / Simulação por Computador / Método de Monte Carlo / Terapia com Prótons / Nêutrons Tipo de estudo: Health_economic_evaluation / Prognostic_studies Idioma: En Revista: PLoS One Assunto da revista: CIENCIA / MEDICINA Ano de publicação: 2017 Tipo de documento: Article