Clinical assessment of aortic valve stenosis: Comparison between 4D flow MRI and transthoracic echocardiography.
J Magn Reson Imaging
; 51(2): 472-480, 2020 02.
Article
em En
| MEDLINE
| ID: mdl-31257647
ABSTRACT
BACKGROUND:
The prevalence of valvular aortic stenosis (AS) increases as the population ages. Echocardiographic measurements of peak jet velocity (Vpeak ), mean pressure gradient (Pmean ), and aortic valve area (AVA) determine AS severity and play a pivotal role in the stratification towards valvular replacement. A multimodality imaging approach might be needed in cases of uncertainty about the actual severity of the stenosis.PURPOSE:
To compare four-dimensional phase-contrast magnetic resonance (4D PC-MR), two-dimensional (2D) PC-MR, and transthoracic echocardiography (TTE) for quantification of AS. STUDY TYPE Prospective. POPULATION Twenty patients with various degrees of AS (69.3 ± 5.0 years). FIELD STRENGTH/SEQUENCES 4D PC-MR and 2D PC-MR at 3T. ASSESSMENT We compared Vpeak , Pmean , and AVA between TTE, 4D PC-MR, and 2D PC-MR. Flow eccentricity was quantified by means of normalized flow displacement, and its influence on the accuracy of TTE measurements was investigated. STATISTICAL TESTS Pearson's correlation, Bland-Altman analysis, paired t-test, and intraclass correlation coefficient.RESULTS:
4D PC-MR measured higher Vpeak (r = 0.95, mean difference + 16.4 ± 10.7%, P <0.001), and Pmean (r = 0.92, mean difference + 14.9 ± 16.0%, P = 0.013), but a less critical AVA (r = 0.80, mean difference + 19.9 ± 20.6%, P = 0.002) than TTE. In contrast, unidirectional 2D PC-MR substantially underestimated AS severity when compared with TTE. Differences in Vpeak between 4D PC-MR and TTE showed to be strongly correlated with the eccentricity of the flow jet (r = 0.89, P <0.001). Use of 4D PC-MR improved the concordance between Vpeak and AVA (from 0.68 to 0.87), and between PGmean and AVA (from 0.68 to 0.86). DATACONCLUSION:
4D PC-MR improves the concordance between the different AS parameters and could serve as an additional imaging technique next to TTE. Future studies should address the potential value of 4D PC-MR in patients with discordant echocardiographic parameters. LEVEL OF EVIDENCE 2 Technical Efficacy Stage 2 J. Magn. Reson. Imaging 2020;51472-480.Palavras-chave
Texto completo:
1
Temas:
ECOS
/
Aspectos_gerais
Bases de dados:
MEDLINE
Assunto principal:
Valva Aórtica
/
Estenose da Valva Aórtica
Tipo de estudo:
Observational_studies
/
Risk_factors_studies
Limite:
Humans
Idioma:
En
Revista:
J Magn Reson Imaging
Assunto da revista:
DIAGNOSTICO POR IMAGEM
Ano de publicação:
2020
Tipo de documento:
Article
País de afiliação:
Holanda