Your browser doesn't support javascript.
loading
The digital scent device as a new concept for olfactory assessment.
Nakanishi, Marcio; Fornazieri, Marco Aurélio; Lança Gomes, Pedro; Dias, Luis Augusto de Miranda; Freire, Gustavo Subtil Magalhães; Vinha, Luís Gustavo do Amaral; Barbosa de Sa, Leonardo Conrado; Voegels, Richard Louis; Galvão, Claudia; Lima, Wilma Terezinha Anselmo.
Afiliação
  • Nakanishi M; Department of Otorhinolaryngology, University Hospital of Brasília UnB-EBSERH, Brasilia, Federal District, Brazil.
  • Fornazieri MA; Department of Ophthalmology and Otorhinolaryngology, Ribeirão Preto Medical School-USP, Ribeirão Preto, São Paulo, Brazil.
  • Lança Gomes P; Division of Otorhinolaryngology, D'Or Institute of Education and Research, Hospital Santa Luzia, Brasilia, Brazil.
  • Dias LAM; Department of Otorhinolaryngology State University of Londrina, Pontifical Catholic University of Paraná, Londrina, PR, Brazil.
  • Freire GSM; Division of Otorhinolaryngology, D'Or Institute of Education and Research, Hospital Santa Luzia, Brasilia, Brazil.
  • Vinha LGDA; Division of Neurosurgery, Hospital de Base of Distrito Federal, Brasilia, DF, Brazil.
  • Barbosa de Sa LC; Department of Otorhinolaryngology, University Hospital of Brasília UnB-EBSERH, Brasilia, Federal District, Brazil.
  • Voegels RL; Department of Statistics, University of Brasilia, Brasília, DF, Brazil.
  • Galvão C; Division of Otorhinolaryngology, D'Or Institute of Education and Research, Hospital Santa Luzia, Brasilia, Brazil.
  • Lima WTA; Department of Ophthalmology and Otorhinolaryngology, University of São Paulo, São Paulo, Brazil.
Int Forum Allergy Rhinol ; 12(10): 1263-1272, 2022 10.
Article em En | MEDLINE | ID: mdl-35224885
ABSTRACT

BACKGROUND:

There are major challenges in olfactory measurements in clinical practice; therefore, a handheld digital scent device (DSD; Noar MultiScent 20) was developed as a tablet with an integrated storage system for odors. The DSD is a self-administered, handheld device that controls the duration of odor release to the nasal cavity through a touchscreen digital interface with automatic database generation. In this study we aimed to determine the feasibility of this DSD as an olfactory assessment test.

METHODS:

We recruited 180 participants (age [mean ± standard deviation], 34.58 ± 9.71 years; 114 women and 66 men) to participate in smell tests using both the DSD and the 40-item Smell Identification Test (SIT-40), which contained the same type and order of odors and the same multiple-choice answers. The scores were compared and evaluated for correlation between the tests, and test-retest reliability was calculated.

RESULTS:

The DSD test scores were higher than the SIT-40 scores (median [interquartile range], 32 [5.0] vs 31 [7.0]; p = 0.005). The completion time was less for the DSD test than for the SIT-40 (12.5 [5.0] vs 16 [6.0] minutes; p < 0.001). The tests were strongly correlated (Spearman rho = 0.74; p < 0.001) and exhibited a high level of agreement (Bland-Altman regression coefficient = 0.672; p = 0.003). The DSD test-retest reliability coefficient was 0.820.

CONCLUSION:

The DSD is feasible as an olfactory assessment test. The digitalization of olfactory assessment combined with data science may enable new research perspectives in the field of olfaction.
Assuntos
Palavras-chave

Texto completo: 1 Temas: ECOS / Aspectos_gerais Bases de dados: MEDLINE Assunto principal: Olfato / Transtornos do Olfato Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Adult / Female / Humans / Male Idioma: En Revista: Int Forum Allergy Rhinol Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Brasil

Texto completo: 1 Temas: ECOS / Aspectos_gerais Bases de dados: MEDLINE Assunto principal: Olfato / Transtornos do Olfato Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Adult / Female / Humans / Male Idioma: En Revista: Int Forum Allergy Rhinol Ano de publicação: 2022 Tipo de documento: Article País de afiliação: Brasil