Your browser doesn't support javascript.
loading
A practical method of estimating medium-heterogeneity corrected dose without a Monte Carlo calculation in ocular brachytherapy using 125I COMS plaques.
Lee, Yongsook C; Nik Akhtar, Mehran; Kim, Yongbok; Jung, Jae Won.
Afiliação
  • Lee YC; Department of Radiation Oncology, Miami Cancer Institute, Baptist Health South Florida, Miami, FL; Department of Radiation Oncology, Herbert Wertheim College of Medicine, Florida International University, Miami, FL.
  • Nik Akhtar M; Department of Physics, East Carolina University, Greenville, NC.
  • Kim Y; Department of Radiation Oncology, Duke University, Durham, NC. Electronic address: yongbok.kim@duke.edu.
  • Jung JW; Department of Radiation Oncology, East Carolina University, Greenville, NC.
Brachytherapy ; 23(3): 377-386, 2024.
Article em En | MEDLINE | ID: mdl-38336557
ABSTRACT

PURPOSE:

To provide a practical method of estimating medium-heterogeneity corrected dose without a Monte Carlo (MC) calculation in ocular brachytherapy using 125I Collaborative Ocular Melanoma Study (COMS) plaques. METHODS AND MATERIALS Using egs_brachy, MC simulations (1) under task group-43 assumptions with fully loaded seed configurations in water (HOMO) and (2) with effects of plaque backing, insert and inter-seed interactions (HETERO) were performed for seven 125I COMS plaques (10 mm-22 mm in diameter), and homogeneous dose (DHOMO) and heterogeneous dose (DHETERO) for central-axis and off-axis points were determined. For DHOMO, 85 Gy was normalized to a depth of 5 mm. Central-axis heterogeneity correction factors (HCFs) from a depth of 0 mm (inner sclera) to 22 mm (opposite retina) were derived from a ratio of DHETERO to DHOMO. Off-axis HCFs for optic disc/macula and lens as a function of distance from optic disc/macula (DT/MT) for various basal dimensions of tumor (BD/BM) were derived from DHETERO/DHOMO.

RESULTS:

Central-axis HCF varied with a dose reduction of 10.3-19.8% by heterogeneity. Off-axis HCF for optic disc/macula varied significantly depending on DT/MT and BD/BM with a dose reduction of 11.3-38.3%. Off-axis HCF for lens had a dependence on MT and BM with its variation of 11.0-19.0%. A clinical example of using HCFs to estimate DHETERO was presented.

CONCLUSIONS:

The practical method of using depth-dependent central-axis HCF and DT/MT- and BD/BM-dependent off-axis HCF provided in this study will facilitate a heterogeneous dose estimate for 125I COMS plaques without MC calculations.
Assuntos
Palavras-chave

Texto completo: 1 Temas: ECOS / Financiamentos_gastos Bases de dados: MEDLINE Assunto principal: Dosagem Radioterapêutica / Braquiterapia / Método de Monte Carlo / Neoplasias Oculares / Radioisótopos do Iodo / Melanoma Limite: Humans Idioma: En Revista: Brachytherapy Assunto da revista: RADIOTERAPIA Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Temas: ECOS / Financiamentos_gastos Bases de dados: MEDLINE Assunto principal: Dosagem Radioterapêutica / Braquiterapia / Método de Monte Carlo / Neoplasias Oculares / Radioisótopos do Iodo / Melanoma Limite: Humans Idioma: En Revista: Brachytherapy Assunto da revista: RADIOTERAPIA Ano de publicação: 2024 Tipo de documento: Article