1.
Bioanalysis ; 16(6): 389-402, 2024 Mar.
Article in English | MEDLINE | ID: mdl-38334082

ABSTRACT

Validation of biomarker assays is crucial for effective drug development and clinical applications. Interlaboratory reproducibility is vital for reliable comparison and combination of data from different centers. This review summarizes interlaboratory studies of quantitative LC-MS-based biomarker assays using reference standards for calibration curves. The following points are discussed: trends in reports, reference and internal standards, evaluation of analytical validation parameters, study sample analysis and normalization of biomarker assay data. Full evaluation of these parameters in interlaboratory studies is limited, necessitating further research. Some reports suggest methods to address variations in biomarker assay data among laboratories, facilitating organized studies and data combination. Method validation across laboratories is crucial for reducing interlaboratory differences and reflecting target biomarker responses.


Subject(s)
Liquid Chromatography-Mass Spectrometry , Tandem Mass Spectrometry , Chromatography, Liquid/methods , Tandem Mass Spectrometry/methods , Reproducibility of Results , Reference Standards