ABSTRACT
PURPOSE: to assess overranging (OR) as a function of pitch, collimation and rotation time for three commonly used CT models in France. METHODS: OR was measured using XR-SP2 Gafchromic films (Ashland, USA) and computed from exposure data for three widely used CT models in France. The impact of collimation, pitch and rotation time on OR was analyzed while comparing the three CT models. RESULTS: for a typical head protocol, measured OR was found to be equal to 1.64 cm for CT scan A, 0.5 cm for CT scan B and 3.44 cm for CT scan C. OR values were respectively of 4.47, 3.24 and 7.81 cm for the typical chest protocol and the 3 CT models. OR was lowest at pitch values <1 for CT scan A, while it linearly increased with pitch for CT scan B and was lowest for high pitch values for CT scan C. Additionally, OR increase with collimation (24 mm to 38.4 mm) was most pronounced for CT scan B (factor of 2.2 for low pitch (0.5) and 2.8 for high pitch (1.2)) while it increased linearly regardless of the pitch for CT scans A and C. Rotation time also proved to slightly affect OR mainly for CT scan B (up to 2.2 mm). CONCLUSION: the present study shows that overranging remains an issue in modern CT scanners with large disparities across the vendors.