Diabetes Care ; 45(7): 1666-1669, 2022 07 07.
Article in English | MEDLINE | ID: mdl-35485908

ABSTRACT

OBJECTIVE: Continuous glucose monitoring (CGM) improves diabetes management, but its reliability in individuals on hemodialysis is poorly understood and potentially affected by interstitial and intravascular volume variations. RESEARCH DESIGN AND METHODS: We assessed the accuracy of a factory-calibrated CGM by using venous blood glucose measurements (vBGM) during hemodialysis sessions and self-monitoring blood glucose (SMBG) at home. RESULTS: Twenty participants completed the protocol. The mean absolute relative difference of the CGM was 13.8% and 14.4%, when calculated on SMBG (n = 684) and on vBGM (n = 624), and 98.7% and 100% of values in the Parkes error grid A/B zones, respectively. Throughout 181 days of CGM monitoring, the median time in range (70-180 mg/dL) was 38.5% (interquartile range 29.3-57.9), with 28.7% (7.8-40.6) of the time >250 mg/dL. CONCLUSIONS: The overall performance of a factory-calibrated CGM appears reasonably accurate and clinically relevant for use in practice by individuals on hemodialysis and health professionals to improve diabetes management.


Subject(s)
Blood Glucose , Diabetes Mellitus, Type 1 , Blood Glucose Self-Monitoring/methods , Humans , Renal Dialysis , Reproducibility of Results