1.
Eur Radiol ; 34(4): 2647-2657, 2024 Apr.
Article in English | MEDLINE | ID: mdl-37672056

ABSTRACT

OBJECTIVES: Evaluation of in-stent restenosis (ISR), especially for small stents, remains challenging during computed tomography (CT) angiography. We used deep learning reconstruction to quantify stent strut thickness and lumen vessel diameter at the stent and compared it with values obtained using conventional reconstruction strategies. METHODS: We examined 166 stents in 85 consecutive patients who underwent CT and invasive coronary angiography (ICA) within 3 months of each other from 2019-2021 after percutaneous coronary intervention with coronary stent placement. The presence of ISR was defined as percent diameter stenosis ≥ 50% on ICA. We compared a super-resolution deep learning reconstruction, Precise IQ Engine (PIQE), and a model-based iterative reconstruction, Forward projected model-based Iterative Reconstruction SoluTion (FIRST). All images were reconstructed using PIQE and FIRST and assessed by two blinded cardiovascular radiographers. RESULTS: PIQE had a larger full width at half maximum of the lumen and smaller strut than FIRST. The image quality score in PIQE was higher than that in FIRST (4.2 ± 1.1 versus 2.7 ± 1.2, p < 0.05). In addition, the specificity and accuracy of ISR detection were better in PIQE than in FIRST (p < 0.05 for both), with particularly pronounced differences for stent diameters < 3.0 mm. CONCLUSION: PIQE provides superior image quality and diagnostic accuracy for ISR, even with stents measuring < 3.0 mm in diameter. CLINICAL RELEVANCE STATEMENT: With improvements in the diagnostic accuracy of in-stent stenosis, CT angiography could become a gatekeeper for ICA in post-stenting cases, obviating ICA in many patients after recent stenting with infrequent ISR and allowing non-invasive ISR detection in the late phase. KEY POINTS: • Despite CT technology advancements, evaluating in-stent stenosis severity, especially in small-diameter stents, remains challenging. • Compared with conventional methods, the Precise IQ Engine uses deep learning to improve spatial resolution. • Improved diagnostic accuracy of CT angiography helps avoid invasive coronary angiography after coronary artery stenting.


Subject(s)
Coronary Restenosis , Deep Learning , Humans , Computed Tomography Angiography/methods , Coronary Angiography/methods , Coronary Restenosis/diagnostic imaging , Feasibility Studies , Constriction, Pathologic , Tomography, X-Ray Computed/methods , Stents
2.
Eur Radiol ; 34(2): 1065-1076, 2024 Feb.
Article in English | MEDLINE | ID: mdl-37580601

ABSTRACT

OBJECTIVE: The purpose of this study was thus to compare capabilities for quantitative differentiation of non- and minimally invasive adenocarcinomas from other of pulmonary MRIs with ultra-short TE (UTE) obtained with single- and dual-echo techniques (UTE-MRISingle and UTE-MRIDual) and thin-section CT for stage IA lung cancer patients. METHODS: Ninety pathologically diagnosed stage IA lung cancer patients who underwent thin-section standard-dose CT, UTE-MRISingle, and UTE-MRIDual, surgical treatment and pathological examinations were included in this retrospective study. The largest dimension (Dlong), solid portion (solid Dlong), and consolidation/tumor (C/T) ratio of each nodule were assessed. Two-tailed Student's t-tests were performed to compare all indexes obtained with each method between non- and minimally invasive adenocarcinomas and other lung cancers. Receiver operating characteristic (ROC)-based positive tests were performed to determine all feasible threshold values for distinguishing non- or minimally invasive adenocarcinoma (MIA) from other lung cancers. Sensitivity, specificity, and accuracy were then compared by means of McNemar's test. RESULTS: Each index showed significant differences between the two groups (p < 0.0001). Specificities and accuracies of solid Dlong for UTE-MRIDual2nd echo and CTMediastinal were significantly higher than those of solid Dlong for UTE-MRISingle and UTE-MRIDual1st echo and all C/T ratios except CTMediastinal (p < 0.05). Moreover, the specificities and accuracies of solid Dlong and C/T ratio were significantly higher than those of Dlong for each method (p < 0.05). CONCLUSION: Pulmonary MRI with UTE is considered at least as valuable as thin-section CT for quantitative differentiation of non- and minimally invasive adenocarcinomas from other stage IA lung cancers. CLINICAL RELEVANCE STATEMENT: Pulmonary MRI with UTE's capability for quantitative differentiation of non- and minimally invasive adenocarcinomas from other lung cancers in stage IA lung cancer patients is equal or superior to that of thin-section CT. KEY POINTS: • Correlations were excellent for pathologically examined nodules with the largest dimensions (Dlong) and a solid component (solid Dlong) for all indexes (0.95 ≤ r ≤ 0.99, p < 0.0001). • Pathologically examined Dlong and solid Dlong obtained with all methods showed significant differences between non- and minimally invasive adenocarcinomas and other lung cancers (p < 0.0001). • Solid tumor components are most accurately measured by UTE-MRIDual2nd echo and CTMediastinal, whereas the ground-glass component is imaged by UTE-MRIDual1st echo and CTlung with high accuracy. UTE-MRIDual predicts tumor invasiveness with 100% sensitivity and 87.5% specificity at a C/T threshold of 0.5.


Subject(s)
Adenocarcinoma , Lung Diseases , Lung Neoplasms , Humans , Lung Neoplasms/pathology , Retrospective Studies , Tomography, X-Ray Computed/methods , Lung/pathology , Adenocarcinoma/pathology , Magnetic Resonance Imaging/methods
3.
Magn Reson Med ; 90(5): 2001-2010, 2023 11.
Article in English | MEDLINE | ID: mdl-37288577

ABSTRACT

PURPOSE: To develop 3D ultrashort-TE (UTE) sequences with tight TE intervals (δTE), allowing for accurate T 2 * $$ {\mathrm{T}}_2^{\ast } $$ mapping of lungs under free breathing. METHODS: We have implemented a four-echo UTE sequence with δTE (< 0.5 ms). A Monte-Carlo simulation was performed to identify an optimal number of echoes that would result in a significant improvement in the accuracy of the T 2 * $$ {\mathrm{T}}_2^{\ast } $$ fit within an acceptable scan time. A validation study was conducted on a phantom with known short T 2 * $$ {\mathrm{T}}_2^{\ast } $$ values (< 5 ms). The scanning protocol included a combination of a standard multi-echo UTE with six echoes (2.2-ms intervals) and a new four-echo UTE (TE < 2 ms) with tight TE intervals δTE. The human imaging was performed at 3 T on 6 adult volunteers. T 2 * $$ {\mathrm{T}}_2^{\ast } $$ mapping was performed with mono-exponential and bi-exponential models. RESULTS: The simulation for the proposed 10-echo acquisition predicted over 2-fold improvement in the accuracy of estimating the short T 2 * $$ {\mathrm{T}}_2^{\ast } $$ compared with the regular six-echo acquisition. In the phantom study, the T 2 * $$ {\mathrm{T}}_2^{\ast } $$ was measured up to three times more accurately compared with standard six-echo UTE. In human lungs, T 2 * $$ {\mathrm{T}}_2^{\ast } $$ maps were successfully obtained from 10 echoes, yielding average values T 2 * $$ {\mathrm{T}}_2^{\ast } $$ = 1.62 ± 0.48 ms for mono-exponential and T 2 s * $$ {\mathrm{T}}_{2s}^{\ast } $$ = 1.00 ± 0.53 ms for bi-exponential models. CONCLUSION: A UTE sequence using δTE was implemented and validated on short T 2 * $$ {\mathrm{T}}_2^{\ast } $$ phantoms. The sequence was successfully applied for lung imaging; the bi-exponential signal model fit for human lung imaging may provide valuable insights into the diseased human lungs.


Subject(s)
Imaging, Three-Dimensional , Magnetic Resonance Imaging , Adult , Humans , Imaging, Three-Dimensional/methods , Magnetic Resonance Imaging/methods , Phantoms, Imaging , Image Interpretation, Computer-Assisted/methods , Lung/diagnostic imaging
4.
J Magn Reson Imaging ; 58(1): 174-186, 2023 07.
Article in English | MEDLINE | ID: mdl-36971493

ABSTRACT

BACKGROUND: Amide proton transfer (APT) weighted chemical exchange saturation transfer CEST (APTw/CEST) magnetic resonance imaging (MRI) has been suggested as having the potential for assessing the therapeutic effect of brain tumors or rectal cancer. Moreover, diffusion-weighted imaging (DWI) and positron emission tomography fused with computed tomography by means of 2-[fluorine-18]-fluoro-2-deoxy-D-glucose (FDG-PET/CT) have been suggested as useful in same setting. PURPOSE: To compare the capability of APTw/CEST imaging, DWI, and FDG-PET/CT for predicting therapeutic effect of chemoradiotherapy (CRT) on stage III non-small cell lung cancer (NSCLC) patients. STUDY TYPE: Prospective. POPULATION: Eighty-four consecutive patients with Stage III NSCLC, 45 men (age range, 62-75 years; mean age, 71 years) and 39 women (age range, 57-75 years; mean age, 70 years). All patients were then divided into two groups (Response Evaluation Criteria in Solid Tumors [RECIST] responders, consisting of the complete response and partial response groups, and RECIST non-responders, consisting of the stable disease and progressive disease groups). FIELD STRENGTH/SEQUENCE: 3 T, echo planar imaging or fast advanced spin-echo (FASE) sequences for DWI and 2D half Fourier FASE sequences with magnetization transfer pulses for CEST imaging. ASSESSMENT: Magnetization transfer ratio asymmetry (MTRasym ) at 3.5 ppm, apparent diffusion coefficient (ADC), and maximum standard uptake value (SUVmax, ) on PET/CT were assessed by means of region of interest (ROI) measurements at primary tumor. STATISTICAL TESTS: Kaplan-Meier method followed by log-rank test and Cox proportional hazards regression analysis with multivariate analysis. A P value <0.05 was considered statistically significant. RESULTS: Progression-free survival (PFS) and overall survival (OS) had significant difference between two groups. MTRasym at 3.5 ppm (hazard ratio [HR] = 0.70) and SUVmax (HR = 1.41) were identified as significant predictors for PFS. Tumor staging (HR = 0.57) was also significant predictors for OS. DATA CONCLUSION: APTw/CEST imaging showed potential performance as DWI and FDG-PET/CT for predicting the therapeutic effect of CRT on stage III NSCLC patients. LEVEL OF EVIDENCE: 2 TECHNICAL EFFICACY: Stage 1.


Subject(s)
Carcinoma, Non-Small-Cell Lung , Lung Neoplasms , Male , Humans , Female , Middle Aged , Aged , Carcinoma, Non-Small-Cell Lung/diagnostic imaging , Carcinoma, Non-Small-Cell Lung/therapy , Positron Emission Tomography Computed Tomography/methods , Lung Neoplasms/diagnostic imaging , Lung Neoplasms/therapy , Lung Neoplasms/pathology , Fluorodeoxyglucose F18 , Prospective Studies , Magnetic Resonance Imaging/methods , Chemoradiotherapy , Radiopharmaceuticals
5.
J Magn Reson Imaging ; 57(1): 259-272, 2023 01.
Article in English | MEDLINE | ID: mdl-35753082

ABSTRACT

BACKGROUND: Computed diffusion-weighted imaging (cDWI) is a mathematical computation technique that generates DWIs for any b-value by using actual DWI (aDWI) data with at least two different b-values and may improve differentiation of metastatic from nonmetastatic lymph nodes. PURPOSE: To determine the appropriate b-value for cDWI to achieve a better diagnostic capability for lymph node staging (N-staging) in non-small cell lung cancer (NSCLC) patients compared to aDWI, short inversion time (TI) inversion recovery (STIR) imaging, or positron emission tomography with 2-[fluorine-18] fluoro-2-deoxy-d-glucose combined with computed tomography (FDG-PET/CT). STUDY TYPE: Prospective. SUBJECTS: A total of 245 (127 males and 118 females; mean age 72 years) consecutive histopathologically confirmed NSCLC patients. FIELD STRENGTH/SEQUENCE: A 3 T, half-Fourier single-shot turbo spin-echo sequence, electrocardiogram (ECG)-triggered STIR fast advanced spin-echo (FASE) sequence with black blood and STIR acquisition and DWI obtained by FASE with b-values of 0 and 1000 sec/mm2 . ASSESSMENT: From aDWIs with b-values of 0 and 1000 (aDWI1000 ) sec/mm2 , cDWI using 400 (cDWI400 ), 600 (cDWI600 ), 800 (cDWI800 ), and 2000 (cDWI2000 ) sec/mm2 were generated. Then, 114 metastatic and 114 nonmetastatic nodes (mediastinal and hilar lymph nodes) were selected and evaluated with a contrast ratio (CR) for each cDWI and aDWI, apparent diffusion coefficient (ADC), lymph node-to-muscle ratio (LMR) on STIR, and maximum standard uptake value (SUVmax ). STATISTICAL TESTS: Receiver operating characteristic curve (ROC) analysis, Youden index, and McNemar's test. RESULTS: Area under the curve (AUC) of CR600 was significantly larger than the CR400 , CR800 , CR2000 , aCR1000 , and SUVmax . Comparison of N-staging accuracy showed that CR600 was significantly higher than CR400 , CR2000 , ADC, aCR1000 , and SUVmax , although there were no significant differences with CR800 (P = 0.99) and LMR (P = 0.99). DATA CONCLUSION: cDWI with b-value at 600 sec/mm2 may have potential to improve N-staging accuracy as compared with aDWI, STIR, and PET/CT. EVIDENCE LEVEL: 2 TECHNICAL EFFICACY: Stage 2.


Subject(s)
Carcinoma, Non-Small-Cell Lung , Lung Neoplasms , Male , Female , Humans , Aged , Carcinoma, Non-Small-Cell Lung/diagnostic imaging , Positron Emission Tomography Computed Tomography , Prospective Studies , Lung Neoplasms/diagnostic imaging , Lung Neoplasms/pathology , Positron-Emission Tomography/methods , Diffusion Magnetic Resonance Imaging/methods , Deoxyglucose , Fluorodeoxyglucose F18 , Radiopharmaceuticals , Neoplasm Staging
6.
Eur Radiol ; 33(1): 368-379, 2023 Jan.
Article in English | MEDLINE | ID: mdl-35841417

ABSTRACT

OBJECTIVE: Ultra-high-resolution CT (UHR-CT), which can be applied normal resolution (NR), high-resolution (HR), and super-high-resolution (SHR) modes, has become available as in conjunction with multi-detector CT (MDCT). Moreover, deep learning reconstruction (DLR) method, as well as filtered back projection (FBP), hybrid-type iterative reconstruction (IR), and model-based IR methods, has been clinically used. The purpose of this study was to directly compare lung CT number and airway dimension evaluation capabilities of UHR-CT using different scan modes with those of MDCT with different reconstruction methods as investigated in a lung density and airway phantom design recommended by QIBA. MATERIALS AND METHODS: Lung CT number, inner diameter (ID), inner area (IA), and wall thickness (WT) were measured, and mean differences between measured CT number, ID, IA, WT, and standard reference were compared by means of Tukey's HSD test between all UHR-CT data and MDCT reconstructed with FBP as 1.0-mm section thickness. RESULTS: For each reconstruction method, mean differences in lung CT numbers and all airway parameters on 0.5-mm and 1-mm section thickness CTs obtained with SHR and HR modes showed significant differences with those obtained with the NR mode on UHR-CT and MDCT (p < 0.05). Moreover, the mean differences on all UHR-CTs obtained with SHR, HR, or NR modes were significantly different from those of 1.0-mm section thickness MDCTs reconstructed with FBP (p < 0.05). CONCLUSION: Scan modes and reconstruction methods used for UHR-CT were found to significantly affect lung CT number and airway dimension evaluations as did reconstruction methods used for MDCT. KEY POINTS: • Scan and reconstruction methods used for UHR-CT showed significantly higher CT numbers and smaller airway dimension evaluations as did those for MDCT in a QIBA phantom study (p < 0.05). • Mean differences in lung CT number for 0.25-mm, 0.5-mm, and 1.0-mm section thickness CT images obtained with SHR and HR modes were significantly larger than those for CT images at 1.0-mm section thickness obtained with MDCT and reconstructed with FBP (p < 0.05). • Mean differences in inner diameter (ID), inner area (IA), and wall thickness (WT) measured with SHR and HR modes on 0.5- and 1.0-mm section thickness CT images were significantly smaller than those obtained with NR mode on UHR-CT and MDCT (p < 0.05).


Subject(s)
Deep Learning , Humans , Phantoms, Imaging , Tomography, X-Ray Computed/methods , Lung/diagnostic imaging , Thorax , Radiation Dosage , Radiographic Image Interpretation, Computer-Assisted/methods , Algorithms
7.
Neuroradiology ; 65(1): 65-75, 2023 Jan.
Article in English | MEDLINE | ID: mdl-35851924

ABSTRACT

PURPOSE: Bayesian estimation with advanced noise reduction (BEANR) in CT perfusion (CTP) could deliver more reliable cerebral blood flow (CBF) measurements than the commonly used reformulated singular value decomposition (rSVD). We compared the efficacy of CBF measurement by CTP using BEANR and rSVD, evaluating both relative to N-isopropyl-p-[(123) I]- iodoamphetamine (123I-IMP) single-photon emission computed tomography (SPECT) as a reference standard, in patients with cerebrovascular disease. METHODS: Thirty-one patients with suspected cerebrovascular disease underwent both CTP on a 320 detector-row CT system and SPECT. We applied rSVD and BEANR in the ischemic and contralateral regions to create CBF maps and calculate CBF ratios from the ischemic side to the healthy contralateral side (CBF index). The analysis involved comparing the CBF index between CTP methods and SPECT using Pearson's correlation and limits of agreement determined with Bland-Altman analyses, before comparing the mean difference in the CBF index between each CTP method and SPECT using the Wilcoxon matched pairs signed-rank test. RESULTS: The CBF indices of BEANR and 123I-IMP SPECT were significantly and positively correlated (r = 0.55, p < 0.0001), but there was no significant correlation between the rSVD method and SPECT (r = 0.15, p > 0.05). BEANR produced smaller limits of agreement for CBF than rSVD. The mean difference in the CBF index between BEANR and SPECT differed significantly from that between rSVD and SPECT (p < 0.001). CONCLUSIONS: BEANR has a better potential utility for CBF measurement in CTP than rSVD compared to SPECT in patients with cerebrovascular disease.


Subject(s)
Cerebrovascular Disorders , Humans , Bayes Theorem , Tomography, Emission-Computed, Single-Photon/methods , Tomography, X-Ray Computed/methods , Cerebrovascular Circulation , Perfusion Imaging
8.
J Comput Assist Tomogr ; 47(3): 494-499, 2023.
Article in English | MEDLINE | ID: mdl-37185016

ABSTRACT

OBJECTIVE: Although amide proton transfer-weighted (APTw) imaging is reported by 2-dimensional (2D) spin-echo-based sequencing, 3-dimensional (3D) APTw imaging can be obtained by gradient-echo-based sequencing. The purpose of this study was to compare the efficacy of APTw imaging between 2D and 3D imaging in patients with various brain tumors. METHODS: A total of 49 patients who had undergone 53 examinations [5 low-grade gliomas (LGG), 16 high-grade gliomas (HGG), 6 malignant lymphomas, 4 metastases, and 22 meningiomas] underwent APTw imaging using 2D and 3D sequences. The magnetization transfer ratio asymmetry (MTR asym ) was assessed by means of region of interest measurements. Pearson correlation was performed to determine the relationship between MTR asym for the 2 methods, and Student's t test to compare MTR asym for LGG and HGG. The diagnostic accuracy to differentiate HGG from LGG of the 2 methods was compared by means of the McNemar test. RESULTS: Three-dimensional APTw imaging showed a significant correlation with 2D APTw imaging ( r = 0.79, P < 0.0001). The limits of agreement between the 2 methods were -0.021 ± 1.42%. The MTR asym of HGG (2D: 1.97 ± 0.96, 3D: 2.11 ± 0.95) was significantly higher than those of LGG (2D: 0.46 ± 0.89%, P < 0.01; 3D: 0.15 ± 1.09%, P < 0.001). The diagnostic performance of the 2 methods to differentiate HGG from LGG was not significantly different ( P = 1). CONCLUSIONS: The potential capability of 3D APTw imaging is equal to or greater than that of 2D APTw imaging and is considered at least as valuable in patients with brain tumors.


Subject(s)
Brain Neoplasms , Glioma , Meningeal Neoplasms , Humans , Protons , Magnetic Resonance Imaging/methods , Amides , Brain Neoplasms/diagnostic imaging , Brain Neoplasms/pathology , Glioma/diagnostic imaging , Glioma/pathology , Imaging, Three-Dimensional
9.
Eur Spine J ; 32(12): 4153-4161, 2023 12.
Article in English | MEDLINE | ID: mdl-37837558

ABSTRACT

PURPOSE: It is still unclear how lumbar spinal surgery affects the lipid metabolism of patients with lumbar spinal disorders (LSDs) such as lumbar spinal canal stenosis and lumbar disk herniation. The present study aimed to assess the impact of lumbar spinal surgery on lipid metabolism in patients with LSDs and clarify the factors associated with changes in visceral fat (VF) accumulation before and after lumbar spinal surgery. METHODS: Consecutive patients with lumbar spinal surgery for LSDs were prospectively included. Abdominal computed tomography images and blood examination of the participants were evaluated before surgery and at 6 months and 1 year after surgery. The cross-sectional VF area (VFA) was measured at the level of the navel using computed tomography images. Blood examination items included triglycerides and high-density lipoprotein (HDL). RESULTS: The study enrolled a total of 138 patients. Female patients with LSDs had significantly increased VFA and serum triglyceride levels after lumbar spinal surgery. On multivariable analysis, the group with > 100 cm2 of preoperative VFA and a postoperative decrease in VFA had a significantly worse preoperative walking ability based on the Japanese Orthopaedic Association Back Pain Evaluation Questionnaire (relative risk 2.1; 95% confidence intervals 1.1-4.1). CONCLUSIONS: The present study demonstrated that patients with LSDs did not necessarily improve their lipid metabolism after lumbar spinal surgery. Instead, female patients with LSDs had significantly deteriorated lipid metabolism after lumbar spinal surgery. Finally, a worse preoperative walking ability was associated with the improvement in excess VF accumulation after lumbar spinal surgery.


Subject(s)
Decompression, Surgical , Spinal Stenosis , Female , Humans , Cross-Sectional Studies , Decompression, Surgical/methods , Lipid Metabolism , Lumbar Vertebrae/diagnostic imaging , Lumbar Vertebrae/surgery , Spinal Stenosis/complications , Spinal Stenosis/diagnostic imaging , Spinal Stenosis/surgery , Treatment Outcome , Prospective Studies
10.
Medicina (Kaunas) ; 59(1)2023 Jan 05.
Article in English | MEDLINE | ID: mdl-36676735

ABSTRACT

Background and Objectives: Modic type 1 is known to be associated with lower back pain (LBP), but at present, a treatment has not been fully established. Meanwhile, platelet-rich plasma (PRP) has been used for tissue regeneration and repair in the clinical setting. There is no clinical PRP injection trial for the intervertebral disc of LBP patients with Modic type 1. Thus, this study aimed to verify PRP injection safety and efficacy in LBP patients with Modic type 1. As a preliminary experiment, two LBP cases with Modic type 1 are presented. Materials and Methods: PRP was administered intradiscally to two LBP patients with Modic type 1. PRP was obtained from the patients' anticoagulated blood. Primary endpoints were physical condition, laboratory data, and X-ray for safety evaluation. Secondary endpoints were pain scores using the visual analog scale (VAS), the Oswestry Disability Index (ODI), and the Roland-Morris Disability Questionnaire (RDQ) to evaluate PRP efficacy. The observation period was 24 weeks after the PRP injection. In addition, changes in Modic type 1 using MRI were evaluated. Results: This study assessed two LBP patients with Modic type 1. There were no adverse events in physical condition, laboratory data, or lumbar X-rays after injection. Follow-up MRI showed a decrease of high signal intensity on T2WI compared to before PRP administration. The pain scores tended to improve after the injection. Conclusions: PRP injection into the intervertebral disc of LBP patients with Modic type 1 might be safe and effective. This analysis will be continued as a prospective study to establish the efficacy.


Subject(s)
Intervertebral Disc Displacement , Intervertebral Disc , Low Back Pain , Platelet-Rich Plasma , Humans , Low Back Pain/therapy , Low Back Pain/complications , Prospective Studies , Intervertebral Disc Displacement/complications , Magnetic Resonance Imaging , Lumbar Vertebrae , Treatment Outcome
11.
Radiology ; 303(2): 373-381, 2022 05.
Article in English | MEDLINE | ID: mdl-35103536

ABSTRACT

Background Deep learning reconstruction (DLR) may improve image quality. However, its impact on diffusion-weighted imaging (DWI) of the prostate has yet to be assessed. Purpose To determine whether DLR can improve image quality of diffusion-weighted MRI at b values ranging from 1000 sec/mm2 to 5000 sec/mm2 in patients with prostate cancer. Materials and Methods In this retrospective study, images of the prostate obtained at DWI with a b value of 0 sec/mm2, DWI with a b value of 1000 sec/mm2 (DWI1000), DWI with a b value of 3000 sec/mm2 (DWI3000), and DWI with a b value of 5000 sec/mm2 (DWI5000) from consecutive patients with biopsy-proven cancer from January to June 2020 were reconstructed with and without DLR. Image quality was assessed using signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) from region-of-interest analysis and qualitatively assessed using a five-point visual scoring system (1 [very poor] to 5 [excellent]) for each high-b-value DWI sequence with and without DLR. The SNR, CNR, and visual score for DWI with and without DLR were compared with the paired t test and the Wilcoxon signed rank test with Bonferroni correction, respectively. Apparent diffusion coefficients (ADCs) from DWI with and without DLR were also compared with the paired t test with Bonferroni correction. Results A total of 60 patients (mean age, 67 years; age range, 49-79 years) were analyzed. DWI with DLR showed significantly higher SNRs and CNRs than DWI without DLR (P < .001); for example, with DWI1000 the mean SNR was 38.7 ± 0.6 versus 17.8 ± 0.6, respectively (P < .001), and the mean CNR was 18.4 ± 5.6 versus 7.4 ± 5.6, respectively (P < .001). DWI with DLR also demonstrated higher qualitative image quality than DWI without DLR (mean score: 4.8 ± 0.4 vs 4.0 ± 0.7, respectively, with DWI1000 [P = .001], 3.8 ± 0.7 vs 3.0 ± 0.8 with DWI3000 [P = .002], and 3.1 ± 0.8 vs 2.0 ± 0.9 with DWI5000 [P < .001]). ADCs derived with and without DLR did not differ substantially (P > .99). Conclusion Deep learning reconstruction improves the image quality of diffusion-weighted MRI scans of prostate cancer with no impact on apparent diffusion coefficient quantitation with a 3.0-T MRI system. © RSNA, 2022 Online supplemental material is available for this article. See also the editorial by Turkbey in this issue.


Subject(s)
Deep Learning , Prostatic Neoplasms , Aged , Diffusion Magnetic Resonance Imaging/methods , Humans , Male , Middle Aged , Prostate , Prostatic Neoplasms/diagnostic imaging , Retrospective Studies
12.
Radiology ; 302(3): 697-706, 2022 03.
Article in English | MEDLINE | ID: mdl-34846203

ABSTRACT

Background Pulmonary MRI with ultrashort echo time (UTE) has been compared with chest CT for nodule detection and classification. However, direct comparisons of these methods' capabilities for Lung CT Screening Reporting and Data System (Lung-RADS) evaluation remain lacking. Purpose To compare the capabilities of pulmonary MRI with UTE with those of standard- or low-dose thin-section CT for Lung-RADS classification. Materials and Methods In this prospective study, standard- and low-dose chest CT (270 mA and 60 mA, respectively) and MRI with UTE were used to examine consecutive participants enrolled between January 2017 and December 2020 who met American College of Radiology Appropriateness Criteria for lung cancer screening with low-dose CT. Probability of nodule presence was assessed for all methods with a five-point visual scoring system by two board-certified radiologists. All nodules were then evaluated in terms of their Lung-RADS classification using each method. To compare nodule detection capability of the three methods, consensus for performances was rated by using jackknife free-response receiver operating characteristic analysis, and sensitivity was compared by means of the McNemar test. In addition, weighted κ statistics were used to determine the agreement between Lung-RADS classification obtained with each method and the reference standard generated from standard-dose CT evaluated by two radiologists who were not included in the image analysis session. Results A total of 205 participants (mean age: 64 years ± 7 [standard deviation], 106 men) with 1073 nodules were enrolled. Figure of merit (FOM) (P < .001) had significant differences among three modalities (standard-dose CT: FOM = 0.91, low-dose CT: FOM = 0.89, pulmonary MRI with UTE: FOM = 0.94), with no evidence of false-positive findings in participants with all modalities (P > .05). Agreements for Lung-RADS classification between all modalities and the reference standard were almost perfect (standard-dose CT: κ = 0.82, P < .001; low-dose CT: κ = 0.82, P < .001; pulmonary MRI with UTE: κ = 0.82, P < .001). Conclusion In a lung cancer screening population, ultrashort echo time pulmonary MRI was comparable to standard- or low-dose CT for Lung CT Screening Reporting and Data System classification. © RSNA, 2021 Online supplemental material is available for this article. See also the editorial by Wielpütz in this issue.


Subject(s)
Lung Neoplasms/diagnostic imaging , Magnetic Resonance Imaging/methods , Multiple Pulmonary Nodules/diagnostic imaging , Tomography, X-Ray Computed/methods , Early Detection of Cancer , Female , Humans , Male , Middle Aged , Prospective Studies
13.
Eur Radiol ; 32(10): 6658-6667, 2022 Oct.
Article in English | MEDLINE | ID: mdl-35687136

ABSTRACT

OBJECTIVE: To compare the utility of deep learning reconstruction (DLR) for improving acquisition time, image quality, and intraductal papillary mucinous neoplasm (IPMN) evaluation for 3D MRCP obtained with parallel imaging (PI), multiple k-space data acquisition for each repetition time (TR) technique (Fast 3D mode multiple: Fast 3Dm) and compressed sensing (CS) with PI. MATERIALS AND METHODS: A total of 32 IPMN patients who had undergone 3D MRCPs obtained with PI, Fast 3Dm, and CS with PI and reconstructed with and without DLR were retrospectively included in this study. Acquisition time, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) obtained with all protocols were compared using Tukey's HSD test. Results of endoscopic ultrasound, ERCP, surgery, or pathological examination were determined as standard reference, and distribution classifications were compared among all 3D MRCP protocols by McNemar's test. RESULTS: Acquisition times of Fast 3Dm and CS with PI with and without DLR were significantly shorter than those of PI with and without DLR (p < 0.05). Each MRCP sequence with DLR showed significantly higher SNRs and CNRs than those without DLR (p < 0.05). IPMN distribution accuracy of PI with and without DLR and Fast 3Dm with DLR was significantly higher than that of Fast 3Dm without DLR and CS with PI without DLR (p < 0.05). CONCLUSION: DLR is useful for improving image quality and IPMN evaluation capability on 3D MRCP obtained with PI, Fast 3Dm, or CS with PI. Moreover, Fast 3Dm and CS with PI may play as substitution to PI for MRCP in patients with IPMN. KEY POINTS: • Mean examination times of multiple k-space data acquisitions for each TR and compressed sensing with parallel imaging were significantly shorter than that of parallel imaging (p < 0.0001). • When comparing image quality of 3D MRCPs with and without deep learning reconstruction, deep learning reconstruction significantly improved signal-to-noise ratio and contrast-to-noise ratio (p < 0.05). • IPMN distribution accuracies of parallel imaging with and without deep learning reconstruction (with vs. without: 88.0% vs. 88.0%) and multiple k-space data acquisitions for each TR with deep learning reconstruction (86.0%) were significantly higher than those of others (p < 0.05).


Subject(s)
Deep Learning , Pancreatic Intraductal Neoplasms , Pancreatic Neoplasms , Humans , Pancreatic Neoplasms/diagnostic imaging , Retrospective Studies , Signal-To-Noise Ratio
14.
AJR Am J Roentgenol ; 218(5): 899-908, 2022 05.
Article in English | MEDLINE | ID: mdl-34877872

ABSTRACT

BACKGROUND. Whole-body MRI and FDG PET/MRI have shown encouraging results for staging of thoracic malignancy but are poorly studied for staging of small cell lung cancer (SCLC). OBJECTIVE. The purpose of our study was to compare the performance of conventional staging tests, FDG PET/CT, whole-body MRI, and FDG PET/MRI for staging of SCLC. METHODS. This prospective study included 98 patients (64 men, 34 women; median age, 74 years) with SCLC who underwent conventional staging tests (brain MRI; neck, chest, and abdominopelvic CT; and bone scintigraphy), FDG PET/CT, and whole-body MRI within 2 weeks before treatment; coregistered FDG PET/MRI was generated. Two nuclear medicine physicians independently reviewed conventional tests and FDG PET/CT examinations in separate sessions, and two chest radiologists independently reviewed whole-body MRI and FDG PET/MRI examinations in separate sessions. Readers assessed T, N, and M categories; TNM stage; and Veterans Administration Lung Cancer Study Group (VALSG) stage. Reader pairs subsequently reached consensus. Stages determined clinically during tumor board sessions served as the reference standard. RESULTS. Accuracy for T category was higher (p < .05) for whole-body MRI (94.9%) and FDG PET/MRI (94.9%) than for FDG PET/CT (85.7%). Accuracy for N category was higher (p < .05) for whole-body MRI (84.7%), FDG PET/MRI (83.7%), and FDG PET/CT (81.6%) than for conventional staging tests (75.5%). Accuracy for M category was higher (p < .05) for whole-body MRI (94.9%), FDG PET/MRI (94.9%), and FDG PET/CT (94.9%) than for conventional staging tests (84.7%). Accuracy for TNM stage was higher (p < .05) for whole-body MRI (88.8%) and FDG PET/MRI (86.7%) than for FDG PET/CT (77.6%) and conventional staging tests (72.4%). Accuracy for VALSG stage was higher (p < .05) for whole-body MRI (95.9%), FDG PET/MRI (95.9%), and FDG PET/CT (98.0%) than for conventional staging tests (82.7%). Interobserver agreement, expressed as kappa coefficients, ranged from 0.81 to 0.94 across imaging tests and staging endpoints. CONCLUSION. FDG PET/CT, whole-body MRI, and coregistered FDG PET/MRI outperformed conventional tests for various staging endpoints in patients with SCLC. Whole-body MRI and FDG PET/MRI outperformed FDG PET/CT for T category and thus TNM stage, indicating the utility of MRI for assessing extent of local invasion in SCLC. CLINICAL IMPACT. Incorporation of either MRI approach may improve initial staging evaluation in SCLC.


Subject(s)
Lung Neoplasms , Small Cell Lung Carcinoma , Aged , Female , Fluorodeoxyglucose F18 , Humans , Lung Neoplasms/diagnostic imaging , Lung Neoplasms/pathology , Magnetic Resonance Imaging/methods , Male , Neoplasm Staging , Positron Emission Tomography Computed Tomography/methods , Positron-Emission Tomography/methods , Prospective Studies , Radiopharmaceuticals , Small Cell Lung Carcinoma/diagnostic imaging , Small Cell Lung Carcinoma/pathology , Whole Body Imaging/methods
15.
Acta Radiol ; 63(10): 1363-1373, 2022 Oct.
Article in English | MEDLINE | ID: mdl-34636644

ABSTRACT

BACKGROUND: The need for quantitative assessment of interstitial lung involvement on thin-section computed tomography (CT) has arisen in interstitial lung diseases including connective tissue disease (CTD). PURPOSE: To evaluate the capability of machine learning (ML)-based CT texture analysis for disease severity and treatment response assessments in comparison with qualitatively assessed thin-section CT for patients with CTD. MATERIAL AND METHODS: A total of 149 patients with CTD-related ILD (CTD-ILD) underwent initial and follow-up CT scans (total 364 paired serial CT examinations), pulmonary function tests, and serum KL-6 level tests. Based on all follow-up examination results, all paired serial CT examinations were assessed as "Stable" (n = 188), "Worse" (n = 98) and "Improved" (n = 78). Next, quantitative index changes were determined by software, and qualitative disease severity scores were assessed by consensus of two radiologists. To evaluate differences in each quantitative index as well as in disease severity score between paired serial CT examinations, Tukey's honestly significant difference (HSD) test was performed among the three statuses. Stepwise regression analyses were performed to determine changes in each pulmonary functional parameter and all quantitative indexes between paired serial CT scans. RESULTS: Δ% normal lung, Δ% consolidation, Δ% ground glass opacity, Δ% reticulation, and Δdisease severity score showed significant differences among the three statuses (P < 0.05). All differences in pulmonary functional parameters were significantly affected by Δ% normal lung, Δ% reticulation, and Δ% honeycomb (0.16 ≤r2 ≤0.42; P < 0.05). CONCLUSION: ML-based CT texture analysis has better potential than qualitatively assessed thin-section CT for disease severity assessment and treatment response evaluation for CTD-ILD.


Subject(s)
Connective Tissue Diseases , Lung Diseases, Interstitial , Connective Tissue Diseases/diagnosis , Connective Tissue Diseases/drug therapy , Humans , Lung/diagnostic imaging , Lung Diseases, Interstitial/diagnostic imaging , Lung Diseases, Interstitial/therapy , Machine Learning , Severity of Illness Index , Tomography, X-Ray Computed/methods
16.
Radiology ; 299(3): 508-523, 2021 06.
Article in English | MEDLINE | ID: mdl-33825513

ABSTRACT

Over the past few decades, pulmonary imaging technologies have advanced from chest radiography and nuclear medicine methods to high-spatial-resolution or low-dose chest CT and MRI. It is currently possible to identify and measure pulmonary pathologic changes before these are obvious even to patients or depicted on conventional morphologic images. Here, key technological advances are described, including multiparametric CT image processing methods, inhaled hyperpolarized and fluorinated gas MRI, and four-dimensional free-breathing CT and MRI methods to measure regional ventilation, perfusion, gas exchange, and biomechanics. The basic anatomic and physiologic underpinnings of these pulmonary functional imaging techniques are explained. In addition, advances in image analysis and computational and artificial intelligence (machine learning) methods pertinent to functional lung imaging are discussed. The clinical applications of pulmonary functional imaging, including both the opportunities and challenges for clinical translation and deployment, will be discussed in part 2 of this review. Given the technical advances in these sophisticated imaging methods and the wealth of information they can provide, it is anticipated that pulmonary functional imaging will be increasingly used in the care of patients with lung disease. © RSNA, 2021 Online supplemental material is available for this article.


Subject(s)
Lung Diseases/diagnostic imaging , Lung Diseases/physiopathology , Magnetic Resonance Imaging/methods , Tomography, X-Ray Computed/methods , Artificial Intelligence , Contrast Media , Humans , Image Enhancement/methods , Image Interpretation, Computer-Assisted/methods , Respiratory Function Tests
17.
Radiology ; 299(3): 524-538, 2021 06.
Article in English | MEDLINE | ID: mdl-33847518

ABSTRACT

Pulmonary functional imaging may be defined as the regional quantification of lung function by using primarily CT, MRI, and nuclear medicine techniques. The distribution of pulmonary physiologic parameters, including ventilation, perfusion, gas exchange, and biomechanics, can be noninvasively mapped and measured throughout the lungs. This information is not accessible by using conventional pulmonary function tests, which measure total lung function without viewing the regional distribution. The latter is important because of the heterogeneous distribution of virtually all lung disorders. Moreover, techniques such as hyperpolarized xenon 129 and helium 3 MRI can probe lung physiologic structure and microstructure at the level of the alveolar-air and alveolar-red blood cell interface, which is well beyond the spatial resolution of other clinical methods. The opportunities, challenges, and current stage of clinical deployment of pulmonary functional imaging are reviewed, including applications to chronic obstructive pulmonary disease, asthma, interstitial lung disease, pulmonary embolism, and pulmonary hypertension. Among the challenges to the deployment of pulmonary functional imaging in routine clinical practice are the need for further validation, establishment of normal values, standardization of imaging acquisition and analysis, and evidence of patient outcomes benefit. When these challenges are addressed, it is anticipated that pulmonary functional imaging will have an expanding role in the evaluation and management of patients with lung disease.


Subject(s)
Lung Diseases/diagnostic imaging , Lung Diseases/physiopathology , Contrast Media , Early Diagnosis , Humans , Image Enhancement/methods , Image Interpretation, Computer-Assisted/methods , Quality Improvement , Respiratory Function Tests
18.
Radiology ; 298(3): 531-549, 2021 03.
Article in English | MEDLINE | ID: mdl-33399507

ABSTRACT

Pulmonary hypertension (PH) is defined by a mean pulmonary artery pressure greater than 20 mm Hg and classified into five different groups sharing similar pathophysiologic mechanisms, hemodynamic characteristics, and therapeutic management. Radiologists play a key role in the multidisciplinary assessment and management of PH. A working group was formed from within the Fleischner Society based on expertise in the imaging and/or management of patients with PH, as well as experience with methodologies of systematic reviews. The working group identified key questions focusing on the utility of CT, MRI, and nuclear medicine in the evaluation of PH: (a) Is noninvasive imaging capable of identifying PH? (b) What is the role of imaging in establishing the cause of PH? (c) How does imaging determine the severity and complications of PH? (d) How should imaging be used to assess chronic thromboembolic PH before treatment? (e) Should imaging be performed after treatment of PH? This systematic review and position paper highlights the key role of imaging in the recognition, work-up, treatment planning, and follow-up of PH. This article is a simultaneous joint publication in Radiology and European Respiratory Journal. The articles are identical except for stylistic changes in keeping with each journal's style. Either version may be used in citing this article. © 2021 RSNA and the European Respiratory Society. Online supplemental material is available for this article.

19.
Eur Respir J ; 57(1)2021 01.
Article in English | MEDLINE | ID: mdl-33402372

ABSTRACT

Pulmonary hypertension (PH) is defined by a mean pulmonary artery pressure greater than 20 mmHg and classified into five different groups sharing similar pathophysiologic mechanisms, haemodynamic characteristics, and therapeutic management. Radiologists play a key role in the multidisciplinary assessment and management of PH. A working group was formed from within the Fleischner Society based on expertise in the imaging and/or management of patients with PH, as well as experience with methodologies of systematic reviews. The working group identified key questions focusing on the utility of CT, MRI, and nuclear medicine in the evaluation of PH: a) Is noninvasive imaging capable of identifying PH? b) What is the role of imaging in establishing the cause of PH? c) How does imaging determine the severity and complications of PH? d) How should imaging be used to assess chronic thromboembolic PH before treatment? e) Should imaging be performed after treatment of PH? This systematic review and position paper highlights the key role of imaging in the recognition, work-up, treatment planning, and follow-up of PH.


Subject(s)
Hypertension, Pulmonary , Adult , Hemodynamics , Humans , Hypertension, Pulmonary/diagnostic imaging , Magnetic Resonance Imaging , Systematic Reviews as Topic
20.
J Magn Reson Imaging ; 53(4): 1042-1051, 2021 04.
Article in English | MEDLINE | ID: mdl-33205499

ABSTRACT

BACKGROUND: Oxygen (O2 )-enhanced MRI is mainly performed by a 2D sequence using 1.5T MR systems but trying to be obtained by a 3D sequence using a 3T MR system. PURPOSE: To compare the capability of 3D O2 -enhanced MRI and that of thin-section computed tomography (CT) for pulmonary functional loss assessment and clinical stage classification of chronic obstructive pulmonary disease (COPD) in smokers. STUDY TYPE: Prospective study. POPULATION: Fifty six smokers were included. FIELD STRENGTH/ SEQUENCE: 3T, 3D O2 -enhanced MRIs were performed with a 3D T1 -weighted fast field echo pulse sequence using the multiple flip angles. ASSESSMENTS: Smokers were classified into four stages ("Without COPD," "Mild COPD," "Moderate COPD," "Severe or very severe COPD"). Maps of regional changes in T1 values were generated from O2 -enhanced MR data. Regions of interest (ROIs) were then placed over the lung on all slices and averaged to determine mean T1 value change (ΔT1 ). Quantitative CT used the percentage of low attenuation areas within the entire lung (LAA%). STATISTICAL TESTS: ΔT1 and LAA% were correlated with pulmonary functional parameters, and compared for four stages using Tukey's Honestly Significant Difference test. Discrimination analyses were performed and McNemar's test was used for a comparison of the accuracy of the indexes. RESULTS: There were significantly higher correlations between ΔT1 and pulmonary functional parameters (-0.83 ≤ r ≤ -0.71, P < 0.05) than between LAA% and the same pulmonary functional parameters (-0.76 ≤ r ≤ -0.69, P < 0.05). ΔT1 and LAA% of the "Mild COPD" and "Moderate COPD" groups were significantly different from those of the "Severe or Very Severe COPD" group (P < 0.05). Discriminatory accuracy of ΔT1 (62.5%) and ΔT1 with LAA% (67.9%) was significantly greater than that of LAA% (48.2%, P < 0.05). DATA CONCLUSION: Compared with thin-section CT, 3D O2 -enhanced MRI has a similar capability for pulmonary functional assessment but better potential for clinical stage classification in smokers. LEVEL OF EVIDENCE: 2 TECHNICAL EFFICACY STAGE: 1.


Subject(s)
Pulmonary Disease, Chronic Obstructive , Smokers , Humans , Lung/diagnostic imaging , Magnetic Resonance Imaging , Oxygen , Prospective Studies , Pulmonary Disease, Chronic Obstructive/diagnostic imaging , Respiratory Function Tests , Smoking , Tomography, X-Ray Computed
SELECTION OF CITATIONS