ABSTRACT
The problem of underdosing lesions adjacent to upper respiratory cavities and a proposal to correct it are presented in this work. The EGS4 Monte Carlo code was used to simulate a 6 MV x-ray beam passing through a block of tissues with air cavities 2, 4, and 6 cm wide. The geometry used approximates the tracheal geometry used by previous researchers who investigated the underdosing phenomenon. A uniform longitudinal magnetic field of 0.5 T strength is used to reduce secondary electron outscatter caused by the presence of an air gap, and thus improving the dose at the distal surface of air cavities. We introduce the term "percent dose reduction" (PDR), which is defined as the difference between the dose after the air cavity and the dose at the same depth in soft-tissue phantom normalized to the dose in the tissue phantom, to quantify the reduction in dose after an air gap. We also introduce the term dose improvement ratio (DIR), which is defined as the dose ratio with magnetic field to the dose, at the same point, without magnetic field, to quantify the improvement in dose when the magnetic field is applied. For 2 x 2 x 20 cm3 and 4 x 4 x 20 cm3 air cavities irradiated by 2 x 2 cm2 beams, we found PDRs of 38% and 52%, respectively. This means that for these cavities, there is a 38% and a 52% reduction in dose at the cavity edge compared to the same dose in tissue at the same depth for each cavity. The dose improved by 30% (DIR= 1.3) and 87% (DIR= 1.87), respectively, when applying the magnetic field. The worst effect on dose at the distal side came from larger cavities irradiated with small fields. In these situations, the improvement in dose due to the presence of magnetic field was the largest. This article deals with "ideal" head and neck geometries with a uniform magnetic field. In a paper to follow we will use a CT-based phantom to study the effect in realistic geometries with the presence of a magnetic field from a Helmholtz coil pair.