1.
Ann Clin Biochem ; 39(Pt 2): 145-7, 2002 Mar.
Artículo en Inglés | MEDLINE | ID: mdl-11928763

RESUMEN

BACKGROUND AND METHODS: Cardiac troponin I (cTnl) is well recognized as a specific marker for myocardial infarction. A fully automated, random access, fluorescent immunoassay for cTnl was evaluated in comparison with an established assay (Stratus cTnl) using samples from healthy subjects and from patients with cardiac disease. RESULTS: The detection and precision were acceptable, and no interference was observed from bilirubin, triglycerides, haemoglobin, rheumatoid factor, drugs (aspirin, dopamine, digoxin) or heparin. Results on the Kryptor cTnl assay correlated with those on the Stratus II cTnI assay. The optimum Kryptor cTnl concentration for acute myocardial infarction was 0.63 microg/L, with a sensitivity of 89% [95% confidence interval (CI): 75.9-96.3] and specificity of 98.9% (95% CI: 93.9-99.8). CONCLUSION: Kryptor cTnl immunoassay is suitable for use in the diagnosis of myocardial infarction.


Asunto(s)
Infarto del Miocardio/diagnóstico , Troponina I/sangre , Adolescente , Adulto , Anciano , Anciano de 80 o más Años , Biomarcadores/sangre , Femenino , Humanos , Inmunoensayo , Masculino , Persona de Mediana Edad , Infarto del Miocardio/sangre , Miocardio/química , Sensibilidad y Especificidad