RESUMEN
The American Association of Physicists in Medicine (AAPM) is a nonprofit professional society whose primary purposes are to advance the science, education, and professional practice of medical physics. The AAPM has more than 8000 members and is the principal organization of medical physicists in the United States. The AAPM will periodically define new practice guidelines for medical physics practice to help advance the science of medical physics and to improve the quality of service to patients throughout the United States. Existing medical physics practice guidelines will be reviewed for the purpose of revision or renewal, as appropriate, on their fifth anniversary or sooner. Each medical physics practice guideline represents a policy statement by the AAPM, has undergone a thorough consensus process in which it has been subjected to extensive review, and requires the approval of the Professional Council. The medical physics practice guidelines recognize that the safe and effective use of diagnostic and therapeutic radiology requires specific training, skills, and techniques, as described in each document. Reproduction or modification of the published practice guidelines and technical standards by those entities not providing these services is not authorized. The following terms are used in the AAPM practice guidelines: Must and Must Not: Used to indicate that adherence to the recommendation is considered necessary to conform to this practice guideline. While must is the term to be used in the guidelines, if an entity that adopts the guideline has shall as the preferred term, the AAPM considers that must and shall have the same meaning. Should and Should Not: Used to indicate a prudent practice to which exceptions may occasionally be made in appropriate circumstances.
Asunto(s)
Electrones , Oncología por Radiación , Humanos , Fotones , Física , Estados UnidosRESUMEN
The dosimetric stability of six TomoTherapy units was analyzed to investigate changes in performance over time and with system upgrades. Energy and output were tracked using monitor chamber signal, onboard megavoltage computed tomography (MVCT) detector profile, and external ion chamber measurements. The systems (and monitoring periods) include three Hi-Art (67, 61, and 65 mos.), two TomoHDA (31 and 26 mos.), and one Radixact unit (11 mos.), representing approximately 10 years of clinical use. The four newest systems use the Dose Control Stability (DCS) system and Fixed Target Linear Accelerator (linac) (FTL). The output stability is reported as deviation from reference monitor chamber signal for all systems and/or from an external chamber signal. The energy stability was monitored using relative (center versus off-axis) MVCT detector signal (beam profile) and/or the ratio of chamber measurements at 2 depths. The clinical TomoHDA data were used to benchmark the Radixact stability, which has the same FTL but runs at a higher dose rate. The output based on monitor chamber data of all systems is very stable. The standard deviation of daily output on the non-DCS systems was 0.94-1.52%. As expected, the DCS systems had improved standard deviation: 0.004-0.06%. The beam energy was also very stable for all units. The standard deviation in profile flatness was 0.23-0.62% for rotating target systems and 0.04-0.09% for FTL. Ion chamber output and PDD ratios supported these results. The output stability on the Radixact system during extended treatment delivery (20, 30, and 40 min) was comparable to a clinical TomoHDA system. For each system, results are consistent between different measurement tools and techniques, proving not only the dosimetric stability, but also these quality parameters can be confirmed with various metrics. The replacement history over extended time periods of the major dosimetric components of the different delivery systems (target, linac, and magnetron) is also reported.
Asunto(s)
Dosificación Radioterapéutica , Radioterapia de Intensidad Modulada/instrumentación , Humanos , Mantenimiento , Aceleradores de Partículas , Radiometría , Factores de Tiempo , Tomografía Computarizada por Rayos XRESUMEN
It was previously demonstrated that dose delivered by a conventional linear accelerator using IMRT or VMAT can be reconstructed - on patient or phantom datasets - using helical diode array measurements and a technique called planned dose perturbation (PDP). This allows meaningful and intuitive analysis of the agreement between the planned and delivered dose, including direct comparison of the dose-volume histograms. While conceptually similar to modulated arc techniques, helical tomotherapy introduces significant challenges to the PDP formalism, arising primarily from TomoTherapy delivery dynamics. The temporal characteristics of the delivery are of the same order or shorter than the dosimeter's update interval (50 ms). Additionally, the prevalence of often small and complex segments, particularly with the 1 cm Y jaw setting, lead to challenges related to detector spacing. Here, we present and test a novel method of tomotherapy-PDP (TPDP) designed to meet these challenges. One of the novel techniques introduced for TPDP is organization of the subbeams into larger subunits called sectors, which assures more robust synchronization of the measurement and delivery dynamics. Another important change is the optional application of a correction based on ion chamber (IC) measurements in the phantom. The TPDP method was validated by direct comparisons to the IC and an independent, biplanar diode array dosimeter previously evaluated for tomotherapy delivery quality assurance. Nineteen plans with varying complexity were analyzed for the 2.5 cm tomotherapy jaw setting and 18 for the 1 cm opening. The dose differences between the TPDP and IC were 1.0% ± 1.1% and 1.1% ± 1.1%, for 2.5 and 1.0 cm jaw plans, respectively. Gamma analysis agreement rates between TPDP and the independent array were: 99.1%± 1.8% (using 3% global normalization/3 mm criteria) and 93.4% ± 7.1% (using 2% global/2 mm) for the 2.5 cm jaw plans; for 1 cm plans, they were 95.2% ± 6.7% (3% G/3) and 83.8% ± 12% (2% G/2). We conclude that TPDP is capable of volumetric dose reconstruction with acceptable accuracy. However, the challenges of fast tomotherapy delivery dynamics make TPDP less precise than the IMRT/VMAT PDP version, particularly for the 1 cm jaw setting.
Asunto(s)
Algoritmos , Aceleradores de Partículas/instrumentación , Fantasmas de Imagen , Garantía de la Calidad de Atención de Salud , Radiometría/métodos , Planificación de la Radioterapia Asistida por Computador/métodos , Radioterapia de Intensidad Modulada/métodos , Neoplasias Endometriales/radioterapia , Femenino , Neoplasias de la Vesícula Biliar/radioterapia , Neoplasias de Cabeza y Cuello/radioterapia , Humanos , Dosificación RadioterapéuticaRESUMEN
This work describes a novel application of MR-guided online adaptive radiotherapy (MRgoART) in the management of patients whom urgent palliative care is indicated using statum-adaptive radiotherapy (STAT-ART). The implementation of STAT-ART, as performed at our institution, is presented including a discussion of the advantages and limitations compared to the standard of care for palliative radiotherapy on conventional c-arm linacs. MR-based treatment planning techniques of STAT-ART for density overrides and deformable image registration (DIR) of diagnostic CT to the treatment MR are also addressed.
RESUMEN
Magnetic resonance-guided radiation therapy (MRgRT) offers advantages for image guidance for radiotherapy treatments as compared to conventional computed tomography (CT)-based modalities. The superior soft tissue contrast of magnetic resonance (MR) enables an improved visualization of the gross tumor and adjacent normal tissues in the treatment of abdominal and thoracic malignancies. Online adaptive capabilities, coupled with advanced motion management of real-time tracking of the tumor, directly allow for high-precision inter-/intrafraction localization. The primary aim of this case series is to describe MR-based interventions for localizing targets not well-visualized with conventional image-guided technologies. The abdominal and thoracic sites of the lung, kidney, liver, and gastric targets are described to illustrate the technological advancement of MR-guidance in radiotherapy.
RESUMEN
PURPOSE: To determine and analyze the dosimetric consequences of current portal imaging practices for pediatric patients, and make specific recommendations for reducing exposure from portal imaging procedures. METHODS AND MATERIALS: A survey was sent to approximately 250 Children's Oncology Group (COG) member institutions asking a series of questions about their portal imaging practices. Three case studies are presented with dosimetric analysis to illustrate the magnitude of unintended dose received by nontarget tissues using the most common techniques from the survey. RESULTS: The vast majority of centers use double-exposure portal image techniques with a variety of open field margins. Only 17% of portal images were obtained during treatment, and for other imaging methods, few centers subtract monitor units from the treatment delivery. The number of monitor units used was nearly the same regardless of imager type, including electronic portal imaging devices. Eighty-six percent imaged all fields the first week and 17% imaged all fields every week. An additional 1,112 cm3 of nontarget tissue received 1 Gy in one of the example cases. Eight new recommendations are made, which will lower nontarget radiation doses with minimal impact on treatment verification accuracy. CONCLUSION: Based on the survey, changes can be made in portal imaging practices that will lower nontarget doses. It is anticipated that treatment verification accuracy will be minimally affected. Specific recommendations made to decrease the imaging dose and help lower the rate of radiation-induced secondary cancers in children are proposed for inclusion in future COG protocols using radiation therapy.
Asunto(s)
Neoplasias/diagnóstico por imagen , Protección Radiológica/métodos , Radiometría/métodos , Encuestas y Cuestionarios , Tecnología Radiológica/métodos , Adolescente , Calibración , Niño , Femenino , Humanos , Masculino , Neoplasias/radioterapia , Oncología por Radiación/normas , Protección Radiológica/normas , Radiometría/normas , Dosificación Radioterapéutica , Tecnología Radiológica/normas , Tomografía Computarizada por Rayos X/métodos , Película para Rayos XRESUMEN
PURPOSE: To assess the sensitivity of two commercial dosimetry systems in detecting Helical TomoTherapy (HT) delivery errors. METHOD: Two commercial dosimeters i) MatriXXEvolution and ii) ArcCHECK® were considered. Ten retrospective nasopharynx HT patients were analysed. For each patient, error plans were created by independently introducing systematic offsets in: a) Jaw width error ±1, ±1.5 and ±2mm, b) Couch speed error ±2%, ±2.5, ±3% and ±4%, and c) MLC Leaf Open Time (MLCLOT) errors (3 separate MLC errors: either leaf 32 open or leaf 42 remains open during delivery, and 4% random reductions in MLCLOT). All error plans, along with the no error plan for each patient, were measured using both dosimeters in the same session. Gamma evaluation (3%/3mm) was applied to quantitatively compare the measured dose from each dosimeter to the treatment planning system. The error sensitivity was quantified as the rate of decrease in gamma pass rate. RESULTS: The gamma pass rate decreases with increase in error magnitude for both dosimeters. ArcCHECK was insensitive for couch speed error up to 2.5% and jaw width error up to -1.5mm while MatriXXEvolution was found to be insensitive to couch speed error up to 2% and couch speed up to -1mm. Both of the detectors show similar sensitivity to all the MLCLOT errors that were clinically relevant. CONCLUSION: No statistically significant (p>0.05) differences exist in detecting the simulated delivery errors between MatriXXEvolution and ArcCHECK dosimeter systems for HT plans. Both dosimeters were able to pick up clinically relevant delivery errors.
Asunto(s)
Dosímetros de Radiación , Monitoreo de Radiación/instrumentación , Planificación de la Radioterapia Asistida por Computador , Radioterapia de Intensidad Modulada , Humanos , Fantasmas de Imagen , Garantía de la Calidad de Atención de Salud , Estudios Retrospectivos , Sensibilidad y EspecificidadRESUMEN
PURPOSE: Develop a framework to characterize helical TomoTherapy (HT) machine delivery errors and their clinical significance. METHOD AND MATERIALS: Ten nasopharynx HT plans were edited to introduce errors in Jaw width (JW), couch speed (CS), gantry period (GP), gantry start position (GSP), multi leaf collimator leaf open times (MLC LOT). In case of MLC LOT only, both systematic and random delivery errors were investigated. Each error type was simulated independently for a range of magnitudes. Dose distributions for the clinical reference plans and the error simulated plans were compared to establish the magnitude for each error type which resulted in a change in clinical tolerance, defined as 5% variation in D95 of PTV70, D0.1cc of spinal cord, D0.1cc of brainstem and the smallest value of either a 10% or 3.6Gy dose variation in mean parotid dose. RESULTS: Dose variation from systematic delivery errors in JW ±0.5mm, CS ranges between -1% to 1.5%, GP ±1s, GSP ranges between -20 to 2.50 and MLC LOT random error up to 2% from the planned value relative to the clinical reference plan was within the set tolerance values for all the patient cohorts. GSP errors and the random MLC LOT errors with up to 10% standard deviation were found to be relatively insensitive compared to other delivery errors. CONCLUSION: This work has established a framework to characterize HT machine delivery errors. This framework could be applied to any patient dataset to determine clinically relevant HT QA tolerances.
Asunto(s)
Neoplasias Nasofaríngeas/radioterapia , Errores de Configuración en Radioterapia , Radioterapia de Intensidad Modulada , Humanos , Garantía de la Calidad de Atención de Salud , Radiometría , Dosificación Radioterapéutica , Planificación de la Radioterapia Asistida por ComputadorRESUMEN
PURPOSE: The aim of this study was to validate the accuracy of an exit detector-based dose reconstruction tool for helical tomotherapy (HT) delivery quality assurance (DQA). METHODS AND MATERIAL: Exit detector-based DQA tool was developed for patient-specific HT treatment verification. The tool performs a dose reconstruction on the planning image using the sinogram measured by the HT exit detector with no objects in the beam (i.e., static couch), and compares the reconstructed dose to the planned dose. Vendor supplied (three "TomoPhant") plans with a cylindrical solid water ("cheese") phantom were used for validation. Each "TomoPhant" plan was modified with intentional multileaf collimator leaf open time (MLC LOT) errors to assess the sensitivity and robustness of this tool. Four scenarios were tested; leaf 32 was "stuck open," leaf 42 was "stuck open," random leaf LOT was closed first by mean values of 2% and then 4%. A static couch DQA procedure was then run five times (once with the unmodified sinogram and four times with modified sinograms) for each of the three "TomoPhant" treatment plans. First, the original optimized delivery plan was compared with the original machine agnostic delivery plan, then the original optimized plans with a known modification applied (intentional MLC LOT error) were compared to the corresponding error plan exit detector measurements. An absolute dose comparison between calculated and ion chamber (A1SL, Standard Imaging, Inc., WI, USA) measured dose was performed for the unmodified "TomoPhant" plans. A 3D gamma evaluation (2%/2 mm global) was performed by comparing the planned dose ("original planned dose" for unmodified plans and "adjusted planned dose" for each intentional error) to exit detector-reconstructed dose for all three "Tomophant" plans. Finally, DQA for 119 clinical (treatment length <25 cm) and three cranio-spinal irradiation (CSI) plans were measured with both the ArcCHECK phantom (Sun Nuclear Corp., Melbourne, FL, USA) and the exit detector DQA tool to assess the time required for DQA and similarity between two methods. RESULTS: The measured ion chamber dose agreed to within 1.5% of the reconstructed dose computed by the exit detector DQA tool on a cheese phantom for all unmodified "Tomophant" plans. Excellent agreement in gamma pass rate (>95%) was observed between the planned and reconstructed dose for all "Tomophant" plans considered using the tool. The gamma pass rate from 119 clinical plan DQA measurements was 94.9% ± 1.5% and 91.9% ± 4.37% for the exit detector DQA tool and ArcCHECK phantom measurements (P = 0.81), respectively. For the clinical plans (treatment length <25 cm), the average time required to perform DQA was 24.7 ± 3.5 and 39.5 ± 4.5 min using the exit detector QA tool and ArcCHECK phantom, respectively, whereas the average time required for the 3 CSI treatments was 35 ± 3.5 and 90 ± 5.2 min, respectively. CONCLUSION: The exit detector tool has been demonstrated to be faster for performing the DQA with equivalent sensitivity for detecting MLC LOT errors relative to a conventional phantom-based QA method. In addition, comprehensive MLC performance evaluation and features of reconstructed dose provide additional insight into understanding DQA failures and the clinical relevance of DQA results.
Asunto(s)
Dosis de Radiación , Radioterapia de Intensidad Modulada , Humanos , Control de Calidad , Dosificación RadioterapéuticaRESUMEN
PURPOSE: Magnetic resonance imaging-guided radiation therapy has entered clinical practice at several major treatment centers. Treatment of early-stage non-small cell lung cancer with stereotactic body radiation therapy is one potential application of this modality, as some form of respiratory motion management is important to address. We hypothesize that magnetic resonance imaging-guided tri-cobalt-60 radiation therapy can be used to generate clinically acceptable stereotactic body radiation therapy treatment plans. Here, we report on a dosimetric comparison between magnetic resonance imaging-guided radiation therapy plans and internal target volume-based plans utilizing volumetric-modulated arc therapy. MATERIALS AND METHODS: Ten patients with early-stage non-small cell lung cancer who underwent radiation therapy planning and treatment were studied. Following 4-dimensional computed tomography, patient images were used to generate clinically deliverable plans. For volumetric-modulated arc therapy plans, the planning tumor volume was defined as an internal target volume + 0.5 cm. For magnetic resonance imaging-guided plans, a single mid-inspiratory cycle was used to define a gross tumor volume, then expanded 0.3 cm to the planning tumor volume. Treatment plan parameters were compared. RESULTS: Planning tumor volumes trended larger for volumetric-modulated arc therapy-based plans, with a mean planning tumor volume of 47.4 mL versus 24.8 mL for magnetic resonance imaging-guided plans ( P = .08). Clinically acceptable plans were achievable via both methods, with bilateral lung V20, 3.9% versus 4.8% ( P = .62). The volume of chest wall receiving greater than 30 Gy was also similar, 22.1 versus 19.8 mL ( P = .78), as were all other parameters commonly used for lung stereotactic body radiation therapy. The ratio of the 50% isodose volume to planning tumor volume was lower in volumetric-modulated arc therapy plans, 4.19 versus 10.0 ( P < .001). Heterogeneity index was comparable between plans, 1.25 versus 1.25 ( P = .98). CONCLUSION: Magnetic resonance imaging-guided tri-cobalt-60 radiation therapy is capable of delivering lung high-quality stereotactic body radiation therapy plans that are clinically acceptable as compared to volumetric-modulated arc therapy-based plans. Real-time magnetic resonance imaging provides the unique capacity to directly observe tumor motion during treatment for purposes of motion management.
Asunto(s)
Carcinoma de Pulmón de Células no Pequeñas/radioterapia , Neoplasias Pulmonares/radioterapia , Radiocirugia/métodos , Radioterapia Guiada por Imagen/métodos , Carcinoma de Pulmón de Células no Pequeñas/diagnóstico por imagen , Carcinoma de Pulmón de Células no Pequeñas/patología , Radioisótopos de Cobalto/uso terapéutico , Femenino , Humanos , Pulmón/diagnóstico por imagen , Pulmón/patología , Pulmón/efectos de la radiación , Neoplasias Pulmonares/diagnóstico por imagen , Neoplasias Pulmonares/patología , Imagen por Resonancia Magnética/métodos , Masculino , Dosificación Radioterapéutica , Planificación de la Radioterapia Asistida por Computador/métodos , Radioterapia Guiada por Imagen/normas , Radioterapia de Intensidad Modulada/métodosRESUMEN
PURPOSE: External beam radiation therapy (EBRT) is effective for early glottic cancers, with cure rates of â¼90% for T1 tumors. EBRT has strengths but also disadvantages including radiation to healthy tissues and duration of 5-7 weeks. With advances in laryngeal framework surgery, new devices can provide reliable, minimally invasive access to the larynx. Such devices could be modified to insert brachytherapy catheters. Brachytherapy could provide focused radiation while limiting dose to normal structures in the larynx and neck. As a preliminary step, we performed simulations comparing EBRT to high-dose-rate brachytherapy to assess if this approach could provide dosimetric advantage. METHODS AND MATERIALS: One- and 2-catheter brachytherapy simulations were performed for 3 patients with T1 glottic carcinoma. Percentage of dose delivered to the target and adjacent structures was compared with conventional EBRT using 3D and intensity-modulated radiation therapy approaches. RESULTS: Percentage of structures exposed to 50% of the dose was lower for brachytherapy compared with 3D EBRT and intensity-modulated radiation therapy, particularly for the cricoid and contralateral arytenoid. Dose was also lower for the carotid-internal jugular vein complexes compared with 3D EBRT. Dose profiles did not differ significantly between 1- and 2-catheter simulations. CONCLUSION: Brachytherapy can decrease radiation to normal tissues including laryngeal cartilages and carotid-internal jugular vein complexes. Recent advancements allowing catheter placement may afford the potential to decrease radiation to healthy tissues with decreased treatment time. However, careful, stepwise evaluation of feasibility and outcomes in model systems is required before recommending this approach for such high cure rate cancers in humans.
Asunto(s)
Braquiterapia/efectos adversos , Carcinoma/radioterapia , Neoplasias Laríngeas/radioterapia , Dosis de Radiación , Traumatismos por Radiación/etiología , Anciano , Cartílago Aritenoides/efectos de la radiación , Braquiterapia/métodos , Arteria Carótida Interna/efectos de la radiación , Catéteres , Simulación por Computador , Cartílago Cricoides/efectos de la radiación , Fraccionamiento de la Dosis de Radiación , Glotis , Humanos , Venas Yugulares/efectos de la radiación , Masculino , Persona de Mediana Edad , Modelos Biológicos , Radioterapia de Intensidad Modulada/efectos adversosRESUMEN
SBRT is increasingly utilized in liver tumor treatment. MRI-guided RT allows for real-time MRI tracking during therapy. Liver tumors are often poorly visualized and most contrast agents are transient. Gadoxetate may allow for sustained tumor visualization. Here, we report on the first use of gadoxetate during real-time MRI-guided SBRT.
Asunto(s)
Neoplasias Hepáticas/cirugía , Radiocirugia/métodos , Anciano , Anciano de 80 o más Años , Medios de Contraste , Evaluación de Medicamentos/métodos , Estudios de Factibilidad , Gadolinio DTPA , Humanos , Neoplasias Hepáticas/diagnóstico , Imagen por Resonancia Magnética Intervencional/métodos , Persona de Mediana EdadRESUMEN
The multi-leaf collimator (MLC) assembly present on TomoTherapy (Accuray, Madison WI) radiation therapy (RT) and mega voltage CT machines is well suited to perform fluence field modulated CT (FFMCT). In addition, there is a demand in the RT environment for FFMCT imaging techniques, specifically volume of interest (VOI) imaging. A clinical TomoTherapy machine was programmed to perform VOI. Four different size ROIs were placed at varying distances from isocenter. Projections intersecting the VOI received 'full dose' while those not intersecting the VOI received 30% of the dose (i.e. the incident fluence for non VOI projections was 30% of the incident fluence for projections intersecting the VOI). Additional scans without fluence field modulation were acquired at 'full' and 30% dose. The noise (pixel standard deviation) and mean CT number were measured inside the VOI region and compared between the three scans. Dose maps were generated using a dedicated TomoTherapy treatment planning dose calculator. The VOI-FFMCT technique produced an image noise 1.05, 1.00, 1.03, and 1.05 times higher than the 'full dose' scan for ROI sizes of 10 cm, 13 cm, 10 cm, and 6 cm respectively within the VOI region. The VOI-FFMCT technique required a total imaging dose equal to 0.61, 0.69, 0.60, and 0.50 times the 'full dose' acquisition dose for ROI sizes of 10 cm, 13 cm, 10 cm, and 6 cm respectively within the VOI region. Noise levels can be almost unchanged within clinically relevant VOIs sizes for RT applications while the integral imaging dose to the patient can be decreased, and/or the image quality in RT can be dramatically increased with no change in dose relative to non-FFMCT RT imaging. The ability to shift dose away from regions unimportant for clinical evaluation in order to improve image quality or reduce imaging dose has been demonstrated. This paper demonstrates that FFMCT can be performed using the MLC on a clinical TomoTherapy machine for the first time.