J Cardiovasc Electrophysiol ; 32(1): 67-70, 2021 01.
Artículo en Inglés | MEDLINE | ID: mdl-33155334

RESUMEN

Deviation of the esophagus during atrial fibrillation (AF) ablation can reduce esophageal injury. This study reports upon a novel esophageal retractor that utilizes vacuum suction and mechanical deflection to deviate the esophagus. The device was used in seven patients undergoing cryoballoon AF ablation. The esophagus was deviated 31.9 ± 4.4 mm to the right and 28.2 ± 5.9 mm to the left. Endoscopy at 4.4 ± 1.5 days postablation showed no esophageal injury. This study demonstrates the safe and effective deviation of the esophagus without a trailing edge with an esophageal retractor utilizing vacuum suction and mechanical deflection.


Asunto(s)
Fibrilación Atrial , Ablación por Catéter , Fibrilación Atrial/diagnóstico , Fibrilación Atrial/cirugía , Ablación por Catéter/efectos adversos , Esófago/diagnóstico por imagen , Esófago/cirugía , Humanos , Succión , Vacio