1.
J Mass Spectrom ; 52(7): 472-479, 2017 Jul.
Artículo en Inglés | MEDLINE | ID: mdl-28544043

RESUMEN

A field-portable gas chromatography-mass spectrometry (GC-MS) system (Hapsite ER) was evaluated for the detection of nonvolatile V-type nerve agents (VX and Russian VX (RVX)) in the vapor phase. The Hapsite ER system consists of a Tri-Bed concentrator gas sampler, a nonpolar low thermal-mass capillary GC column and a hydrophobic membrane-interfaced electron ionization quadrupole mass spectrometer evacuated by a non-evaporative getter pump. The GC-MS system was attached to a VX-G fluoridating conversion tube containing silver nitrate and potassium fluoride. Sample vapors of VX and RVX were converted into O-ethyl methylphosphonofluoridate (EtGB) and O-isobutyl methylphosphonofluoridate (iBuGB), respectively. These fluoridated derivatives were detected within 10 min. No compounds were detected when the VX and RVX samples were analyzed without the conversion tube. A vapor sample of tabun (GA) was analyzed, in which GA and O-ethyl N,N-dimethylphosphoramidofluoridate were detected. The molar recovery percentages of EtGB and iBuGB from VX and RVX vapors varied from 0.3 to 17%, which was attributed to variations in the vaporization efficiency of the glass vapor container. The conversion efficiencies of the VX-G conversion tube for VX and RVX to their phosphonate derivatives were estimated to be 40%. VX and RVX vapors were detected at concentrations as low as 0.3 mg m-3 . Gasoline vapor was found to interfere with the analyses of VX and RVX. In the presence of 160 mg m-3 gasoline, the detection limits of VX and RVX vapor were increased to 20 mg m-3 . Copyright © 2017 John Wiley & Sons, Ltd.


Asunto(s)
Agentes Nerviosos/análisis , Compuestos Organotiofosforados/análisis , Fluoruros/química , Cromatografía de Gases y Espectrometría de Masas/instrumentación , Cromatografía de Gases y Espectrometría de Masas/métodos , Agentes Nerviosos/química , Organofosfatos/análisis , Organofosfatos/química , Compuestos Organotiofosforados/química , Compuestos de Potasio/química , Nitrato de Plata/química , Espectrometría de Masa por Ionización de Electrospray/métodos
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA