RESUMEN
OBJECTIVE: The purpose of this study is to assess the ability of a novel automatic tube potential selection tool to reduce radiation dose while maintaining diagnostic quality in CT angiography (CTA) and contrast-enhanced abdominopelvic CT. MATERIALS AND METHODS: One hundred one CTA examinations and 90 contrastenhanced abdominopelvic examinations were performed using an automatic tube potential selection tool on a 128-MDCT scanner. Two vascular radiologists and two abdominal radiologists evaluated the image quality for sharpness, noise, artifact, and diagnostic confidence. In a subset of patients who had undergone prior studies (CTA, 28 patients; abdominopelvic CT, 25 patients), a side-by-side comparison was performed by a separate radiologist. Dose reduction and iodine contrast-to-noise ratio resulting from use of the tool were calculated. RESULTS: For CTA, 80 or 100 kV was selected for 73% of the scans, with a mean dose reduction of 36% relative to the reference 120-kV protocol. For abdominopelvic CT examinations, 80 or 100 kV was used for 55% of the scans, with a mean dose reduction of 25%. Overall dose reduction relative to the reference 120-kV protocol was 25% and 13% for CTA and abdominopelvic CT scans, respectively. Over 98% of scans had acceptable sharpness, noise texture, artifact, and diagnostic confidence for both readers and diagnostic tasks; 94-100% of scans had acceptable noise. Iodine contrast-to-noise ratio was significantly higher than (p < 0.001) or similar to (p = 0.11) that of prior scans, and equivalent quality was achieved despite the dose reduction. CONCLUSION: Automatic tube potential selection provides an efficient and quantitativeway to guide the selection of the optimal tube potential for CTA and abdominopelvic CT examinations.