1.
Artigo em Inglês | MEDLINE | ID: mdl-34092806

RESUMO

We have evaluated a Josephson arbitrary waveform synthesizer (JAWS) in the voltage range from 1 µV to 1 mV at frequencies from 60 to 1000 Hz for the use in the calibration of lock-in amplifiers. The uncertainty contribution from the JAWS system is 45 nV for 1 mV at 1000 Hz and k = 2.0. We anticipate that the JAWS will help extend the lower voltage and frequency range of ac voltage metrology and improve the uncertainties by one order of magnitude compared to conventional techniques.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA