1.
J Appl Clin Med Phys ; 22(6): 241-252, 2021 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-34028955

RESUMO

The aim of this study was to benchmark the accuracy of the VIrtual Phantom Epid dose Reconstruction (VIPER) software for pre-treatment dosimetric verification of multiple-target stereotactic radiosurgery (SRS). VIPER is an EPID-based method to reconstruct a 3D dose distribution in a virtual phantom from in-air portal images. Validation of the VIPER dose calculation was assessed using several MLC-defined fields for a 6 MV photon beam. Central axis percent depth doses (PDDs) and output factors were measured with an ionization chamber in a water tank, while dose planes at a depth of 10 cm in a solid flat phantom were acquired with radiochromic films. The accuracy of VIPER for multiple-target SRS plan verification was benchmarked against Monte Carlo simulations. Eighteen multiple-target SRS plans designed with the Eclipse treatment planning system were mapped to a cylindrical water phantom. For each plan, the 3D dose distribution reconstructed by VIPER within the phantom was compared with the Monte Carlo simulation, using a 3D gamma analysis. Dose differences (VIPER vs. measurements) generally within 2% were found for the MLC-defined fields, while film dosimetry revealed gamma passing rates (GPRs) ≥95% for a 3%/1 mm criteria. For the 18 multiple-target SRS plans, average 3D GPRs greater than 93% and 98% for the 3%/2 mm and 5%/2 mm criteria, respectively. Our results validate the use of VIPER as a dosimetric verification tool for pre-treatment QA of single-isocenter multiple-target SRS plans. The method requires no setup time on the linac and results in an accurate 3D characterization of the delivered dose.


Assuntos
Radiocirurgia , Radioterapia de Intensidade Modulada , Humanos , Imagens de Fantasmas , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador , Software , Água
2.
Rep Pract Oncol Radiother ; 24(5): 421-427, 2019.
Artigo em Inglês | MEDLINE | ID: mdl-31367194

RESUMO

AIM: To investigate whether the target-isocenter deviations reported by a conventional Winston-Lutz (WL) test actually reflect the shifts of the measured prescription isodose line with respect to the target. BACKGROUND: A conventional WL test uses a metallic ball as a target that aims at several fields. But this test does not report information on the accuracy of the delivery in terms of dose. MATERIALS AND METHODS: A conventional WL test using a metallic pointer as a target (Pointer-WL test) has been recreated in the Eclipse treatment planning system over an acrylic phantom containing a radiochromic film (Dose-WL test). After Dose-WL test delivery, the shift of the 80% prescription isodose line with respect to the target center (d80%-center) was measured using film dosimetry. The Pointer-WL and Dose-WL tests were performed in 10 different sessions. The isocenter deviation reported by the Pointer-WL test was compared to the d80%-center vector, according to the three patient's directions (Left-Right or LR; Anterior-Posterior or AP; and Superior-Inferior or SI). RESULTS: The deviations (mean ± SD) found for the Dose-WL tests (LR: 0.5 ± 0.4 mm; AP: 0.5 ± 0.4 mm; SI: 0.6 ± 0.2 mm) were in most cases less than 1 mm, and they were significantly smaller (all p < 0.05) than the maximum deviations reported by the Pointer-WL tests (LR: 1.3 ± 0.3 mm; AP: 1.2 ± 0.4 mm; SI: 1.1 ± 0.3 mm). CONCLUSIONS: The Dose-WL test described in this study allows estimating the spatial accuracy of the prescription isodose line.

3.
J Med Imaging Radiat Sci ; 55(4): 101719, 2024 Jul 30.
Artigo em Inglês | MEDLINE | ID: mdl-39084157

RESUMO

INTRODUCTION: The aim of this study was to assess the results of the local pre-treatment verifications of online adaptive prostate SBRT plans performed by dosimetrists METHODS AND MATERIALS: Prostate SBRT treatments are planned in our department using an online adaptive method developed and validated by our group. The adaptive plans were computed on the daily CBCT scan using the Acuros XB v. 16.1 algorithm of the Varian Eclipse treatment planning system. Adaptive plans consisted of a single VMAT with 6 MV flattening-filter-free (FFF) energy performed on a Varian TrueBeam linac. Pre-treatment verification of the adaptive "plan-of-the-day" (POD) created in each treatment session was performed using the Mobius 3D v. 3.1 secondary dose calculation program (M3D). Commissioning of M3D included the tuning of the dosimetric leaf gap correction (DLGc) parameter. Generic and specific DLGc values were then derived using a set of plans for typical sites (prostate, head and neck, brain, lung and bone palliative) and another set were determined for specific online SBRT PODs (gDLGc and sDLGc, respectively). The first 50 prostate patients treated with the PACE-B schedule (5 × 7.25 Gy) were included, i.e., 250 adaptive SBRT PODs were collected in this study. For each online adaptive POD, a global 3D gamma comparison between the Eclipse 3D dose and the M3D dose in the patient CBCT was performed. Gamma passing rates (GPRs) for the whole external patient contour (Body) and the PTV were recorded, using the 5 % global /3 mm criteria. The target mean dose and target coverage differences between the Eclipse and M3D doses were also analyzed (ΔDmean and ΔD90 %, respectively). The accuracy of M3D was assessed against PRIMO Monte Carlo software. Twenty-five online prostate SBRT PODs were randomly selected from the set of 250 adaptive plans and simulated with PRIMO. RESULTS: Values of -1 mm and -0.14 mm were found as optimal gDLGc and sDLGc, respectively. Over the 250 online adaptive PODs, excellent GPR values ∼ 100 % were obtained for the Body and PTV structures, regardless the type of DLGc used. The use of the sDLGc instead of the gDLGc provided better results for ΔDmean (0.1 % ± 0.5% vs. -1.9 ± 0.7 %) and ΔD90 % (-1.0 % ± 0.5 %. vs. -3.5 % ± 0.8 %). This issue was also observed when M3D calculations were compared to PRIMO simulations. CONCLUSIONS: M3D can be effectively used for independent pre-treatment verifications of online adaptive prostate SBRT plans. The use of a specific DLGc value is advised for this SBRT online adaptive technique.

4.
Phys Med ; 102: 19-26, 2022 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-36037748

RESUMO

PURPOSE: To investigate the feasibility of using the free PRIMO Monte Carlo software for independent dose check of cranial SRS plans designed with the Varian HyperArc (HA) technique. MATERIALS AND METHODS: In this study, the PRIMO Monte Carlo software v. 0.3.64.1800 was used with the phase-space files (v. 2, Feb. 27, 2013) provided by Varian for 6 MV flattening-filter-free (FFF) photon beams from a Varian TrueBeam linear accelerator (linac), equipped with a Millennium 120 multileaf collimator (MLC). This configuration was validated by comparing the percentage depth doses (PDDs), lateral profiles and relative output factors (OFs) simulated in a water phantom against measurements for field sizes from 1 × 1 to 40 × 40 cm2. The agreement between simulated and experimental relative dose curves was evaluated using a global (G) gamma index analysis. In addition, the accuracy of PRIMO to model the MLC was investigated (dosimetric leaf gap, tongue and groove, leaf transmission and interleaf leakage). Thirty-five HA SRS plans computed in the Eclipse treatment planning system (TPS) were simulated in PRIMO. The Acuros XB algorithm v. 16.10 (dose to medium) was used in Eclipse. Sixty targets with diameters ranging from 6 to 33 mm were included. Agreement between the dose distributions given by Eclipse and PRIMO was evaluated in terms of 3D global gamma passing rates (GPRs) for the 2 %/2 mm criteria. RESULTS: Average GPR greater than 95 % with the 2 %(G)/1 mm criteria were obtained over the PDD and profiles of each field size. Differences between PRIMO calculated and measured OFs were within 0.5 % in all fields, except for the 1 × 1 cm2 with a discrepancy of 1.5 %. Regarding the MLC modeling in PRIMO, an agreement within 3 % was achieved between calculated and experimental doses. Excellent agreement between PRIMO and Eclipse was found for the 35 HA plans. The 3D global GPRs (2 %/2 mm) for the targets and external patient contour were 99.6 % ± 1.1 % and 99.8 % ± 0.5 %, respectively. CONCLUSIONS: According to the results described in this study, the PRIMO Monte Carlo software, in conjunction with the 6X FFF Varian phase-space files, can be used as secondary dose calculation software to check stereotactic radiosurgery plans from Eclipse using the HyperArc technique.


Assuntos
Radiocirurgia , Humanos , Método de Monte Carlo , Aceleradores de Partículas , Imagens de Fantasmas , Radiometria , Radiocirurgia/métodos , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador/métodos , Água
5.
Pract Radiat Oncol ; 12(2): e144-e152, 2022.
Artigo em Inglês | MEDLINE | ID: mdl-34670139

RESUMO

PURPOSE: This study aimed to describe the feasibility of the online adaptive radiation therapy (oART) method developed at the Hospital Quirónsalud Barcelona for prostate cancer, using a standard C-arm linear accelerator (linac) and without the support of artificial intelligence. METHODS AND MATERIALS: The first 18 patients treated at the Hospital Quirónsalud Barcelona with the developed oART method were included. An ultrahypofractionated radiation therapy scheme consisting of 7 × 6.1 Gy was used. Patients were treated on 2 conventional Varian C-arm linacs. For each patient, a reference plan based on a planning computed tomography (pCT) scan was generated using the Eclipse system. On each treatment session, the pCT scan was rigidly registered with the daily cone beam computed tomography (CT) scan. The pCT-based target (prostate) and organs at risk were mapped onto the cone beam CT images and manually adapted to take into account the anatomy of the day. The reference plan was then copied to the cone beam CT scan, and a full reoptimization was done for the current anatomy (adapted plan). For each treatment session, the unaltered reference plan was recomputed on the daily cone beam CT scan by mimicking the soft-tissue alignment performed per our standard procedure (nonadapted plan). Over the 126 adapted sessions from the 18 patients, a dosimetric comparison of adapted against nonadapted plans was done. RESULTS: A significant difference in the target coverage was found between the adapted and nonadapted plans (97.1 vs 90.4; P < .001) in favor of adapting. Without online adaptation, the optimal coverage of the prostate was not attained in 35% of fractions. Adapting allows for the improvement of the target coverage with compliance of all organ-at-risk dose constraints in all treatment fractions. CONCLUSIONS: The oART technique described in this study is technically feasible with a C-arm linac. To our knowledge, this is the first clinical experience with oART for prostate cancer including full replanning and delivered with a C-arm linac without artificial intelligence capability.


Assuntos
Neoplasias da Próstata , Radioterapia de Intensidade Modulada , Inteligência Artificial , Humanos , Masculino , Próstata/diagnóstico por imagem , Neoplasias da Próstata/diagnóstico por imagem , Neoplasias da Próstata/radioterapia , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador/métodos , Radioterapia de Intensidade Modulada/métodos