Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros

Bases de dados
Ano de publicação
Tipo de documento
País de afiliação
Intervalo de ano de publicação
1.
J Comput Assist Tomogr ; 26(6): 1013-8, 2002.
Artigo em Inglês | MEDLINE | ID: mdl-12488752

RESUMO

PURPOSE: To determine the importance of polyp size, orientation to the scan plane, collimation, scanner type (single or multislice helical), and radiation dose on computed tomography (CT) colonography computer-aided detection. MATERIALS AND METHODS: Eight tissue-equivalent simulated polyps were placed into the interior of an air-filled acrylic tube placed within a water-filled box. Their sizes, expressed by diameter and height in millimeters, were 10 x 10, 10 x 7, 10 x 5, 10 x 3, 7 x 7, 7 x 5, 7 x 3, and 5 x 5. Detection of the polyps was performed by applying our prototype automated polyp detector software to 48 CT colonography data sets of the phantom acquired with different CT scanner settings. RESULTS: We detected at least six of the eight polyps in 47 of 48 experiments. The two most frequently undetected polyps (7 x 7 and 5 x 5) had extreme eccentricity (their height was twice the radius of the base) and were most commonly missed for 90 degrees tube orientation, 5-mm collimation, and high table speed. False-positive detections occurred in only 5 of 48 experiments. CONCLUSION: Clinically significant 10-mm polyps can be detected with 100% sensitivity in all orientations, doses, collimations, and modes that we examined.


Assuntos
Pólipos do Colo/diagnóstico por imagem , Processamento de Imagem Assistida por Computador , Tomografia Computadorizada por Raios X/métodos , Artefatos , Pólipos do Colo/patologia , Humanos , Imageamento Tridimensional , Sensibilidade e Especificidade
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA