Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros

Bases de dados
Ano de publicação
Tipo de documento
Assunto da revista
País de afiliação
Intervalo de ano de publicação
1.
Molecules ; 26(2)2021 Jan 08.
Artigo em Inglês | MEDLINE | ID: mdl-33430078

RESUMO

This work presents the long-term determination accuracy study of ICP-QMS for rare earth elements (REEs) in geological matrices. Following high-pressure closed acidic decomposition, REEs are measured repetitively across seven months by ICP-QMS. Under optimum experimental conditions (including spray chamber temperature, gas flow rate, sampling depth, etc.), the REE contents in geological standard materials from basic (basalt BCR-2 and BE-N) to intermediate (andesite AGV-2) and up to acidic (granite GSR-1) show good agreement with the certified values, giving relative errors below 10%. Here, the influence of two storage materials (perfluoroalkoxy PFA and polypropylene PP) on the long-term determination accuracy of REEs has also been monitored. It is found that the relative errors of REEs using a PFA container range from -6.6 to 6.3% (RSDs < 6.0%), while that using a PP container are within -4.0 to 3.9% (RSDs < 4.6%). By using PP material as a solution storage container, the accuracy of REEs quantification in a series of real geological samples are checked, showing the RSDs of less than 5.0%. This work first clarifies the long-term stability of REEs quantification by ICP-QMS covering two types of storage materials, confirming the reasonability of PP material as a daily storage container in terms of higher data precision and lower cost.


Assuntos
Metais Terras Raras/química , Gases em Plasma/química , Espectrometria de Massas , Análise Espectral
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA