1.
Phys Eng Sci Med ; 2024 Feb 29.
Artigo em Inglês | MEDLINE | ID: mdl-38421582

RESUMO

Workforce modelling for Radiation Oncology Medical Physicists (ROMPs) is evolving and challenging, prompting the development of the 2021 Australasian College of Physical Scientists and Engineers in Medicine (ACPSEM) ROMP Workforce (ARW) Model. In the exploration of this model at Sir Charles Gairdner Hospital, a comprehensive productivity exercise was conducted to obtain a detailed breakdown of ROMP time at a granular level. The results provide valuable insights into ROMP activities and enabled an evaluation of ARW Model calculations. The findings also capture the changing ROMP role as evidenced by an increasing involvement in consultation and advisory tasks with other professionals in the field. They also suggest that CyberKnife QA time requirements in the data utilised by the model may need to be revised. This study emphasises features inherent in the model, that need to be understood if the model is to be applied correctly.

2.
Phys Eng Sci Med ; 47(2): 455-463, 2024 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-38285271

RESUMO

Single plan techniques for multiple brain targets (MBT) stereotactic radiosurgery (SRS) are now routine. Patient specific quality assurance (QA) for MBT poses challenges due to the limited capabilities of existing QA tools which necessitates several plan redeliveries. This study sought to develop an SRS QA phantom that enables flexible MBT patient specific QA in a single delivery, along with complex SRS commissioning. PLA marble and PLA StoneFil materials were selected based on the literature and previous research conducted in our department. The HU numbers were investigated to determine the appropriate percentage infill for skull and soft-tissue equivalence. A Prusa MK3S printer in conjunction with the above-mentioned filaments were used to print the SRS QA phantom. Quality control (QC) was performed on the printed skull, film inserts and plugs for point dose measurements. EBT3 film and point dose measurements were performed using a CC04 ionisation chamber. QC demonstrated that the SRS QA phantom transverse, coronal and sagittal film planes were orthogonal within 0.5°. HU numbers for the skull, film inserts and plugs were 858 ± 20 and 35 ± 12 respectively. Point and EBT3 film dose measurements were within 2.5% and 3%/2 mm 95% gamma pass rate, respectively except one Gross Tumour Volume (GTV) that had a slightly lower gamma pass rate. Dose distributions to five GTVs were measured with EBT3 film in a single plan delivery on CyberKnife. In conclusion, an SRS QA phantom was designed, and 3D printed and its use for performing complex MBT patient specific QA in a single delivery was demonstrated.


Assuntos
Encéfalo , Imagens de Fantasmas , Impressão Tridimensional , Garantia da Qualidade dos Cuidados de Saúde , Radiocirurgia , Radiocirurgia/instrumentação , Humanos , Encéfalo/cirurgia , Encéfalo/diagnóstico por imagem , Garantia da Qualidade dos Cuidados de Saúde/normas , Controle de Qualidade , Planejamento da Radioterapia Assistida por Computador , Dosagem Radioterapêutica
3.
PLoS One ; 19(8): e0307088, 2024.
Artigo em Inglês | MEDLINE | ID: mdl-39121064

RESUMO

BACKGROUND: GammaKnife (GK) and CyberKnife (CK) have been the mainstay stereotactic radiosurgery (SRS) solution for multiple brain metastases (MBM) for several years. Recent technological advancement has seen an increase in single-isocentre C-arm linac-based SRS. This systematic review focuses on dosimetric and geometric insights into contemporary MBM SRS and thereby establish if linac-based SRS has matured to match the mainstay SRS delivery systems. METHODS: The PubMed, Web of Science and Scopus databases were interrogated which yielded 891 relevant articles that narrowed to 20 articles after removing duplicates and applying the inclusion and exclusion criteria. Primary studies which reported the use of SRS for treatment of MBM SRS and reported the technical aspects including dosimetry were included. The review was limited to English language publications from January 2015 to August 2023. Only full-length papers were included in the final analysis. Opinion papers, commentary pieces, letters to the editor, abstracts, conference proceedings and editorials were excluded. The Preferred Reporting Items for Systematic Reviews and Meta-Analyses guidelines were followed. The reporting of conformity indices (CI) and gradient indices, V12Gy, monitor units and the impact of translational and rotational shifts were extracted and analysed. RESULTS: The single-isocentre technique for MBM dominated recent SRS studies and the most studied delivery platforms were Varian. The C-arm linac-based SRS plan quality and normal brain tissue sparing was comparable to GK and CK and in some cases better. The most used nominal beam energy was 6FFF, and optimised couch and collimator angles could reduce mean normal brain dose by 11.3%. Reduction in volume of the healthy brain receiving a certain dose was dependent on the number and size of the metastases and the relative geometric location. GK and CK required 4.5-8.4 times treatment time compared with linac-based SRS. Rotational shifts caused larger changes in CI in C-arm linac-based single-isocentre SRS. CONCLUSION: C-arm linac-based SRS produced comparable MBM plan quality and the delivery is notably shorter compared to GK and CK SRS.


Assuntos
Neoplasias Encefálicas , Radiocirurgia , Radiocirurgia/métodos , Humanos , Neoplasias Encefálicas/secundário , Neoplasias Encefálicas/radioterapia , Neoplasias Encefálicas/cirurgia , Radiometria/métodos , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador/métodos