RESUMO
PURPOSE: The influence of the Integral Quality Monitor (IQM) transmission detector on photon beam properties was evaluated in a preclinical phase, using data from nine participating centres: (i) the change of beam quality (beam hardening), (ii) the influence on surface dose, and (iii) the attenuation of the IQM detector. METHODS: For 6 different nominal photon energies (4 standard, 2 FFF) and square field sizes from 1×1cm2 to 20×20cm2, the effect of IQM on beam quality was assessed from the PDD20,10 values obtained from the percentage dose depth (PDD) curves, measured with and without IQM in the beam path. The change in surface dose with/without IQM was assessed for all available energies and field sizes from 4×4cm2 to 20×20cm2. The transmission factor was calculated by means of measured absorbed dose at 10cm depth for all available energies and field sizes. RESULTS: (i) A small (0.11-0.53%) yet statistically significant beam hardening effect was observed, depending on photon beam energy. (ii) The increase in surface dose correlated with field size (p<0.01) for all photon energies except for 18MV. The change in surface dose was smaller than 3.3% in all cases except for the 20×20cm2 field and 10MV FFF beam, where it reached 8.1%. (iii) For standard beams, transmission of the IQM showed a weak dependence on the field size, and a pronounced dependence on the beam energy (0.9412 for 6MV to 0.9578 for 18MV and 0.9440 for 6MV FFF; 0.9533 for 10MV FFF). CONCLUSIONS: The effects of the IQM detector on photon beam properties were found to be small yet statistically significant. The magnitudes of changes which were found justify treating IQM either as tray factors within the treatment planning system (TPS) for a particular energy or alternatively as modified outputs for specific beam energy of linear accelerators, which eases the introduction of the IQM into clinical practice.
Assuntos
Aceleradores de Partículas/normas , Fótons/uso terapêutico , Garantia da Qualidade dos Cuidados de Saúde , Protocolos Clínicos , Eletrodos , Humanos , Imagens de Fantasmas , RadiometriaRESUMO
BACKGROUND AND PURPOSE: IMRT necessitates extension of existing inter-centre quality assurance programs due to its increased complexity. We assessed the feasibility of an inter-centre verification method for different IMRT techniques. MATERIALS AND METHODS: Eight European radiotherapy institutions of the QUASIMODO network, have designed an IMRT plan for a horseshoe-shaped PTV surrounding a cylindrical OAR in a simplified pelvic phantom. All centres applied common plan objectives but used their own equipment for planning and delivery. They verified the delivery of this plan according to a common protocol with radiographic film and ionisation chamber measurements. The irradiated films, the results of the ionisation chamber measurements and the computed dose distributions were sent to one analysis centre that compared the measured and computed dose distributions with the gamma method and composite dose-area histograms. RESULTS: 4% (relative to the prescribed dose) and 3mm (distance-to-agreement) were decided feasible gamma criteria. The composite dose-area histograms showed a maximum local deviation of 3.5% in the mean dose of the PTV and 5% in the OAR. Systematic differences could be identified, and in some cases explained. CONCLUSIONS: This multi-centre dosimetric verification study demonstrated both the feasibility of a multi-centre quality assurance network to evaluate any IMRT planning and delivery system combination, as well as the validity of the methodology involved.
Assuntos
Redes Comunitárias , Garantia da Qualidade dos Cuidados de Saúde/organização & administração , Radioterapia/normas , Fracionamento da Dose de Radiação , Humanos , Modelos Teóricos , Neoplasias/radioterapia , Radiometria , Reprodutibilidade dos Testes , Interface Usuário-ComputadorRESUMO
BACKGROUND AND PURPOSE: The purpose of this work was a comparison of realistic IMRT plans based on the same CT-image data set and a common predefined set of dose objectives for the planning target volume and the organs at risk. This work was part of the larger European QUASIMODO IMRT verification project. MATERIALS AND METHODS: Eleven IMRT plans were produced by nine different European groups, each applying a representative set of clinically used IMRT treatment planning systems. The plans produced were to be deliverable in a clinically acceptable treatment time with the local technical equipment. All plans were characterized using a set of different quality measures such as dose-volume histograms, number of monitor units and treatment time. RESULTS: Only one plan was able to fulfil all dose objectives strictly; six plans failed some of the objectives but were still considered to be clinically acceptable; four plans were not able to reach the objectives. Additional quality scores such as the number of monitor units and treatment time showed large variations, which mainly depend on the delivery technique. CONCLUSION: The presented planning study showed that with nearly all presently available IMRT planning and delivery systems comparable dose distributions could be achieved if the planning goals are clearly defined in advance.
Assuntos
Garantia da Qualidade dos Cuidados de Saúde , Radioterapia/métodos , Radioterapia/normas , Antropometria , Fracionamento da Dose de Radiação , Europa (Continente) , Humanos , Neoplasias/radioterapia , Radiometria , Reprodutibilidade dos TestesRESUMO
The dose rate due to activity induced by (gamma, n) reactions around an ELEKTA Precise accelerator running at 18 MV is reported. A model to calculate the induced dose rate for a variety of working practices has been derived and compared to the measured values. From this model, the dose received by the staff using the machine can be estimated. From measured dose rates at the face of the linear accelerator for a 10 x 10 cm2 jaw setting at 18 MV an activation coefficient per MU was derived for each of the major activation products. The relative dose rates at points around the linac head, for different energy and jaw settings, were measured. Dose rates adjacent to the patient support system and portal imager were also measured. A model to calculate the dose rate at these points was derived, and compared to those measured over a typical working week. The model was then used to estimate the maximum dose to therapists for the current working schedule on this machine. Calculated dose rates at the linac face agreed to within +/- 12% of those measured over a week, with a typical dose rate of 4.5 microSv h(-1) 2 min after the beam has stopped. The estimated maximum annual whole body dose for a treatment therapist, with the machine treating at only 18 MV, for 60000 MUs per week was 2.5 mSv. This compares well with value of 2.9 mSv published for a Clinac 21EX. A model has been derived to calculate the dose from the four dominant activation products of an ELEKTA Precise 18 MV linear accelerator. This model is a useful tool to calculate the induced dose rate around the treatment head. The model can be used to estimate the dose to the staff for typical working patterns.