Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros

Bases de dados
Ano de publicação
Tipo de documento
País de afiliação
Intervalo de ano de publicação
1.
J Clin Microbiol ; 61(11): e0085923, 2023 11 21.
Artigo em Inglês | MEDLINE | ID: mdl-37921475

RESUMO

The diagnosis of acute gastroenteritis is an ongoing clinical challenge in terms of identification of the etiologic agent, time to results, and appropriate treatment. Rapid detection of gastrointestinal pathogens is needed to improve patient care. This study evaluates the performance of the QIAstat-Dx gastrointestinal panel (Q-GP; Investigational Use Only) compared to the Luminex xTAG gastrointestinal pathogen panel (L-GPP; US-IVD). Using 245 stool specimens, we evaluated 10 different targets including rotavirus, norovirus, Salmonella, Shigella, Campylobacter, Giardia, Cryptosporidium, Escherichia coli O157, enterotoxigenic E. coli (ETEC), and Shiga toxin-producing E. coli (STEC). For the viral targets, the percent positive agreement (PPA) for rotavirus was 100% (n = 19) and that for norovirus was 91% (20/22). For the parasitic targets, the PPA was 100% for Giardia and Cryptosporidium (n = 18 and n = 23, respectively). The PPA was 96% for Salmonella (22/23) and Campylobacter (22/23), and the PPA for Shigella was 100% (n = 23). For the E. coli targets, a PPA of 94% was achieved for STEC (32/34) and 96% for ETEC (24/25). We did not assess PPA for the E. coli O157 target as the Q-GP O157 call is stx dependent. The negative percent agreement across all targets was 99.1%. Our study suggests that QIAstat-Dx GP provides comparable results to Luminex GPP based on the analysis of targets found on both panels.


Assuntos
Campylobacter , Criptosporidiose , Cryptosporidium , Escherichia coli O157 , Gastroenterite , Norovirus , Rotavirus , Humanos , Gastroenterite/diagnóstico , Fezes
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA