Gynecol Obstet Invest ; 69(4): 264-9, 2010.
Artigo em Inglês | MEDLINE | ID: mdl-20090357

RESUMO

AIMS: To compare the performance of the four latest models of glucose meters in capillary blood glucose monitoring during pregnancy. METHODS: 208 pregnant women with gestational diabetes were recruited. Each subject had simultaneous capillary glucose monitoring by two study glucose meters and venous plasma glucose assay. The performance of four glucose meters was compared using error grid analysis (EGA) and the agreement between the meter readings and plasma glucose by Bland-Altman plot analysis. RESULTS: Elite, Advantage II and CareSens had more than 90% of readings in the acceptable target range of EGA. CareSens had the lowest mean bias by Bland-Altman analysis while Advantage II had the highest proportion of readings within 5% difference from plasma glucose. Readings from all glucose meters except Optium were not influenced by the change in maternal hematocrit levels. CONCLUSIONS: The performance of four study glucose meters appeared very similar.


Assuntos
Glicemia/análise , Diabetes Gestacional/sangue , Adulto , Capilares , Técnicas de Diagnóstico Endócrino/instrumentação , Feminino , Humanos , Monitorização Fisiológica , Gravidez , Sensibilidade e Especificidade , Veias
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA