Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros

Bases de dados
Ano de publicação
Tipo de documento
País de afiliação
Intervalo de ano de publicação
1.
Pacing Clin Electrophysiol ; 45(11): 1295-1298, 2022 11.
Artigo em Inglês | MEDLINE | ID: mdl-35687737

RESUMO

A 39-years old ventricular lead of a right-sided single-chamber pacemaker required extraction for infection. Angulation at the right subclavian-superior vena cava junction coupled with calcified fibrotic encapsulating tissue prevented advancement of a rotational dissecting sheath. To straighten the lead, it was pulled from the subclavian and out of the right internal jugular vein, whilst the Needle's-Eye Snare via the femoral access provided counter-traction. A 13-french rotational dissecting sheath was successfully advanced over the lead via the jugular access to complete the lead extraction without any complication.


Assuntos
Marca-Passo Artificial , Veia Cava Superior , Humanos , Adulto , Remoção de Dispositivo , Veias Jugulares/cirurgia , Artéria Femoral
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA