Iterative reconstructions in multiphasic CT imaging of the liver: qualitative and task-based analyses of image quality.
Pasquier, H; Gardavaud, F; Chiaradia, M; Zanca, F; Hérin, E; Mulé, S; Rahmouni, A; Luciani, A.
Affiliation
  • Pasquier H; Université Paris Est, Ecole Doctorale Sciences de la Vie et de la Santé - ED402, Créteil, F-94010, France; AP-HP, Groupe Henri Mondor Albert Chenevier, Imagerie Médicale, Créteil, F-94010, France. Electronic address: hugo.pasq@gmail.com.
  • Gardavaud F; AP-HP, Hôpital Tenon, Imagerie Médicale, Paris, F-75020, France.
  • Chiaradia M; AP-HP, Groupe Henri Mondor Albert Chenevier, Imagerie Médicale, Créteil, F-94010, France.
  • Zanca F; DoseWatch, GE Healthcare, Buc, F-78530, France.
  • Hérin E; AP-HP, Groupe Henri Mondor Albert Chenevier, Imagerie Médicale, Créteil, F-94010, France; Université Paris Est Créteil, Créteil, Faculté de Médecine, F-94010, France.
  • Mulé S; AP-HP, Groupe Henri Mondor Albert Chenevier, Imagerie Médicale, Créteil, F-94010, France.
  • Rahmouni A; AP-HP, Groupe Henri Mondor Albert Chenevier, Imagerie Médicale, Créteil, F-94010, France; Université Paris Est Créteil, Créteil, Faculté de Médecine, F-94010, France.
  • Luciani A; AP-HP, Groupe Henri Mondor Albert Chenevier, Imagerie Médicale, Créteil, F-94010, France; Université Paris Est Créteil, Créteil, Faculté de Médecine, F-94010, France; INSERM Unité U955 IMRB, Equipe 18, Créteil, F-94010, France.
Clin Radiol ; 73(9): 834.e9-834.e16, 2018 09.
Article in En | MEDLINE | ID: mdl-29929903
ABSTRACT

AIM:

To evaluate the clinical benefits on image quality (IQ) of adaptive statistical iterative reconstruction (ASIR) and model-based iterative reconstruction (MBIR) in multiphasic liver CT compared to filtered back-projection (FBP), in patients and on phantoms using a novel task-based metric. MATERIALS AND

METHODS:

Image data of 65 patients who underwent a routine multiphasic liver CT during a 1-month period were reconstructed with FBP, ASIR50, ASIR80, and MBIR. IQ was assessed qualitatively by ranking the most distal hepatic artery (HA) and portal vein (PV) visible; and quantitatively by measuring contrast-to-noise ratio (CNR) of the liver parenchyma, HA and PV. IQ was compared between each reconstruction and correlated to CNR and detectability index (d') measurements computed on phantoms scanned with the same CT protocol as for patients.

RESULTS:

HA and PV were seen more distally on MBIR and ASIR80 compared to FBP (p≤0.001). The CNR correlated weakly between patient and phantom (r=0.76 and 0.80 for HA and PV, respectively), whereas d' correlated strongly with the division order of HA and PV (r=0.96 and 0.95, respectively).

CONCLUSION:

MBIR and ASIR significantly improve the IQ of multiphasic liver CT, especially through better distal detection of HA and PV, in agreement with the adapted task-based metric d' estimated on phantoms.
Subject(s)

Full text: 1 Database: MEDLINE Main subject: Radiographic Image Interpretation, Computer-Assisted / Tomography, X-Ray Computed / Liver Type of study: Guideline / Qualitative_research Limits: Adolescent / Adult / Female / Humans / Male Language: En Journal: Clin Radiol Year: 2018 Type: Article