Recurrent HeartMate 3 Right Ventricular Assist Device Stoppages in a Biventricular Assist Device Carrier.
Gasparovic, Hrvoje; Tokic, Tomislav; Cikes, Maja; Urlic, Marjan; Kopjar, Tomislav; Cerina, Petra; Knezevic, Ivan; Perkov, Drazen; Milicic, Davor; Biocina, Bojan.
Affiliation
  • Gasparovic H; University of Zagreb School of Medicine, Department of Cardiac Surgery, University Hospital Center Zagreb, Zagreb, Croatia. hgasparovic@gmail.com.
  • Tokic T; University of Zagreb School of Medicine, Department of Cardiac Surgery, University Hospital Center Zagreb, Zagreb, Croatia. tokic.klc@hotmail.com.
  • Cikes M; University of Zagreb School of Medicine, Department of Cardiovascular Diseases, University Hospital Center Zagreb, Zagreb, Croatia. maja.cikes@gmail.com.
  • Urlic M; University of Zagreb School of Medicine, Department of Cardiac Surgery, University Hospital Center Zagreb, Zagreb, Croatia. urlic.marjan@gmail.com.
  • Kopjar T; University of Zagreb School of Medicine, Department of Cardiac Surgery, University Hospital Center Zagreb, Zagreb, Croatia. tkopjar@gmail.com.
  • Cerina P; University of Zagreb School of Medicine, Department of Cardiac Surgery, University Hospital Center Zagreb, Zagreb, Croatia. petra.cerina@hotmail.com.
  • Knezevic I; Department of Cardiac Surgery, University Hospital Center Ljubljana, Ljubljana, Slovenia. ivan.knezevic@kclj.si.
  • Perkov D; University of Zagreb School of Medicine, Department of Radiology, University Hospital Center Zagreb, Zagreb, Croatia. drazen.perkov@hevi.hr.
  • Milicic D; University of Zagreb School of Medicine, Department of Cardiovascular Diseases, University Hospital Center Zagreb, Zagreb, Croatia. davormilicic01@gmail.com.
  • Biocina B; University of Zagreb School of Medicine, Department of Cardiac Surgery, University Hospital Center Zagreb, Zagreb, Croatia. bbiocina@gmail.com.
Heart Surg Forum ; 26(1): E038-E039, 2023 Jan 31.
Article in En | MEDLINE | ID: mdl-36856506
ABSTRACT
A 44-year-old female patient with chemotherapy-induced cardiomyopathy presented with acute cardiogenic shock requiring ECMO support. Multiple failed weaning trials from temporary mechanical circulatory assistance prompted a transition to staged durable biventricular support. Her course was complicated with recurrent RVAD stoppages. The initial event was treated with pump exchange, while for the subsequent RVAD standstill, we employed a device wash-out and reimplantation strategy. A brief period of circulatory arrest was employed to explore the right-sided cardiac chambers using a single-use bronchoscope.
Subject(s)

Full text: 1 Database: MEDLINE Main subject: Heart-Assist Devices / Dextrocardia / Heart Arrest Limits: Adult / Female / Humans Language: En Journal: Heart Surg Forum Journal subject: CARDIOLOGIA Year: 2023 Type: Article Affiliation country: Croatia

Full text: 1 Database: MEDLINE Main subject: Heart-Assist Devices / Dextrocardia / Heart Arrest Limits: Adult / Female / Humans Language: En Journal: Heart Surg Forum Journal subject: CARDIOLOGIA Year: 2023 Type: Article Affiliation country: Croatia