Practical value of cardiac magnetic resonance imaging for clinical quantification of aortic valve stenosis: comparison with echocardiography.
Caruthers, Shelton D; Lin, Shiow Jiuan; Brown, Peggy; Watkins, Mary P; Williams, Todd A; Lehr, Katherine A; Wickline, Samuel A.
Afiliación
  • Caruthers SD; Cardiovascular Magnetic Resonance Laboratories, Barnes-Jewish Hospital, Cardiovascular Division, Washington University School of Medicine, St Louis, MO 63110, USA. saw@howdy.wustl.edu
Circulation ; 108(18): 2236-43, 2003 Nov 04.
Article en En | MEDLINE | ID: mdl-14568899
ABSTRACT

BACKGROUND:

Valvular pathology can be analyzed quickly and accurately through the use of Doppler ultrasound. For aortic stenosis, the continuity equation approach with Doppler velocity-time integral (VTI) data is by far the most commonly used clinical method of quantification. In view of the emerging popularity of cardiac magnetic resonance (CMR) as a routine clinical imaging tool, the purposes of this study were to define the reliability of velocity-encoded CMR as a routine method for quantifying stenotic aortic valve area, to compare this method with the accepted standard, and to evaluate its reproducibility. METHODS AND

RESULTS:

Patients (n=24) with aortic stenosis (ranging from 0.5 to 1.8 cm2) were imaged with CMR and echocardiography. Velocity-encoded CMR was used to obtain velocity information in the aorta and left ventricular outflow tract. From this flow data, pressure gradients were estimated by means of the modified Bernoulli equation, and VTIs were calculated to estimate aortic valve orifice dimensions by means of the continuity equation. The correlation coefficients between modalities for pressure gradients were r=0.83 for peak and r=0.87 for mean. The measurements of VTI correlated well, leading to an overall strong correlation between modalities for the estimation of valve dimension (r=0.83, by means of the identified best approach). For 5 patients, the CMR examination was repeated using the best approach. The repeat calculations of valve size correlated well (r=0.94).

CONCLUSIONS:

Velocity-encoded CMR can be used as a reliable, user-friendly tool to evaluate stenotic aortic valves. The measurements of pressure gradients, VTIs, and the valve dimension correlate well with the accepted standard of Doppler ultrasound.
Asunto(s)
Buscar en Google
Bases de datos: MEDLINE Asunto principal: Estenosis de la Válvula Aórtica / Imagen por Resonancia Magnética / Corazón Tipo de estudio: Clinical_trials / Diagnostic_studies / Prognostic_studies Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Circulation Año: 2003 Tipo del documento: Article País de afiliación: Estados Unidos
Buscar en Google
Bases de datos: MEDLINE Asunto principal: Estenosis de la Válvula Aórtica / Imagen por Resonancia Magnética / Corazón Tipo de estudio: Clinical_trials / Diagnostic_studies / Prognostic_studies Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Circulation Año: 2003 Tipo del documento: Article País de afiliación: Estados Unidos