Your browser doesn't support javascript.
loading
Coronary guidewire circumcision during use of a Gopher support catheter: potential adverse interaction with polymer-jacketed wire design.
Armstrong, Ehrin J; Shunk, Kendrick A.
Afiliación
  • Armstrong EJ; Department of Medicine, Division of Cardiology, University of California-San Francisco, CA, USA.
Catheter Cardiovasc Interv ; 76(1): 112-6, 2010 Jul 01.
Article en En | MEDLINE | ID: mdl-20578095
ABSTRACT
Support catheters with a threaded, distal tip may be useful for crossing recalcitrant coronary lesions. The interaction between the Gopher support catheter and polymer-coated coronary guidewires may, however, result in circumferential stripping of polymer from the guidewire, with resulting wire damage or fracture. We report two cases of polymer-coated guidewire damage with use of a Gopher support catheter, suggesting that these two devices should not be used together.
Asunto(s)

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Angioplastia Coronaria con Balón / Materiales Biocompatibles Revestidos / Estenosis Coronaria / Oclusión Coronaria Límite: Aged / Humans / Male / Middle aged Idioma: En Revista: Catheter Cardiovasc Interv Asunto de la revista: CARDIOLOGIA Año: 2010 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Angioplastia Coronaria con Balón / Materiales Biocompatibles Revestidos / Estenosis Coronaria / Oclusión Coronaria Límite: Aged / Humans / Male / Middle aged Idioma: En Revista: Catheter Cardiovasc Interv Asunto de la revista: CARDIOLOGIA Año: 2010 Tipo del documento: Article País de afiliación: Estados Unidos