Coupling desorption electrospray ionization with solid-phase microextraction for screening and quantitative analysis of drugs in urine.
Anal Chem
; 82(17): 7502-8, 2010 Sep 01.
Article
en En
| MEDLINE
| ID: mdl-20695439
Direct analysis of silica C(18)-coated solid-phase microextraction (SPME) fibers using desorption electrospray ionization mass spectrometry (DESI-MS) for the purpose of analyzing drugs from raw urine is presented. The method combines a simple, inexpensive, and solvent-less sample preparation technique with the specificity and speed of DESI-MS and MS/MS. Extraction of seven drugs from raw urine is performed using specially designed SPME fibers coated uniformly with silica-C(18) stationary phase. Each SPME device is inserted into unprocessed urine under gentle agitation and, then, removed, rinsed, and analyzed directly by DESI-MS (MS/MS). Rapid screening over a wide mass range is afforded by coupling the method with a time of flight (TOF) mass spectrometer while quantitative analysis is performed using selected reaction monitoring (SRM) using a triple quadrupole mass spectrometer. The performance of the SPME DESI-MS/MS method was evaluated by preparing calibration standards and quality control (QC) samples of the seven drug compounds from urine over a range from 20 to 1000 ng/mL, with the exception of meprobamate which was prepared from 200 to 10000 ng/mL. The calibration curves constructed for each analyte had an R(2) > 0.99. The range of precision (%CV) and accuracy values (% bias) for low QC samples was 1-11% and 3-38%, respectively. Precision and accuracy values for high QC samples range from 0.9 to 8% and -31 to -8%. Results from urine specimens of actual exposure to drugs screened using the SPME DESI-MS/MS method showed good agreement with the conventional immunoassays and GC/MS analysis. Liquid desorption of the SPME fiber followed by LC/MS/MS also showed good agreement with the SPME DESI-MS/MS method.
Texto completo:
1
Bases de datos:
MEDLINE
Asunto principal:
Drogas Ilícitas
/
Espectrometría de Masa por Ionización de Electrospray
/
Microextracción en Fase Sólida
Tipo de estudio:
Diagnostic_studies
/
Screening_studies
Límite:
Humans
Idioma:
En
Revista:
Anal Chem
Año:
2010
Tipo del documento:
Article
País de afiliación:
Estados Unidos