Copper-based electrochemical sensor with palladium electrode for cathodic stripping voltammetry of manganese.
Kang, Wenjing; Pei, Xing; Bange, Adam; Haynes, Erin N; Heineman, William R; Papautsky, Ian.
Afiliación
  • Kang W; BioMicroSystems Lab, Department of Electrical Engineering and Computing Systems, University of Cincinnati , Cincinnati, Ohio 45221-0030, United States.
Anal Chem ; 86(24): 12070-7, 2014 Dec 16.
Article en En | MEDLINE | ID: mdl-25476591
ABSTRACT
In this work, we report on the development of a palladium-based, microfabricated point-of-care electrochemical sensor for the determination of manganese using square wave cathodic stripping voltammetry. Heavy metals require careful monitoring, yet current methods are too complex for a point-of-care system. Voltammetry offers an attractive approach to metal detection on the microscale, but traditional carbon, gold, or platinum electrodes are difficult or expensive to microfabricate, preventing widespread use. Our sensor uses palladium working and auxiliary electrodes and integrates them with a copper-based reference electrode for simple fabrication and compatibility with microfabrication and printed circuit board processing, while maintaining competitive performance in electrochemical detection. Copper electrodes were prepared on glass substrate using a combination of microfabrication procedures followed by electrodeposition of palladium. The disposable sensor system was formed by bonding a poly(dimethylsiloxane) (PDMS) well to the glass substrate. Cathodic stripping voltammetry of manganese using our new disposable palladium-based sensors exhibited 334 nM (18.3 ppb) limit of detection in borate buffer. The sensor was used to demonstrate manganese determination in natural water samples from a pond in Burnet Woods, located in Cincinnati, OH, and the Ohio River.
Asunto(s)

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Paladio / Monitoreo del Ambiente / Cobre / Manganeso Idioma: En Revista: Anal Chem Año: 2014 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Paladio / Monitoreo del Ambiente / Cobre / Manganeso Idioma: En Revista: Anal Chem Año: 2014 Tipo del documento: Article País de afiliación: Estados Unidos