Quality assurance multicenter comparison of different MR scanners for quantitative diffusion-weighted imaging.
Belli, Giacomo; Busoni, Simone; Ciccarone, Antonio; Coniglio, Angela; Esposito, Marco; Giannelli, Marco; Mazzoni, Lorenzo N; Nocetti, Luca; Sghedoni, Roberto; Tarducci, Roberto; Zatelli, Giovanna; Anoja, Rosa A; Belmonte, Gina; Bertolino, Nicola; Betti, Margherita; Biagini, Cristiano; Ciarmatori, Alberto; Cretti, Fabiola; Fabbri, Emma; Fedeli, Luca; Filice, Silvano; Fulcheri, Christian P L; Gasperi, Chiara; Mangili, Paola A; Mazzocchi, Silvia; Meliadò, Gabriele; Morzenti, Sabrina; Noferini, Linhsia; Oberhofer, Nadia; Orsingher, Laura; Paruccini, Nicoletta; Princigalli, Goffredo; Quattrocchi, Mariagrazia; Rinaldi, Adele; Scelfo, Danilo; Freixas, Gloria Vilches; Tenori, Leonardo; Zucca, Ileana; Luchinat, Claudio; Gori, Cesare; Gobbi, Gianni.
Afiliación
  • Belli G; Health Physics Unit, USL 8, Arezzo, Italy.
  • Busoni S; Health Physics Unit, AOU Careggi, Florence, Italy.
  • Ciccarone A; AOU Meyer, Florence, Italy.
  • Coniglio A; Medical Physics Department, Fondazione Fatebenefratelli per la Ricerca e la Formazione sanitaria e sociale, San Giovanni Calibita Hospital, Rome, Italy.
  • Esposito M; Health Physics Unit, ASF, Florence, Italy.
  • Giannelli M; Health Physics Unit, AOU Pisana, Pisa, Italy.
  • Mazzoni LN; Health Physics Unit, AOU Senese, Siena, Italy.
  • Nocetti L; Medical Phisics Department, AOU Policlinico Modena, Italy.
  • Sghedoni R; Medical Physics Unit, Arcispedale Santa Maria Nuova - IRCCS, Reggio Emilia, Italy.
  • Tarducci R; Health Physics Unit, AOU Perugia, Italy.
  • Zatelli G; Health Physics Unit, ASF, Florence, Italy.
  • Anoja RA; A.O."Pugliese-Ciaccio," Catanzaro, Italy.
  • Belmonte G; Health Physics Unit, AOU Senese, Siena, Italy.
  • Bertolino N; UO Direzione Sanitaria, IRCCS Istituto Neurologico Carlo Besta, Milan, Italy.
  • Betti M; Health Physics Unit, Centro Oncologico Fiorentino, Sesto Fiorentino, Italy.
  • Biagini C; Radiodiagnostic Unit, Centro Oncologico Fiorentino, Sesto Fiorentino, Italy.
  • Ciarmatori A; Medical Phisics Department, AOU Policlinico Modena, Italy.
  • Cretti F; Azienda Ospedaliera Papa Giovanni XXIII, Bergamo, Italy.
  • Fabbri E; Health Physics Unit, Ospedale Sant'Orsola-Malpighi, Bologna, Italy.
  • Fedeli L; Health Physics Unit, AOU Careggi, Florence, Italy.
  • Filice S; Health Physics Unit, AOU Parma, Italy.
  • Fulcheri CP; Health Physics Unit, AOU Perugia, Italy.
  • Gasperi C; Health Physics Unit, USL 8, Arezzo, Italy.
  • Mangili PA; Health Physics Unit, IRCCS-San Raffaele, Milan, Italy.
  • Mazzocchi S; Health Physics Unit, ASF, Florence, Italy.
  • Meliadò G; Health Physics Unit, AOU Verona, Italy.
  • Morzenti S; Health Physics Unit, AO San Gerardo, Monza, Italy.
  • Noferini L; Health Physics Unit, AOU Careggi, Florence, Italy.
  • Oberhofer N; Health Physics Unit, Azienda Sanitaria dell'Alto Adige-Ospedale Bolzano, Italy.
  • Orsingher L; Medical Physics Unit, Arcispedale Santa Maria Nuova - IRCCS, Reggio Emilia, Italy.
  • Paruccini N; Health Physics Unit, AO San Gerardo, Monza, Italy.
  • Princigalli G; FISMECO SRL, Rome, Italy.
  • Quattrocchi M; Health Physics Unit, ASL 2-Ospedale Campo di Marte, Lucca, Italy.
  • Rinaldi A; Medical Physics Department, Fondazione Fatebenefratelli per la Ricerca e la Formazione sanitaria e sociale, San Giovanni Calibita Hospital, Rome, Italy.
  • Scelfo D; Department of Developmental Neuroscience, Stella Maris Scientific Institute, Pisa, Italy.
  • Freixas GV; CNAO, Pavia, Italy.
  • Tenori L; Magnetic Resonance Center (CERM), University of Florence, Sesto Fiorentino, Italy; FiorGen Foundation, Sesto Fiorentino, Italy.
  • Zucca I; UO Direzione Scientifica, IRCCS Istituto Neurologico Carlo Besta, Milan, Italy.
  • Luchinat C; Magnetic Resonance Center (CERM), University of Florence, Sesto Fiorentino, Italy; FiorGen Foundation, Sesto Fiorentino, Italy.
  • Gori C; Health Physics Unit, AOU Careggi, Florence, Italy.
  • Gobbi G; Health Physics Unit, AOU Perugia, Italy.
J Magn Reson Imaging ; 43(1): 213-9, 2016 Jan.
Article en En | MEDLINE | ID: mdl-26013043
ABSTRACT

PURPOSE:

To propose a magnetic resonance imaging (MRI) quality assurance procedure that can be used for multicenter comparison of different MR scanners for quantitative diffusion-weighted imaging (DWI). MATERIALS AND

METHODS:

Twenty-six centers (35 MR scanners with field strengths 1T, 1.5T, and 3T) were enrolled in the study. Two different DWI acquisition series (b-value ranges 0-1000 and 0-3000 s/mm(2) , respectively) were performed for each MR scanner. All DWI acquisitions were performed by using a cylindrical doped water phantom. Mean apparent diffusion coefficient (ADC) values as well as ADC values along each of the three main orthogonal directions of the diffusion gradients (x, y, and z) were calculated. Short-term repeatability of ADC measurement was evaluated for 26 MR scanners.

RESULTS:

A good agreement was found between the nominal and measured mean ADC over all the centers. More than 80% of mean ADC measurements were within 5% from the nominal value, and the highest deviation and overall standard deviation were 9.3% and 3.5%, respectively. Short-term repeatability of ADC measurement was found <2.5% for all MR scanners.

CONCLUSION:

A specific and widely accepted protocol for quality controls in DWI is still lacking. The DWI quality assurance protocol proposed in this study can be applied in order to assess the reliability of DWI-derived indices before tackling single- as well as multicenter studies.
Asunto(s)
Palabras clave

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Garantía de la Calidad de Atención de Salud / Interpretación de Imagen Asistida por Computador / Imagen de Difusión por Resonancia Magnética Tipo de estudio: Clinical_trials / Diagnostic_studies / Evaluation_studies País/Región como asunto: Europa Idioma: En Revista: J Magn Reson Imaging Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2016 Tipo del documento: Article País de afiliación: Italia

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Garantía de la Calidad de Atención de Salud / Interpretación de Imagen Asistida por Computador / Imagen de Difusión por Resonancia Magnética Tipo de estudio: Clinical_trials / Diagnostic_studies / Evaluation_studies País/Región como asunto: Europa Idioma: En Revista: J Magn Reson Imaging Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2016 Tipo del documento: Article País de afiliación: Italia