Your browser doesn't support javascript.
loading
Local SAR, global SAR, and power-constrained large-flip-angle pulses with optimal control and virtual observation points.
Vinding, Mads S; Guérin, Bastien; Vosegaard, Thomas; Nielsen, Niels Chr.
Afiliación
  • Vinding MS; Center of Insoluble Protein Structures (inSPIN), Interdisciplinary Nanoscience Center (iNANO) and Department of Chemistry, Aarhus University, Aarhus C, Denmark.
  • Guérin B; A.A. Martinos Center for Biomedical Imaging, Department of Radiology, Massachusetts General Hospital, Charlestown, Massachusetts, USA.
  • Vosegaard T; Harvard Medical School, Boston, Massachusetts, USA.
  • Nielsen NC; Center of Insoluble Protein Structures (inSPIN), Interdisciplinary Nanoscience Center (iNANO) and Department of Chemistry, Aarhus University, Aarhus C, Denmark.
Magn Reson Med ; 77(1): 374-384, 2017 01.
Article en En | MEDLINE | ID: mdl-26715084
ABSTRACT

PURPOSE:

To present a constrained optimal-control (OC) framework for designing large-flip-angle parallel-transmit (pTx) pulses satisfying hardware peak-power as well as regulatory local and global specific-absorption-rate (SAR) limits. The application is 2D and 3D spatial-selective 90° and 180° pulses. THEORY AND

METHODS:

The OC gradient-ascent-pulse-engineering method with exact gradients and the limited-memory Broyden-Fletcher-Goldfarb-Shanno method is proposed. Local SAR is constrained by the virtual-observation-points method. Two numerical models facilitated the optimizations, a torso at 3 T and a head at 7 T, both in eight-channel pTx coils and acceleration-factors up to 4.

RESULTS:

The proposed approach yielded excellent flip-angle distributions. Enforcing the local-SAR constraint, as opposed to peak power alone, reduced the local SAR 7 and 5-fold with the 2D torso excitation and inversion pulse, respectively. The root-mean-square errors of the magnetization profiles increased less than 5% with the acceleration factor of 4.

CONCLUSION:

A local and global SAR, and peak-power constrained OC large-flip-angle pTx pulse design was presented, and numerically validated for 2D and 3D spatial-selective 90° and 180° pulses at 3 T and 7 T. Magn Reson Med 77374-384, 2017. © 2015 Wiley Periodicals, Inc.
Asunto(s)
Palabras clave

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética Tipo de estudio: Prognostic_studies Límite: Adult / Humans / Male Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2017 Tipo del documento: Article País de afiliación: Dinamarca

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Imagen por Resonancia Magnética Tipo de estudio: Prognostic_studies Límite: Adult / Humans / Male Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2017 Tipo del documento: Article País de afiliación: Dinamarca