Does motion affect liver stiffness estimates in shear wave elastography? Phantom and clinical study.
Pellot-Barakat, Claire; Chami, Linda; Correas, Jean Michel; Lefort, Muriel; Lucidarme, Olivier.
Afiliación
  • Pellot-Barakat C; Laboratoire Imagerie Moléculaire In Vivo (IMIV), UMR 1023 Inserm/CEA/Université Paris Sud - ERL 9218 CNRS, CEA/I2BM/SHFJ, Orsay, France. Electronic address: claire.barakat@inserm.fr.
  • Chami L; Radiology department, Hôpital Pitié-Salpétrière, APHP, Paris, France; Sorbonne Universités, UPMC Univ Paris 06, CNRS, INSERM, Laboratoire d'Imagerie Biomédicale (LIB), 75006 (ou 75013), Paris, France. Electronic address: linda.chami@yahoo.com.
  • Correas JM; Radiology department, Hôpital Necker, APHP, Paris, France. Electronic address: jean-michel.correas@nck.aphp.fr.
  • Lefort M; Sorbonne Universités, UPMC Univ Paris 06, CNRS, INSERM, Laboratoire d'Imagerie Biomédicale (LIB), 75006 (ou 75013), Paris, France. Electronic address: Muriel.Lefort@lib.upmc.fr.
  • Lucidarme O; Radiology department, Hôpital Pitié-Salpétrière, APHP, Paris, France; Sorbonne Universités, UPMC Univ Paris 06, CNRS, INSERM, Laboratoire d'Imagerie Biomédicale (LIB), 75006 (ou 75013), Paris, France. Electronic address: olivier.lucidarme@psl.aphp.fr.
Eur J Radiol ; 85(9): 1645-50, 2016 Sep.
Article en En | MEDLINE | ID: mdl-27501901
ABSTRACT
This study was undertaken to evaluate the impact of free-breathing (FB) vs. Apnea on Shear-wave elastography (SWE) measurements. Quantitative liver-stiffness measurements were obtained during FB and Apnea for 97 patients with various body-morphologies and liver textures. Quality indexes of FB and Apnea elasticity maps (percentage of non-filling (PNF), temporal (TV) and spatial (SV) variabilities) were computed. SWE measurements were also obtained from an homogeneous phantom at rest and during a mechanically-induced motion. Liver-stiffness values estimated from FB and Apnea acquisitions were correlated, particularly for homogeneous livers (r=0.76, P<0.001) and favorable body-morphologies (r=0.68, P<0.001). However FB values were consistently 20-25% lower than Apnea ones (P<0.001). FB also systematically resulted in degradation of TV (P<0.005) and PNF (P<0.001) compared to Apnea but had no impact on SV. With the phantom, no differences between SWE measurements at rest and during motion were observed. Apnea and FB measurements are highly correlated, although FB data quality is degraded compared to Apnea and estimated stiffness in FB is systematically lower than in Apnea. These discrepancies between rest and motion states were observed for patients but not for phantom data, suggesting that patient breath-holding impacts liver stiffness.
Asunto(s)
Palabras clave

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Interpretación de Imagen Asistida por Computador / Diagnóstico por Imagen de Elasticidad / Cirrosis Hepática Tipo de estudio: Diagnostic_studies Límite: Female / Humans / Male / Middle aged País/Región como asunto: Europa Idioma: En Revista: Eur J Radiol Año: 2016 Tipo del documento: Article

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Interpretación de Imagen Asistida por Computador / Diagnóstico por Imagen de Elasticidad / Cirrosis Hepática Tipo de estudio: Diagnostic_studies Límite: Female / Humans / Male / Middle aged País/Región como asunto: Europa Idioma: En Revista: Eur J Radiol Año: 2016 Tipo del documento: Article