Factors Affecting Stereotactic Accuracy in Image-Guided Deep Brain Stimulator Electrode Placement.
Ko, Andrew L; Ibrahim, Aly; Magown, Philippe; Macallum, Rebecca; Burchiel, Kim J.
Afiliación
  • Ko AL; Department of Neurological Surgery, University of Washington, Seattle, WA, USA.
Stereotact Funct Neurosurg ; 95(5): 315-324, 2017.
Article en En | MEDLINE | ID: mdl-28889124
ABSTRACT
BACKGROUND/

AIMS:

Intraoperative imaging allows near-real-time assessment of stereotactic accuracy during implantation of deep brain stimulation (DBS) electrodes. Such technology can be used to examine factors impacting stereotactic error.

METHODS:

Intraoperative CT imaging was reviewed in patients undergoing DBS placement at Oregon Health and Sciences University. Coordinates of the target electrode were compared to the operative plan to characterize the magnitude and direction of stereotactic error with respect to side of implantation, target, and electrode approach angles.

RESULTS:

One hundred sixty-nine leads in 94 patients were examined. Targets were GPi (n = 86), STN (n = 31), and Vim (n = 52). The average Euclidean error was 1.63 mm (SD 0.87). The error magnitude was higher for Vim (1.95 mm) than for GPi (1.44 mm), while STN (1.65 mm) did not differ from either Vim or GPi (ANOVA F = 6.15, p = 0.003). Electrodes targeting Vim and STN were significantly more likely to deviate medially compared to those targeting GPi (ANOVA F = 9.13, p < 0.001). The coronal approach angle affected the error when targeting Vim (ρ = 0.338, p = 0.01). These findings were confirmed during multivariate analyses.

CONCLUSIONS:

This study shows a significant effect of target on the accuracy of electrode placement for DBS. Targeting Vim results in a greater Euclidean error and a greater medial deviation off target. These systematic deviations should be taken into account during electrode implantation.
Asunto(s)
Palabras clave

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Enfermedad de Parkinson / Técnicas Estereotáxicas / Estimulación Encefálica Profunda / Electrodos Implantados Tipo de estudio: Observational_studies Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Stereotact Funct Neurosurg Asunto de la revista: NEUROCIRURGIA Año: 2017 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Enfermedad de Parkinson / Técnicas Estereotáxicas / Estimulación Encefálica Profunda / Electrodos Implantados Tipo de estudio: Observational_studies Límite: Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Stereotact Funct Neurosurg Asunto de la revista: NEUROCIRURGIA Año: 2017 Tipo del documento: Article País de afiliación: Estados Unidos