[Evaluation of the accuracy of volume navigation of sonography and computed tomography using a phantom]. / Genauigkeitsevaluierung der Volumennavigation von Sonografie und Computertomografie anhand eines Phantoms.
Moritz, Anne-Kristin; Winter, Karsten; Köhler, Claudia; Alef, Michaele; Kiefer, Ingmar.
Afiliación
  • Moritz AK; Klinik für Kleintiere, Veterinärmedizinische Fakultät der Universität Leipzig.
  • Winter K; Institut für Anatomie, Medizinische Fakultät der Universität Leipzig.
  • Köhler C; Klinik für Kleintiere, Veterinärmedizinische Fakultät der Universität Leipzig.
  • Alef M; Klinik für Kleintiere, Veterinärmedizinische Fakultät der Universität Leipzig.
  • Kiefer I; Klinik für Kleintiere, Veterinärmedizinische Fakultät der Universität Leipzig.
Article en De | MEDLINE | ID: mdl-31627223
ABSTRACT
BACKGROUND AND

OBJECTIVE:

In case of superimpositions of gas in the gastrointestinal tract or the ribs, tissue changes well detectable on computed tomography (CT) cannot be identified sonographically in a number of cases. Combining ultrasonography and CT provides enhanced information compared to sole sonography and volume navigation may be used as an effective tool. Tissue samples easily and safely obtained under sonographic guidance are often necessary to confirm the diagnosis of a suspicious focus. In these cases, the spatial fusion of CT and sonography may also be employed for improved visualization of foci by eliminating superimposition of sonographic images which is a limitation of ultrasound. This study investigated the potential benefit and improved informative value of the fusion of CT and sonography in case of superimpositions and aimed at determining the registration method with the best accuracy. MATERIALS AND

METHODS:

Sixteen models (10 models with peas [low contrast], 6 models with wooden spheres [high contrast] as round structures) were created. These models were examined by computed tomography and fused using 3 volume navigation protocols. Subsequently, volume-guided sonography was performed. The deviation of the specimens was measured.

RESULTS:

In total, 1026 measurements of the pea models and 648 measurements of the wooden sphere models were carried out. A fusion accuracy of 100 % was observed in 9.9 % (102/1026) resp. 9.9 % (64/648) of the models. In 85.4 % (876/1026) resp. 94.1 % (610/648) the deviation was < 5 mm and in 98.1 % (1006/1026) resp. 99.4 % (644/648) it was < 10 mm. The registration protocol in which all reference points were used for spatial fusion proved to be the most accurate

CONCLUSION:

The registration protocols for volume-guided ultrasound have sufficient biopsy accuracy to merge identical sites and provide the basis for improved volume-navigated biopsy sampling.
Asunto(s)

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Tomografía Computarizada por Rayos X / Ultrasonografía / Fantasmas de Imagen Tipo de estudio: Diagnostic_studies / Guideline / Prognostic_studies Idioma: De Revista: Tierarztl Prax Ausg K Kleintiere Heimtiere Asunto de la revista: MEDICINA VETERINARIA Año: 2019 Tipo del documento: Article

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Tomografía Computarizada por Rayos X / Ultrasonografía / Fantasmas de Imagen Tipo de estudio: Diagnostic_studies / Guideline / Prognostic_studies Idioma: De Revista: Tierarztl Prax Ausg K Kleintiere Heimtiere Asunto de la revista: MEDICINA VETERINARIA Año: 2019 Tipo del documento: Article