Comparison of the New, Rapid, and Fully Automated Kryptor TSH receptor Antibodies Assay (B.R.A.H.M.S.) with the Radioimmunological Assay (B.R.A.H.M.S.).
Clin Lab ; 65(11)2019 Nov 01.
Article en En | MEDLINE | ID: mdl-31710438
ABSTRACT

BACKGROUND:

Radioimmunoassays, which are often not automated and time-consuming, are gradually being re-placed in medical laboratories by non-radioactive methods that need to be evaluated. The purpose was to compare the measurement of thyroid-stimulating hormone receptor antibodies (TRAb) by the new Brahms' kit using Kryptor TRACE technology and the Brahms' radioimmunoassay.

METHODS:

We prospectively collected all samples from patients who received thyroid-stimulating hormone receptor antibodies testing in July 2018 at the University Hospital of Brest. The radioimmunoassay used was the Dynotest TRAK human by BRAHMS Diagnostica (Berlin, Germany). The Kryptor method used the BRAHMS TRAK human Kryptor kit performed with the Kryptor Compact Plus system.

RESULTS:

The inter-assay coefficient variations for the radioimmunological and Kryptor methods were 11.07% and 8.36%, respectively, with the low level quality control and 8.36% and 4.38%, respectively, with the high level quality control. Forty-four patients were included in the study including thirty-two Graves' disease patients in follow-up. The sensitivity of the radioimmunological method for the detection of Graves' disease was 0.94 and the specificity was 0.73. The sensitivity of the Kryptor method was 0.91 and the specificity was 0.91. A non-proportional systematic bias in favor of higher values of TRAb concentrations with the radioimmunological method was observed slope of 0.93 (0.74 - 1.07, 95% confidence interval) and an intercept of -0.69 IU/L (-1.58 to -0.30, 95% confidence interval). Compared to the Kryptor method, the radioimmunological method tends to overestimate TRAb concentrations by up to 120%.

CONCLUSIONS:

The fully automated Brahms Kryptor kit using TRACE technology to measure TRAb reduces sampling time and intra- as well as inter-assay variations. The Kryptor kit underestimates the results of TRAb leading to a lower sensitivity and higher specificity compared to the radioimmunoassay. Thus, the new Brahms Kryptor kit has good laboratory performances but the interpretation of the results must still be performed with caution.
Asunto(s)

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Tiroiditis / Receptores de Tirotropina / Radioinmunoensayo / Enfermedad de Graves / Inmunoglobulinas Estimulantes de la Tiroides / Hipotiroidismo Tipo de estudio: Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Lab Asunto de la revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2019 Tipo del documento: Article

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Tiroiditis / Receptores de Tirotropina / Radioinmunoensayo / Enfermedad de Graves / Inmunoglobulinas Estimulantes de la Tiroides / Hipotiroidismo Tipo de estudio: Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Adult / Female / Humans / Male / Middle aged Idioma: En Revista: Clin Lab Asunto de la revista: TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2019 Tipo del documento: Article