Preventing ceramic liner fracture after Delta ceramic-on-ceramic total hip arthroplasty.
Lee, Young-Kyun; Lim, Jae-Young; Ha, Yong-Chan; Kim, Tae-Young; Jung, Woon-Hwa; Koo, Kyung-Hoi.
Afiliación
  • Lee YK; Department of Orthopaedic Surgery, Seoul National University Bundang Hospital, Seongnam, South Korea.
  • Lim JY; Department of Orthopaedic Surgery, Inje University Haeundae Paik Hospital, Busan, South Korea.
  • Ha YC; Department of Orthopaedic Surgery, Chung-Ang University College of Medicine, 102 Heukseok-ro, Dongjak-gu, Seoul, 06973, South Korea. hayongch@naver.com.
  • Kim TY; Department of Orthopaedic Surgery, Konkuk University Hospital, Seoul, South Korea.
  • Jung WH; Department of Orthopaedic Surgery, Murup Hospital, Masan, South Korea.
  • Koo KH; Department of Orthopaedic Surgery, Seoul National University Bundang Hospital, Seongnam, South Korea.
Arch Orthop Trauma Surg ; 141(7): 1155-1162, 2021 Jul.
Article en En | MEDLINE | ID: mdl-32529389
BACKGROUND: The results of total hip arthroplasty (THA) with use of Delta ceramic articulation were successful at medium term follow-up. The use of this newest ceramic has markedly reduced the incidence of ceramic head fractures, but not the incidence of ceramic liner fractures. We tested a hypothesis that the ceramic fractures are prevented by use of a metal shell with 18° inner taper angle and a stem design with a reduced neck geometry. METHODS: We conducted a prospective multicenter study of cementless THA with use of 32/36-mm Delta ceramic bearing, Exceed ABT metal shell and Taperloc stem. We evaluated ceramic fracture, noise, clinical results, radiological changes and survival rate at a minimum of 5-year follow-up. RESULTS: From April 2010 to February 2012, 246 patients (274 THAs) were enrolled. Among them, 224 patients (130 men and 94 women, 250 hips) were followed-up for 5-8 years (mean 6.0 years). Ceramic malseating or fracture did not occur in any patient. Eight patients (8 hips, 3.2%) reported noise. Mean Harris hip score was 84 points at the latest follow-up. All acetabular and femoral components had bone-ingrown stability. Two hips were revised due to recurrent dislocation and periprosthetic fracture. The survival rate was 99.6% at 8 years postoperatively. CONCLUSION: Ceramic fractures can be prevented by a use of (1) 32/36-mm Delta ceramic bearing, (2) metal shell with 18° taper angle, and (2) stem with a reduced neck geometry. However, noise remains a concern of the Delta ceramic bearing. LEVEL OF EVIDENCE: Therapeutic level II.
Asunto(s)
Palabras clave

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Complicaciones Posoperatorias / Falla de Prótesis / Cerámica / Artroplastia de Reemplazo de Cadera / Prótesis de Cadera Tipo de estudio: Clinical_trials / Observational_studies Límite: Humans Idioma: En Revista: Arch Orthop Trauma Surg Año: 2021 Tipo del documento: Article País de afiliación: Corea del Sur

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Complicaciones Posoperatorias / Falla de Prótesis / Cerámica / Artroplastia de Reemplazo de Cadera / Prótesis de Cadera Tipo de estudio: Clinical_trials / Observational_studies Límite: Humans Idioma: En Revista: Arch Orthop Trauma Surg Año: 2021 Tipo del documento: Article País de afiliación: Corea del Sur