Comparative study of a modified double-lumen tube ventilation control connector and traditional connector in clinical use: a randomised-controlled trial.
Liu, Chang; Zhao, Yuanyu; Li, You; Guan, Huiwen; Feng, Junjie; Cheng, Shengquan; Wang, Xin; Wang, Yue; Sun, Xufang.
Afiliación
  • Liu C; Department of Anesthesiology, The Second Hospital of Jilin University, No.218, Ziqiang street, Nanguan District, Changchun City, 130000, Jilin Province, China.
  • Zhao Y; Department of Anesthesiology, The Second Hospital of Jilin University, No.218, Ziqiang street, Nanguan District, Changchun City, 130000, Jilin Province, China.
  • Li Y; Department of Anesthesiology, The Second Hospital of Jilin University, No.218, Ziqiang street, Nanguan District, Changchun City, 130000, Jilin Province, China.
  • Guan H; Department of Anesthesiology, The Second Hospital of Jilin University, No.218, Ziqiang street, Nanguan District, Changchun City, 130000, Jilin Province, China.
  • Feng J; Department of Anesthesiology, The Second Hospital of Jilin University, No.218, Ziqiang street, Nanguan District, Changchun City, 130000, Jilin Province, China.
  • Cheng S; Department of Anesthesiology, The Second Hospital of Jilin University, No.218, Ziqiang street, Nanguan District, Changchun City, 130000, Jilin Province, China.
  • Wang X; Department of Anesthesiology, The Second Hospital of Jilin University, No.218, Ziqiang street, Nanguan District, Changchun City, 130000, Jilin Province, China.
  • Wang Y; Department of Anesthesiology, The Second Hospital of Jilin University, No.218, Ziqiang street, Nanguan District, Changchun City, 130000, Jilin Province, China.
  • Sun X; Department of Anesthesiology, The Second Hospital of Jilin University, No.218, Ziqiang street, Nanguan District, Changchun City, 130000, Jilin Province, China. sunxufang20@163.com.
BMC Anesthesiol ; 22(1): 281, 2022 09 06.
Article en En | MEDLINE | ID: mdl-36068501
ABSTRACT

BACKGROUND:

A Y-shaped rotatable connector (YRC) for double-lumen tubes (DLT) is invented and compared with the traditional connector (Y-shaped connector, YC).

METHODS:

Sixty patients with ASA grade I-III, aged ≥ 18 years, who needed to insert a DLT for thoracic surgery were recruited and assigned into the YRC group (n = 30) and the YC group (n = 30) randomly. The primary endpoints included the inhaled air concentration (Fi) and the exhaled air concentration (Et) of sevoflurane before and after the switch between two-lung ventilation and one-lung ventilation at different times, positioning time, and switching time. The secondary endpoints were the internal gas volume of the two connectors, airway pressure, and the sputum suction time.

RESULTS:

The Et and Fi of the YRC group and the YC group were significantly different (all p < 0.05) at 5s, 10s, and 30s after the patient switched from two-lung ventilation to one-lung ventilation. The positioning time of the YRC group was less than YC group (89.75 ± 14.28 s vs 107.80 ± 14.96 s, p < 0.05), as well as the switching time (3.60 ± 1.20 s vs 9.05 ± 2.53 s, p < 0.05) and the internal gas volume (17.20 ml vs 24.12 ml). There was no difference in airway pressure and the sputum suction time in two groups.

CONCLUSION:

Compared with YC, YRC was beneficial for maintaining depth of anesthesia, improves efficiency for the switch between one-lung and two-lung ventilation, and shortens the tube positioning time.
Asunto(s)
Palabras clave

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Procedimientos Quirúrgicos Torácicos / Ventilación Unipulmonar / Anestesia Tipo de estudio: Clinical_trials Límite: Humans Idioma: En Revista: BMC Anesthesiol Año: 2022 Tipo del documento: Article País de afiliación: China

Texto completo: 1 Bases de datos: MEDLINE Asunto principal: Procedimientos Quirúrgicos Torácicos / Ventilación Unipulmonar / Anestesia Tipo de estudio: Clinical_trials Límite: Humans Idioma: En Revista: BMC Anesthesiol Año: 2022 Tipo del documento: Article País de afiliación: China