Software compatibility analysis for quantitative measures of [18F]flutemetamol amyloid PET burden in mild cognitive impairment.
Pemberton, Hugh G; Buckley, Christopher; Battle, Mark; Bollack, Ariane; Patel, Vrajesh; Tomova, Petya; Cooke, David; Balhorn, Will; Hegedorn, Katherine; Lilja, Johan; Brand, Christine; Farrar, Gill.
Afiliación
  • Pemberton HG; GE Healthcare, Pollards Wood, Chalfont St Giles, Amersham, HP8 4SP, UK. h.pemberton@ucl.ac.uk.
  • Buckley C; Centre for Medical Image Computing (CMIC), Department of Medical Physics and Bioengineering, University College London, London, UK. h.pemberton@ucl.ac.uk.
  • Battle M; UCL Queen Square Institute of Neurology, University College London, London, UK. h.pemberton@ucl.ac.uk.
  • Bollack A; GE Healthcare, Pollards Wood, Chalfont St Giles, Amersham, HP8 4SP, UK.
  • Patel V; GE Healthcare, Pollards Wood, Chalfont St Giles, Amersham, HP8 4SP, UK.
  • Tomova P; Clinical Memory Research Unit, Department of Clinical Sciences, Lund University, Malmö, Sweden.
  • Cooke D; Centre for Medical Image Computing (CMIC), Department of Medical Physics and Bioengineering, University College London, London, UK.
  • Balhorn W; GE Healthcare, Pollards Wood, Chalfont St Giles, Amersham, HP8 4SP, UK.
  • Hegedorn K; GE Healthcare, Pollards Wood, Chalfont St Giles, Amersham, HP8 4SP, UK.
  • Lilja J; Syntermed, Atlanta, GA, USA.
  • Brand C; MIM Software, Cleveland, OH, USA.
  • Farrar G; MIM Software, Cleveland, OH, USA.
EJNMMI Res ; 13(1): 48, 2023 May 24.
Article en En | MEDLINE | ID: mdl-37225974
ABSTRACT
RATIONALE Amyloid-ß (Aß) pathology is one of the earliest detectable brain changes in Alzheimer's disease pathogenesis. In clinical practice, trained readers will visually categorise positron emission tomography (PET) scans as either Aß positive or negative. However, adjunct quantitative analysis is becoming more widely available, where regulatory approved software can currently generate metrics such as standardised uptake value ratios (SUVr) and individual Z-scores. Therefore, it is of direct value to the imaging community to assess the compatibility of commercially available software packages. In this collaborative project, the compatibility of amyloid PET quantification was investigated across four regulatory approved software packages. In doing so, the intention is to increase visibility and understanding of clinically relevant quantitative methods.

METHODS:

Composite SUVr using the pons as the reference region was generated from [18F]flutemetamol (GE Healthcare) PET in a retrospective cohort of 80 amnestic mild cognitive impairment (aMCI) patients (40 each male/female; mean age = 73 years, SD = 8.52). Based on previous autopsy validation work, an Aß positivity threshold of ≥ 0.6 SUVrpons was applied. Quantitative results from MIM Software's MIMneuro, Syntermed's NeuroQ, Hermes Medical Solutions' BRASS and GE Healthcare's CortexID were analysed using intraclass correlation coefficient (ICC), percentage agreement around the Aß positivity threshold and kappa scores.

RESULTS:

Using an Aß positivity threshold of ≥ 0.6 SUVrpons, 95% agreement was achieved across the four software packages. Two patients were narrowly classed as Aß negative by one software package but positive by the others, and two patients vice versa. All kappa scores around the same Aß positivity threshold, both combined (Fleiss') and individual software pairings (Cohen's), were ≥ 0.9 signifying "almost perfect" inter-rater reliability. Excellent reliability was found between composite SUVr measurements for all four software packages, with an average measure ICC of 0.97 and 95% confidence interval of 0.957-0.979. Correlation coefficient analysis between the two software packages reporting composite z-scores was strong (r2 = 0.98).

CONCLUSION:

Using an optimised cortical mask, regulatory approved software packages provided highly correlated and reliable quantification of [18F]flutemetamol amyloid PET with a ≥ 0.6 SUVrpons positivity threshold. In particular, this work could be of interest to physicians performing routine clinical imaging rather than researchers performing more bespoke image analysis. Similar analysis is encouraged using other reference regions as well as the Centiloid scale, when it has been implemented by more software packages.