Assessment of the Efficiency of Measuring Foot and Ankle Edema with a 3D Portable Scanner.
Beldame, Julien; Sacco, Riccardo; Munoz, Marie-Aude; Masse, Marion; Lalevée, Matthieu.
Afiliación
  • Beldame J; ICP-Clinique Blomet, 136 bis Rue Blomet, 75015 Paris, France.
  • Sacco R; Clinique Megival, 1328 Avenue de la Maison Blanche, 76550 Saint Aubin sur Scie, France.
  • Munoz MA; Department of Orthopedic Surgery, Rouen University Hospital, 37 Bd Gambetta, 76000 Rouen, France.
  • Masse M; Centre Médical Achille, 200 Avenue des Prés d'Arènes, 34070 Montpellier, France.
  • Lalevée M; CKS, Centre kiné sport Dieppe, 32 Rue Louis Blériot, Neuville les Dieppe, 76370 Dieppe, France.
Bioengineering (Basel) ; 10(5)2023 May 03.
Article en En | MEDLINE | ID: mdl-37237619
ABSTRACT

Background:

To prospectively evaluate the reliability of a portable optical scanner compared to the water displacement technique for volumetric measurements of the foot and ankle and to compare the acquisition time associated with these two methods.

Methods:

Foot volume was measured in 29 healthy volunteers (58 feet, 24 females and 5 males) by a 3D scanner (UPOD-S 3D Laser Full-Foot Scanner®) and by water displacement volumetry. Measurements were performed on both feet, up to a height of 10 cm above the ground. The acquisition time for each method was evaluated. The Kolmogorov-Smirnov test, Lin's Concordance Correlation Coefficient, and a Student's t-test were performed.

Results:

Mean foot volume was 869.7 +/- 165.1 cm3 (3D scanner) versus 867.9 +/- 155.4 cm3 (water-displacement volumetry) (p < 10-5). The concordance of measurements was 0.93, indicative of a high correlation between the two techniques. Volumes were 47.8 cm3 lower when using the 3D scanner versus water volumetry. After statistically correcting this underestimation, the concordance was improved (0.98, residual bias = -0.03 +/- 35.1 cm3). The mean examination time was 4.2 +/- 1.7 min (3D optical scanner) versus 11.1 +/- 2.9 min (water volumeter) (p < 10-4).

Conclusions:

Ankle/foot volumetric measurements performed using this portable 3D scanner are reliable and fast and can be used in clinical practice and research.
Palabras clave

Texto completo: 1 Bases de datos: MEDLINE Idioma: En Revista: Bioengineering (Basel) Año: 2023 Tipo del documento: Article País de afiliación: Francia